← Product Code [JDS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDS) · K080706

# CONNEXX LOCKING TIBIA NAIL (K080706)

_I.T.S. Implantat-Technologie-Systeme GmbH · JDS · Jun 9, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDS/K080706

## Device Facts

- **Applicant:** I.T.S. Implantat-Technologie-Systeme GmbH
- **Product Code:** [JDS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDS.md)
- **Decision Date:** Jun 9, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The intended use of the I.T.S. CONNEXX Locking Tibia Nail is to stabilize fractures of the proximal and distal tibia and the tibial shaft in a pediatric or adult patient. Indications for use include open and closed tibial shaft fractures; certain pre and post-isthmic fractures, tibial malunions and non-unions, corrective osteotomies, pathologic fractures, pseudoarthrosis of the tibial shaft, metaphyseal and epiphyseal fractures, transverse fractures, oblique fractures, spiral fractures, segmental fractures, comminuted fractures, bone support, and fractures with bone loss.

## Device Story

CONNEXX Locking Tibia Nail is a cannulated intramedullary rod system for tibia fracture fixation. System includes 9mm/10mm diameter nails (255mm-420mm lengths) and 4.7mm cross-screws. Locking key sleeve and end cap mechanism provide angle-stable locking of cross-screws to nail. Used in clinical settings by surgeons; requires X-ray fluoroscopy for alignment and screw placement. Device stabilizes bone fragments to facilitate healing. High-strength 6-4 ELI Titanium Alloy construction.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing and design comparison to predicate devices.

## Technological Characteristics

Cannulated intramedullary nail system; 9mm and 10mm diameters; 6-4 ELI Titanium Alloy (ASTM F-136); angle-stable locking mechanism via cannulated locking key sleeve and end cap; manual surgical instrumentation; X-ray fluoroscopy guided placement.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes Tibial Nail System ([K040762](/device/K040762.md))
- DePuy/ACE VersaNail Tibial Nail ([K032097](/device/K032097.md))
- Synthes Cannulated Tibial Nail ([K962047](/device/K962047.md))
- Synthes Universal Tibial Nail & Unreamed Tibial Nail ([K914453](/device/K914453.md))
- BioPro Wujin #3 Tibial Nail ([K002325](/device/K002325.md))
- Zimmer Sirus Intramedullary Nail-Femoral and Tibial Nail ([K043270](/device/K043270.md))
- Stryker Interlocking Tibial Nailing System ([K915435](/device/K915435.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K080706 p.14

| Section XII:                        | 510(k) Summary of Safety and Effectiveness<br>SAFE MEDICAL DEVICES ACT OF 1990<br>510(k) Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| NAME OF FIRM:                       | JUN - 9 2008<br>I.T.S. Implantat-Technologie-Systeme GmbH.<br>Autal 28.<br>Lassnitzhoehe A 8301<br>AUSTRIA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| 510(k) FIRM CONTACT:                | Al Lippincott<br>Engineering Consulting Services, Inc.<br>3150 E. 200th St.<br>Prior Lake, MN 55372                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| TRADE NAME:                         | CONNEXX Locking Tibia Nail                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| COMMON NAME:                        | Intramedullary Nail System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| CLASSIFICATION:                     | Single/multiple component metallic bone fixation appliances and<br>accessories (see 21 CFR, Sec. 888.3040)<br><br>Rod, Fixation, Intramedullary and Accessories<br>(see 21 CFR, Sec. 888.3020)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| DEVICE PRODUCT CODE:                | HWC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| SUBSEQUENT PRODUCT CODE:            | HSB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| SUBSTANTIALLY<br>EQUIVALENT DEVICES | Synthes Tibial Nail System (K040762)<br>DePuy/ACE VersaNail Tibial Nail (K032097)<br>Synthes Cannulated Tibial Nail (K962047)<br>Synthes Universal Tibial Nail & Unreamed Tibial Nail (K914453)<br>BioPro Wujin #3 Tibial Nail (K002325)<br>Zimmer Sirus Intramedullary Nail-Femoral and Tibial Nail (K043270)<br>Stryker Interlocking Tibial Nailing System (K915435)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| DEVICE DESCRIPTION:                 | The I.T.S. CONNEXX Locking Tibia Nail is a fracture fixation<br>intramedullary rod with locking/nonlocking cross-screw fixation for<br>repairing long bone tibia fractures of a patient. The I.T.S. CONNEXX<br>Locking Tibia Nail System is composed of a cannulated 9mm and 10mm<br>diameter nail in various lengths (255mm to 420mm in 15mm increments)<br>and two styles of cross-screw attachments, a 4.7mm standard cortical<br>screw as a non-locking feature and a 4.7mm Bolt Double Thread as a<br>non-locking feature (two thread diameters on the screw for thread<br>engagement into the proximal and distal cortex of the tibial shaft when<br>inserted). A cannulated 'locking key sleeve' and 'end cap' combination<br>function to firmly fix the above two cross-screw attachments to the 9mm<br>or 10mm nail for 'locking' the cross-screws to the tibia nail - thus<br>making the entire IM rod/cross-screw assembly 'angle-stable locking'.<br>All tibial rods and cross-screws are composed of high strength 6-4 ELI<br>Titanium Alloy to ASTM F-136. A full compliment of instrumentation<br>is available for use with the system. |

{1}------------------------------------------------

Proper alignment instrumentation is incorporated for proper postioning of the locking screw to screw hole in the IM rod -- along with the assistance of X-ray Fluoroscopy. The intended use of the I.T.S. CONNEXX Locking Tibia Nail is to INTENDED USE: stabilize fractures of the proximal and distal tibia and the tibial shaft in a pediatric or adult patient. Indications for use include open and closed tibial shaft fractures; certain pre and post-isthmic fractures, tibial malunions and non-unions, corrective osteotomies, pathologic fractures, pseudoarthrosis of the tibial shaft, metaphyseal and epiphyseal fractures, transverse fractures, oblique fractures, spiral fractures, segmental fractures, comminuted fractures, bone support, and fractures with bone loss. BASIS OF SUBSTANTIAL The I.T.S. CONNEXX Locking Tibia Nail is substantially EQIVLAENCE: equivalent to the Synthes, DePuy/ACE, Zimmer, Stryker, and Biopro IM Tibia Nail systems. SUMMARY OF SAFETY AND EFFECTIVENESS: The I.T.S. CONNEXX Locking Tibia Nail is shown to be safe and effective for use in fracture fixation of the tibia.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol of medicine and health. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 9 2008

I.T.S. Implantat-Technologie-Systeme GmbH % Engineering Consulting Services, Inc. Mr. Al Lippincott 3150 E. 200th Street Prior Lake, Minnesota 55372

Re: K080706

Trade/Device Name: CONNEXX Locking Tibia Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: II Product Code: HSB, JDS, HWC Dated: March 10, 2008 Received: March 12, 2008

Dear Mr. Lippincott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or unly with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set OF I art 807); ideeing (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

## Page 2 - Mr. Al Lippincott

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Milliman

- Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{4}------------------------------------------------

## Indications for Use 510(k) NUMBER: K080706 DEVICE NAME: CONNEXX Locking Tibia Nail INDICATIONS FOR USE:

Indications for use include open and closed tibial shaft fractures; certain pre and post-isthmic fractures, tibial malunions and non-unions, corrective osteotomies, pathologic fractures, pseudoarthrothis of the tibial shaft, mataphyseal and epiphyseal fractures, transverse fractures, oblique fractures, spiral fractures, segmental fractures, comminuted fractures, bone transport, and fractures with bone loss.

Prescription Use Yes AND/OR Over-The-Counter-Use

se No

(Per 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nil R.P.O.sal for mam

(Division Division of General. Restorative. and Neurological Devices

510(k) Number_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Section XI

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDS/K080706](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDS/K080706)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
