← Product Code [JDR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR) · K220181

# StealthFix Intraosseous Fixation System (K220181)

_Nextremity Solutions, Inc. · JDR · Mar 22, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K220181

## Device Facts

- **Applicant:** Nextremity Solutions, Inc.
- **Product Code:** [JDR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR.md)
- **Decision Date:** Mar 22, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The StealthFix Intraosseous Fixation System is indicated for fixation of bone fractures, fusions, or for bone reconstructions, including: - Arthrodesis in hand or foot surgery - Mono or bi-cortical osteotomies in the foot or hand - Fracture management in the foot or hand - Distal or proximal metatarsal or metacarpal osteotomies - Fixation of osteotomies for Hallux Valgus treatment such as scarf, chevron, etc.

## Device Story

StealthFix Intraosseous Fixation System is an orthopedic implant system comprising metallic staples and crossing screws; used for bone fixation, fusion, or reconstruction in hand/foot surgery. Staples feature two legs/posts connected by a bridge; standard staples include a screw slot for optional crossing screw fixation. System includes accessory instruments (inserters, targeting arms) for site preparation and implant insertion. Implants are provided sterile or non-sterile for single use; instruments are provided non-sterile (reusable or single-use) or sterile. Surgeons use the system in clinical settings to stabilize bone segments; implants provide mechanical fixation to facilitate healing. System benefits patients by providing rigid internal fixation for fractures and osteotomies.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing and engineering analysis.

## Technological Characteristics

Metallic bone fixation system (staples and screws). Implants available in various diameters (2.5mm, 3.5mm, 4.5mm) and lengths. Mechanical testing performed per ASTM F564 (pullout force) and ASTM F543 (torsional/pullout strength). Endotoxin testing per AAMI ST72, USP 161, and USP 85. Sterilization required for non-sterile components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Stealth Staple System ([K163440](/device/K163440.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

March 22, 2022

Nextremity Solutions, Inc. Elise Fox Ouality and Regulatory Specialist 1195 Polk Drive Warsaw, Indiana 46582

Re: K220181

Trade/Device Name: StealthFix Intraosseous Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDR, HWC Dated: January 20, 2022 Received: January 21, 2022

Dear Elise Fox:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Limin Sun, Ph.D. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known) K220181

Device Name StealthFix Intraosseous Fixation System

Indications for Use (Describe)

The StealthFix Intraosseous Fixation System is indicated for fixation of bone fractures, fusions, or for bone reconstructions, including:

- · Arthrodesis in hand or foot surgery
- · Mono or bi-cortical osteotomies in the foot or hand
- · Fracture management in the foot or hand
- · Distal or proximal metatarsal or metacarpal osteotomies
- · Fixation of osteotomies for Hallux Valgus treatment such as scarf, chevron, etc.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 510(k) Summary    |                                                                                                                                                                                                           |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared:         | March 22, 2022                                                                                                                                                                                            |
| Submitter:        | Nextremity Solutions, Inc.<br>1195 Polk Drive<br>Warsaw, IN 46582                                                                                                                                         |
| Contact:          | Elise Fox<br>Quality and Regulatory Specialist<br>elise.fox@nextremity.com<br>Phone: 574-376-2062                                                                                                         |
| Proprietary Name: | StealthFix Intraosseous Fixation System                                                                                                                                                                   |
| Common Name:      | Bone Staple System                                                                                                                                                                                        |
| Classification:   | 21 CFR § 888.3030 (Primary) Single/multiple component metallic bone<br>fixation appliances and accessories; Class II<br>21 CFR § 888.3040 Smooth or threaded metallic bone fixation fastener;<br>Class II |
| Product Code:     | JDR (Primary), HWC                                                                                                                                                                                        |
| Predicate Device: | K163440- Stealth Staple System, First Ray LLC                                                                                                                                                             |

### Device Description:

The StealthFix Intraosseous Fixation System is an orthopedic intraosseous staple system consisting of staple and screw implants. The staples consist of two legs or posts connected by a bridge. The staples are available in post diameters of 2.5mm(mini), 3.5mm(small) and 4.5mm(standard). The 2.5mm staples are provided with a bridge span of 10mm and range in post length from 8mm to 12mm. The 3.5mm staples are provided with a bridge span of 15mm and range in post length from 14mm to 20mm. The 4.5mm staples are available in bridge spans of 15mm and 20mm and range in post length from 14mm to 32mm. The system provides crossing screws for optional fixation with the standard staple implants. Standard staples are designed with a screw slot to accept a crossing screw. The screws are 3.5mm in diameter with lengths ranging from 16mm to 38mm in 2mm increments. The system provides accessory instruments designed for preparation of the implant site and insertion of implants into bone, including implant specific inserters and targeting arms. The implants of the system are available packaged both sterile and non-sterile for single use. The instruments are provided non-sterile, reusable or non-sterile, single use and must

{4}------------------------------------------------

be cleaned and sterilized by the end user prior to use. The system also provides some instruments sterile packaged, individually and in sets.

#### Intended Use / Indications:

The StealthFix Intraosseous Fixation System is indicated for fixation of bone fractures, fusions, or for bone reconstructions, including:

- Arthrodesis in hand or foot surgery .
- Mono or bi-cortical osteotomies in the foot or hand
- Fracture management in the foot or hand •
- . Distal or proximal metatarsal or metacarpal osteotomies
- . Fixation of osteotomies for Hallux Valgus treatment such as scarf, chevron, etc.

## Summary of Technologies/Substantial Equivalence:

The StealthFix Intraosseous Fixation System is substantially equivalent to the predicate devices regarding the intended use and indications, material, design, sizes, and mechanical properties. The subject and predicates are both staple and screw implant systems manufactured from the same materials and are available in the same sizes. The subject system differs because the staple posts internal thread length was increased and because it contains non-cannulated screws.

#### Non-Clinical Testing:

Endotoxin testing was performed using the Limulus Amebocyte Lysate (LAL) method according to AAMI ST72, USP 161 and USP 85. Results met the Endotoxin limit of ≤20 EU per device. An engineering analysis was performed to ensure that the modified staples do not create a new worst case for static and dynamic 4 point bend testing, and pullout force (ASTM F564). An engineering analysis was performed to ensure that the modified screws do not create a new worst case for torsional strength, pullout strength or insertion performance (ASTM F543).

### Clinical Testing:

Clinical testing was not necessary to demonstrate substantial equivalence of the StealthFix Intraosseous Fixation System to the predicate device.

### Conclusions/Substantial Equivalence:

The StealthFix Intraosseous Fixation System is substantially equivalent to the predicate devices regarding its intended use, material, design, sizes, and mechanical properties. Differences between the subject device system and the predicate device systems do not raise different types of safety and effectiveness questions.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K220181](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K220181)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
