← Product Code [JDR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR) · K141550

# INSTAFIX SHAPE MEMORY FIXATION SYSTEM (K141550)

_Ot Medical · JDR · Jul 11, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K141550

## Device Facts

- **Applicant:** Ot Medical
- **Product Code:** [JDR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR.md)
- **Decision Date:** Jul 11, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

InstaFix™ Shape Memory Fixation System is intended for use in: Fixation of Osteotomies of the Hand, Foot and Tibia Arthrodesis of the Joints of the Hand and Foot Fixation of Soft Tissue to Bone, as in the case of the Anterior Cruciate Ligament InstaFix™ Shape Memory Fixation System is also indicated for adjunctive fixation of Small Bone Fragments of: The Upper Extremity, such as the Radius, Ulna, Humerus, Clavicle and Scapula The Lower Extremity, such as the Tibia, Fibula and the Femur The Upper Torso, such as the Sternum and the Ribs

## Device Story

InstaFix™ is a shape memory fixation system utilizing NiTinol (nickel-titanium alloy) staples; implants deflect upon implantation to generate continuous compressive force across bone fixation sites. Used in orthopedic procedures for osteotomies, arthrodesis, and soft tissue-to-bone attachment. System includes sterile-packaged staples (8mm-30mm), drill guides, implant holders, and accessory kits (drills, locating pins). Operated by surgeons in clinical/OR settings; provides low-profile fixation with minimal bone disruption compared to traditional screws/plates. Compressive force maintained throughout healing process promotes stability. Device is single-use.

## Clinical Evidence

Bench testing only. Verification and validation included staple fatigue life, pull-out fixation strength, elastic static bending, sterilization validation, packaging validation, sealing, and aging studies.

## Technological Characteristics

NiTinol (nickel-titanium) shape memory alloy; 8mm-30mm staple sizes; gamma radiation sterilization; single-use implants and accessories; low-profile design.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- reVERTO™ Shape Memory Staples ([K071477](/device/K071477.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

·

JUL 1 1 2014

:

## 510(K) SUMMARY

| Owner's Name                     | William Rhoda                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's                      | OT Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Address                          | 1000 Continental Drive, Suite 240<br>King of Prussia, PA 19034                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Phone Number                     | (484) 588-2063                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Fax Number                       | (484) 588-2064                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| 510(k) Owner                     | OT Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Contact Person                   | William Rhoda, President & CEO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Date Prepared                    | June, 10 2014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Trade Name                       | InstaFix™ Shape Memory Fixation System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name                      | Staple, Fixation, Bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification                   | Sec. 888.3030 Single/multiple component metallic bone fixation<br>appliances and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Section                          | Orthopedic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code                     | JDR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Predicate Device                 | reVERTO™ Shape Memory Staples (K071477)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device<br>Description            | InstaFix™ Shape Memory Fixation System is a set of dynamic<br>compression implants made from shape memory metal nickel-<br>titanium alloy called NiTinol. The staples are available in sizes 8mm<br>to 30mm. The NiTinol construction causes the legs of the implant to<br>deflect, thereby creating a compressive force across the site of<br>fixation. InstaFix™ Shape Memory Fixation implants are<br>advantageous to screws and plates in that they require less<br>disruption of the surrounding bone for implantation and provide<br>'low-profile' fixation. The most important factor related to the use<br>of NiTinol is its ability to maintain a compressive force across the<br>site of fixation during the healing process. Each staple is packaged<br>sterile along with the corresponding drill guide and implant holder.<br>The one time use Accessory Kit and the Sizer Kit are packaged<br>sterile to assist in the procedure. The Accessory Kit contains one<br>drill and two locating pins. The Sizer Kit contains samples of the<br>available staple sizes. |
| Reason for<br>510(k)             | Modified device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Indications for<br>Use           | InstaFix™ Shape Memory Fixation System is intended for use in:<br>Fixation of Osteotomies of the Hand, Foot and Tibia<br>Arthrodesis of the Joints of the Hand and Foot<br>Fixation of Soft Tissue to Bone, as in the case of the<br>Anterior Cruciate Ligament<br>InstaFix™ Shape Memory Fixation System is also indicated for<br>adjunctive fixation of Small Bone Fragments of:<br>The Upper Extremity, such as the Radius, Ulna, Humerus,<br>Clavicle and Scapula<br>The Lower Extremity, such as the Tibia, Fibula and the<br>Femur<br>The Upper Torso, such as the Sternum and the Ribs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Technological<br>Characteristics | The InstaFix™ Shape Memory Fixation System implant is made of<br>the same Nitinol material as the predicate. The implant is single use<br>and sterilized by gamma radiation, which is the same as the<br>predicate. The accessories are single use versus reusable. The staple<br>sizes available are the same as the predicate.<br><br>All patient contact components are made from biocompatible<br>materials. This indicates that the components are safe for their<br>intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Substantial<br>Equivalence       | The InstaFix™ Shape Memory Fixation System is substantially<br>equivalent in design, materials, intended use, indications for use,<br>principles of operations and technological characteristics as the<br>predicate reVERTO™ Shape Memory Staples K071477. There are<br>differences in the packaging and the accessories are single use<br>instead of reusable. The shelf life and sterilization method is the<br>same. The risk analysis performed, raised no new issues of safety or<br>efficacy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Nonclinical Test<br>Performed    | The verification and validation testing of the InstaFix™ Shape<br>Memory Fixation implant included staple fatigue life, pull-out<br>fixation strength and elastic static bending. Design verification and<br>validation testing included sterilization validation, packaging<br>validation, sealing and aging study.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Conclusions<br>Drawn             | The substantial equivalence of the InstaFix™ Shape Memory<br>Fixation System is based on the equivalence in intended use,<br>materials, operational principals, technological characteristics and<br>indications for use to reVERTO™ Shape Memory Staples K071477.<br>Non-clinical testing demonstrates that the proposed device is safe,<br>effective, and performs as well as the predicate device, therefore<br>demonstrating substantial equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

.

.

. . . . . .

{1}------------------------------------------------

# END OF 510(K) SUMMARY

and the comments of the comments of

.

:

:

-

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is in a bold, sans-serif font. The text is black and the background is white. 

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 11, 2014

OT Medical Mr. William Rhoda President & CEO 1000 Continental Drive, Suite 240 King of Prussia, Pennsylvania 19406

Re: K141550 Trade/Device Name: InstaFix™ Shape Memory Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR Dated: June 11, 2014 Received: June 12, 2014

Dear Mr. Rhoda:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

#### Page 2 - Mr. William Rhoda

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Lori A. Wigqins

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Statement of Indications for Use

510 (K) NUMBER IF KNOWN: _K141550 DEVICE NAME: InstaFix™ Shape Memory Fixation System

Indications for Use:

InstaFix™ Shape Memory Fixation System is intended for use in:

Fixation of Osteotomies of the Hand, Foot and Tibia

Arthrodesis of the Joints of the Hand and Foot

Fixation of Soft Tissue to Bone, as in the case of the Anterior Cruciate Ligament InstaFix™ Shape Memory Fixation System is also indicated for adjunctive fixation of Small Bone Fragments of:

> The Upper Extremity, such as the Radius, Ulna. Humerus. Clavicle and Scapula The Lower Extremity, such as the Tibia, Fibula and the Femur

The Upper Torso, such as the Sternum and the Ribs

Prescription Use ☑ AND/OR (Part 21 CFR 801 Subpart D)

Over-the-Counter Use O (21 CFR 801 Subpart C)

### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE, IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley, Ph.D.

---

Division of Orthopedic Devices

Page 1/1

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K141550](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K141550)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
