← Product Code [JDR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR) · K122113

# MEMORY METAL STAPLES, EASYCLIP (K122113)

_Stryker Corp. · JDR · Mar 12, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K122113

## Device Facts

- **Applicant:** Stryker Corp.
- **Product Code:** [JDR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR.md)
- **Decision Date:** Mar 12, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The MEMORY METAL STAPLES (MEMOCLIP, EASYCLIP, and FOR FUSION) are indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

## Device Story

Easy Clip is a metallic bone fixation staple used for osteotomy fixation and joint arthrodesis in hand and foot surgery. The device functions as a mechanical implant to stabilize bone fragments. It is intended for use by surgeons in a clinical or hospital setting. The device is a modification of the previously cleared Memometal Memory Staples (K070031), featuring additional sizes and design modifications. It provides mechanical support to facilitate bone healing and fusion.

## Clinical Evidence

No clinical data. Bench testing only, including dimensional analysis, engineering strength analysis, static and dynamic bending tests, and corrosion testing per ASTM F2129.

## Technological Characteristics

Metallic bone fixation staple. Materials and performance characteristics are equivalent to predicate K070031. Corrosion testing performed per ASTM F2129. Mechanical fixation via staple geometry.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Memometal Memory Staples ([K070031](/device/K070031.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

| Proprietary Name:                  | Easy Clip                                                                                       | MAR 1 2 2013 |
|------------------------------------|-------------------------------------------------------------------------------------------------|--------------|
| Common Name:                       | Staple, Fixation, Bone                                                                          |              |
| Classification Name and Reference: | Single/multiple component metallic bone fixation<br>appliances and accessories 21 CFR §888.3030 |              |
| Regulatory Class:                  | Class II                                                                                        |              |
| Product Codes:                     | 87 JDR- Single/multiple component metallic bone fixation<br>appliances and accessories          |              |
| For Information contact:           | Estela Celi, Regulatory Affairs Specialist<br>Stryker Trauma AG                                 |              |
|                                    | Phone: (201) 831-6461 Fax:# (201) 831-3461                                                      |              |
| Date Prepared:                     | July 16,2012                                                                                    |              |

# Description:

· This Traditional 510(k) submission is being supplied to the U.S. FDA to provide authorization to market Easy Clip. Additional sizes to the existing range and modifications are being made existing cleared *Easy Clip' devices (K070031).

# Intended Use:

The MEMORY METAL STAPLES (MEMOCLIP, EASYCLIP, and FOR FUSION) are indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

# Indications:

The MEMORY METAL STAPLES (MEMOCLIP, EASYCLIP, and FOR FUSION) are indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

# Summary of Technologies:

Device comparison showed that the proposed device is substantially equivalent in intended, use, materials and performance characteristics to the following predicate devices:

- . K070031 - Memometal Memory Staples
# Non-Clinical Testing:

Dimensional (geometric cross section) and engineering strength analyses as well as static and dynamic bending tests were conducted and the results support the conclusion that there are no effects of the modifications subject to this premarket notification on the safety and effectiveness of the EasyClip staples. Corrosion Testing as per ASTM F2129 was also performed.

Clinical Testing: Clinical testing was not required for this submission.

Conclusion: The additional Easy Clip staples are substantially equivalent to the predicate devices identified in this premarket notification.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized graphic of an eagle or bird-like figure with flowing lines, representing the department's mission related to health and human services.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Letter dated: March 12, 2013

Stryker. Corporation % Ms. Estela Celi Regulatory Affairs Specialist 132 Corporate Drive Mahwah. New Jersey 07430

Re: K122113

Trade/Device Name: Memory Metal Staples Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: JDR Dated: February 25, 2013

Received: February 26, 2013

Dear Ms. Celi:

i i

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{2}------------------------------------------------

### Page 2 - Ms. Estela Celi

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark NEmelkerson -S

Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### · Indications for Use

510(k) Number (if known): K122113

Device Name: Memory Metal Staples

Indications for Use:

The MEMORY METAL STAPLES (MEMOCLIP, EASYCLIP, and FOR FUSION) are indicated for hand and foot bone fragments osteotomy fixation and joint arthrodesis.

Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED).

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley, Ph.D.
Division of Orthopaedic Devices

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K122113](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K122113)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
