Synthes (Usa) Products, LLC / Depuy Orthopaedics, Inc.
Product Code
JDQ · Orthopedic
Decision Date
Jun 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The DePuy Synthes 5.0mm and 7.3mm Crimp Positioning Pins are intended for use with multifilament cable to augment fracture stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants. The 5.0mm Crimp Positioning Pins are designed for use in the 5.0mm LCP holes of Synthes plates. The 7.3mm Crimp Positioning Pins are designed for use in the 7.3mm locking holes of Synthes plates. The DePuy Synthes Variable Angle Positioning Pins are intended for use with cerclage multifilament cable to augment fracture stabilization with plates used in long bone fixation, when screw placement would be inhibited, as in the presence of intramedullary implant. The Variable Angle Positioning Pins are designed for use with Variable Angle LCP plate implants featuring variable angle locking holes that accept 5.0 mm variable angle bone screws. The DePuy Synthes Wire Mount is intended for use with cerclage wire to augment fracture stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants. The Wire Mount is designed for use in dynamic compression screw holes that accept a 4.5 mm bone screw. The Wire Mount (and cerclage wire) can be used with a variety of plates that include, but are not limited to the following. Dynamic Hip Screw (DHS) Plates, Dynamic Condylar Screw (DCS) Plates, Condylar Buttress Plates, Hook Plates, 4.5 mm Narrow and Broad Dynamic Compression Plates (DCP), 90° Child and Adolescent Osteotomy Plates, 110-130° Adult Osteotomy Plates, and 95° Condylar and 130° Angle Blade Plates (both including small stature plates). The DePuy Synthes Cerclage Positioning Pins are intended for use with cerclage monofilament wire and multifilament cable to augment fracture stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants. Cerclage Positioning Pins are designed for use with plates having locking compression or dynamic compression screw holes that accept a 3.5 mm or 4.5 mm bone screws. The Cerclage Positioning Pin (and cerclage wire or cable) can be used with a variety of Synthes plates.
Device Story
Metallic positioning pins and wire mounts; used as guides and fixation points for cerclage wire or multifilament cable; augment fracture stabilization with orthopedic plates; used in long bone fixation when screws are contraindicated (e.g., intramedullary implants present); operated by surgeons in clinical settings; provides stable structure for wire/cable tensioning; MR conditional status added to existing devices; non-clinical testing confirms safety in MR environment.
Clinical Evidence
Bench testing only. Assessment of magnetically induced displacement force (ASTM F2052-14), torque (ASTM F2213-06), RF heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07) performed to support MR Conditional status.
Technological Characteristics
Metallic implants (stainless steel/titanium); form factors include crimp pins, variable angle pins, wire mounts, and cerclage positioning pins; designed for integration with specific Synthes plate screw holes (3.5mm, 4.5mm, 5.0mm, 7.3mm); MR conditional per ASTM F2052-14, F2213-06, F2182-11a, F2119-07.
Indications for Use
Indicated for patients requiring fracture stabilization with plates in long bone fixation where screw placement is contraindicated or inhibited, such as in the presence of intramedullary implants.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized symbol. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Synthes (USA) Products LLC/DePuy Orthopaedics Inc Suchitra Basu Global Strategy Manager, Regulatory Affairs 1301 Goshen Parkway West Chester, Pennsylvania 19380
June 21, 2018
### Re: K172975
Trade/Device Name: DePuy Synthes (USA) 5.0mm / 7.3mm Crimp Positioning Pins - MR Conditional, DePuy Synthes Variable Angle Positioning Pins - MR Conditional, DePuy Synthes Wire Mount - MR Conditional, DePuy Synthes Cerclage Positioning Pin - MR Conditional Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ Dated: May 11, 2018 Received: May 14, 2018
Dear Suchitra Basu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for DePuy Synthes. The logo consists of a geometric shape on the left and the text "DePuy Synthes" on the right. The geometric shape is a sphere made up of several smaller shapes. The text is in a sans-serif font, with "DePuy" in a darker blue and "Synthes" in a lighter blue.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K172975
Device Name
DePuy Synthes (USA) 5.0mm / 7.3mm Crimp Positioning Pins - MR Conditional
#### Indications for Use (Describe)
The DePuy Synthes 5.0mm and 7.3mm Crimp Positioning Pins are intended for use with multifilament cable to augment fracture stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants.
The 5.0mm Crimp Positioning Pins are designed for use in the 5.0mm LCP holes of Synthes plates. The 7.3mm Crimp Positioning Pins are designed for use in the 7.3mm locking holes of Synthes plates.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------|
| <span style="font-size:12px;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
FORM FDA 3881 (7/17)
PSC Pablishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for DePuy Synthes. The logo consists of a geometric shape on the left and the text "DePuy Synthes" on the right. The geometric shape is a sphere made up of intersecting lines, and the text is in a sans-serif font. The logo is in a blue and gold color scheme.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K172975
Device Name
DePuy Synthes Variable Angle Positioning Pins - MR Conditional
#### Indications for Use (Describe)
The DePuy Synthes Variable Angle Positioning Pins are intended for use with cerclage multifilament cable to augment fracture stabilization with plates used in long bone fixation, when screw placement would be inhibited, as in the presence of intramedullary implant.
The Variable Angle Positioning Pins are designed for use with Variable Angle LCP plate implants featuring variable angle locking holes that accept 5.0 mm variable angle bone screws.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for DePuy Synthes. The logo consists of a circular symbol on the left and the text "DePuy Synthes" on the right. The circular symbol is made up of several curved lines that intersect to form a spherical shape. The text "DePuy Synthes" is written in a sans-serif font, with "DePuy" in a slightly larger font size than "Synthes".
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K172975
Device Name
DePuy Synthes Wire Mount - MR Conditional
Indications for Use (Describe)
The DePuy Synthes Wire Mount is intended for use with cerclage wire to augment fracture stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants.
The Wire Mount is designed for use in dynamic compression screw holes that accept a 4.5 mm bone screw. The Wire Mount (and cerclage wire) can be used with a variety of plates that include, but are not limited to the following. Dynamic Hip Screw (DHS) Plates, Dynamic Condylar Screw (DCS) Plates, Condylar Buttress Plates, Hook Plates, 4.5 mm Narrow and Broad Dynamic Compression Plates (DCP), 90° Child and Adolescent Osteotomy Plates, 110-130° Adult Osteotomy Plates, and 95° Condylar and 130° Angle Blade Plates (both including small stature plates).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for DePuy Synthes. The logo consists of a circular graphic to the left of the company name. The graphic is a tan color and appears to be a stylized representation of a sphere. The company name, "DePuy Synthes", is written in a dark blue font.
T
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Indications for Use |
|---------------------|
|---------------------|
| 510(k) Number (if known) | K172975 |
|--------------------------|--------------------------------------------------------|
| Device Name | DePuy Synthes Cerclage Positioning Pin- MR Conditional |
| Indications for Use (Describe) |
|--------------------------------|
|--------------------------------|
The DePuy Synthes Cerclage Positioning Pins are intended for use with cerclage monofilament wire and multifilament cable to augment fracture stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants.
Cerclage Positioning Pins are designed for use with plates having locking compression or dynamic compression screw holes that accept a 3.5 mm or 4.5 mm bone screws. The Cerclage Positioning Pin (and cerclage wire or cable) can be used with a variety of Synthes plates.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image features the logo of DePuy Synthes, a company specializing in medical devices and orthopedic solutions. The logo consists of two distinct elements: a stylized, abstract symbol on the left and the company name "DePuy Synthes" on the right. The symbol is rendered in a light brown or gold color, while the company name is displayed in a dark blue font.
# 510(K) Summary for K081205:
Date Prepared: June 20, 2018
# 1.1. Submitter
Primary Contact: Suchitra Basu, PhD, RAC Global Strategy Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5448 sbasu23@its.jnj.com
Alternate Contact: Elizabeth Jacobs Senior Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5768 Ejacob10@its.jnj.com
### 1.2. Device
Name of Device: DePuy Synthes (USA) 5.0mm / 7.3mm Crimp Positioning Pins – MR Conditional Classification Name(s): Bone Fixation Cerclage Common or Usual Name(s): crimp, pin, Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II; 888.3010 Product Code(s): JDQ
### 1.3. Predicate Device
Synthes 5.0mm/7.3mm (USA) Crimp Positioning Pins (K081205)
### 1.3.1 Reference Device
DePuy Synthes Trauma Orthopedic Implant Portfolio – MR Conditional (K160553)
### 1.4. Device Description
The DePuy Synthes 5.0 mm and 7.3 mm Crimp Positioning Pins are designed to fit into the screw holes of existing DePuy Synthes fixation plates for the purpose of providing a guide for Cerclage cable positioning as well as a crimping point for Cerclage cable tension fixation. The Crimp Positioning pins are available in versions composed of implant quality stainless steel and titanium.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for DePuy Synthes. The logo consists of a circular graphic on the left and the text "DePuy Synthes" on the right. The circular graphic is made up of several curved lines that intersect to form a spherical shape. The text "DePuy Synthes" is written in a sans-serif font and is colored in dark blue.
# 1.5. Indications for Use
The DePuy Synthes 5.0 mm and 7.3 mm Crimp Positioning Pins are intended for use with multifilament cable to augment fracture stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants.
The 5.0 mm Crimp Positioning Pins are designed for use in the 5.0 mm locking or 4.5 mm LCP holes of Synthes plates. The 7.3 mm Crimp Positioning Pins are designed for use in the 7.3 mm locking holes of Synthes Plates.
# 1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes 5.0 mm and 7.3 mm Crimp Positioning Pins. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes 5.0 mm and 7.3 mm Crimp Positioning Pins in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for DePuy Synthes. The logo consists of a geometric shape on the left and the text "DePuy Synthes" on the right. The geometric shape is a sphere made up of several smaller shapes. The text is in a sans-serif font and is colored blue.
# 510(k) Summary for K162124:
Date Prepared: June 20, 2018
#### 1.1. Submitter
Primary Contact: Suchitra Basu, PhD, RAC Global Strategy Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5448 sbasu23@its.jnj.com
Alternate Contact: Elizabeth Jacobs Senior Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5768 Ejacob10@its.jnj.com
#### 1.2. Device
Name of Device: DePuy Synthes Variable Angle Positioning Pins – MR Conditional Classification Name(s): Cerclage, Fixation Common or Usual Name(s): pin, Smooth or thread metallic bone fixation fastener Regulatory Class: Class II; 888.3010 Product Code(s): JDQ
#### 1.3. Predicate Device
Synthes Variable Angle Positioning Pins (K162124)
### 1.3.1 Reference Device
DePuy Synthes Trauma Orthopedic Implant Portfolio - MR Conditional (K160553)
#### 1.4. Device Description
The VA 5.0 Positioning Pins are designed for use with Variable Angle LCP plate implants featuring variable angle locking holes that accept 5.0mm variable angle bone screws. Positioning or cerclage pins are intended for use with cerclage monofilament wire or multifilament cable to augment fracture stabilization with plates used in long bone fixation, when screw placement would be inhibited, as in the presence of intramedullary implant.
# 1.5. Indications for Use
The DePuy Synthes Variable Angle Positioning Pins are intended for use with cerclage multifilament cable to augment fracture stabilization with plates used in long bone fixation, when screw placement would be inhibited, as in the presence of intramedullary implant.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for DePuy Synthes. The logo consists of a geometric shape on the left and the text "DePuy Synthes" on the right. The geometric shape is a sphere made up of intersecting lines, and the text is in a sans-serif font. The logo is simple and modern, and it is likely used to represent the company's brand.
The Variable Angle Positioning Pins are designed for use with Variable Angle LCP plate implants featuring variable angle locking holes that accept 5.0 mm variable angle bone screws.
# 1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes Variable Angle Positioning Pins. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Variable Angle Positioning Pins in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for DePuy Synthes. The logo consists of a geometric shape on the left and the text "DePuy Synthes" on the right. The geometric shape is a sphere made up of several smaller shapes. The text is in a sans-serif font and is colored blue.
# 510(k) Summary for K953777:
Date Prepared: June 20, 2018
### 1.1. Submitter
Primary Contact:
Suchitra Basu, PhD, RAC Global Strategy Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5448 sbasu23@its.jnj.com
### Alternate Contact:
Elizabeth Jacobs Senior Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5768 Ejacob10@its.jnj.com
### 1.2. Device
Name of Device: DePuy Synthes Wire Mount – MR Conditional Classification Name(s): Cerclage, Fixation Common or Usual Name(s): Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II; 888.3010 Product Code(s): JDQ
1.3. Predicate Device Synthes Wire Mount (K953777)
# 1.3.1 Reference Device
DePuy Synthes Trauma Orthopedic Implant Portfolio – MR Conditional (K160553)
# 1.4. Device Description
The Wire Mount is intended for use with cerclage wire to augment fracture stabilization with plates when use of screws is contraindicated, as in the presence of intramedullary implants. The design is a "T" post with a hole perpendicular to the post axis, and a small triangular ridge on the bottom. Placing the post through the dynamic compression screw hole, inserting from the bone side of the plate, creates a stable structure for cerclage wire fixation. The cerclage wire is passed around the bone, through the wire mount hole above the outer surface of the plate, and then twisted with the opposite wire end.
# 1.5. Indications for Use
The DePuy Synthes Wire Mount is intended for use with cerclage wire to augment fracture
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for DePuy Synthes. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is a stylized representation of a sphere or globe, with intersecting lines forming a pattern. The company name, "DePuy Synthes," is written in a sans-serif font. The color of the geometric shape is a light brown or gold, while the company name is a dark blue.
stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants.
The Wire Mount is designed for use in dynamic compression screw holes that accept a 4.5 mm bone screw. The Wire Mount (and cerclage wire) can be used with a variety of plates that include, but are not limited to the following: Dynamic Hip Screw (DHS) Plates, Dynamic Condylar Screw (DCS) Plates, Condylar Buttress Plates, Cobra Head Plates, 4.5 mm Narrow and Broad Dynamic Compression Plates (DCP), 90° Child and Adolescent Osteotomy Plates, 110-130° Adult Osteotomy Plates, and 95° Condylar and 130° Angle Blade Plates (both including small stature plates).
# 1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the DePuy Synthes Wire Mount. The intended use and technological characteristics of the devices remain unchanged.
Non-clinical testing is provided to support the conditional safety of the DePuy Synthes Wire Mount construct in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for DePuy Synthes. The logo consists of a geometric shape on the left and the text "DePuy Synthes" on the right. The geometric shape is a sphere made up of several smaller shapes. The text is in a sans-serif font and is colored blue.
# 510(k) Summary for K992891:
Date Prepared: June 20, 2018
#### 1.1. Submitter
Primary Contact: Suchitra Basu, PhD, RAC Global Strategy Manager, Regulatory Affairs DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5448 sbasu23@its.jnj.com
Alternate Contact: Elizabeth Jacobs Senior Regulatory Affairs Specialist DePuy Synthes 1301 Goshen Parkway West Chester, PA 19380 Phone: (610) 719-5768 Ejacob10@its.jnj.com
#### 1.2. Device
Name of Device: DePuy Synthes Cerclage Positioning Pin- MR Conditional Classification Name(s): Cerclage, Fixation Common or Usual Name(s): Smooth or thread metallic bone fixation fastener Regulatory Class: Class II; 888.3010 Product Code(s): JDQ
#### 1.3. Predicate Device
Synthes Cerclage Positioning Pin (K992891)
### 1.3.1 Reference Device
DePuy Synthes Trauma Orthopedic Implant Portfolio – MR Conditional (K160553)
# 1.4. Device Description
The design of the cerclage positioning pin is an oval body with a hole running perpendicular to the long axis (for use in compression holes), and has a stud protruding from the bottom. By placing the post through the top of the dynamic compression screw hole into a pre-drilled hole in cortical bone a stable structure is created for cerclage wire or cable fixation. The wire or cable is passed around the bone, through the Cerclage Positioning Pin hole above the outer plate surface, and then the wire is twisted or cable is crimped for final securement.
#### 1.5. Indications for Use
{13}------------------------------------------------
The DePuy Synthes Cerclage Positioning Pins are intended for use with cerclage monofilament wire and multifilament cable to augment fracture stabilization with plates used in long bone fixation when the use of screws is contraindicated, as in the presence of intramedullary implants.
Cerclage Positioning Pins are designed for use with plates having locking compression or dynamic compression screw holes that accept a 3.5 mm or 4.5 mm bone screw. The Cerclage Positioning Pin (and cerclage wire or cable) can be used with a variety of Synthes plates.
#### 1.6. Substantial Equivalence
The purpose of this submission is to add MR Conditional information to the device labeling for the Synthes Cerclage Positioning Pin. The intended use and technological characteristics of the devices remains unchanged.
Non-clinical testing is provided to support the conditional safety of the Synthes Cerclage Positioning Pin construct in the MR environment including assessment of magnetically induced displacement force (ASTM F2052-14) and torque (ASTM F2213-06), radio frequency (RF) heating (ASTM F2182-11a), and image artifacts (ASTM F2119-07). The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.