← Product Code [JDO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO) · K193029

# CONQUEST FN (K193029)

_Smith & Nephew, Inc. · JDO · Feb 28, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO/K193029

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [JDO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO.md)
- **Decision Date:** Feb 28, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Smith & Nephew CONQUEST FN is intended for use in internal fixation of femoral neck fractures.

## Device Story

CONQUEST FN system consists of contoured locking bone plates and compatible locking/non-locking bone screws. Device used for internal fixation of femoral neck fractures. Plates are implanted by surgeons in clinical settings to stabilize bone fractures. System provides mechanical support to facilitate healing. Subject device adds smaller plate sizes to existing system. Manufactured from 316L stainless steel; provided sterile via Gamma irradiation. Single-use only.

## Clinical Evidence

Bench testing only. Construct fatigue testing performed to evaluate stability during bending fatigue; results met acceptance criteria. Bacterial endotoxin testing performed and met limits per ANSI/AAMI ST72.

## Technological Characteristics

Material: 316L stainless steel. Design: Contoured locking bone plates and locking/non-locking bone screws. Sterilization: Gamma irradiation. Single-use. No software or electronic components.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- CONQUEST FN ([K152686](/device/K152686.md))
- Smith & Nephew Cannulated Screws and Washers ([K111994](/device/K111994.md))

## Reference Devices

- POGO Screw design ([K080649](/device/K080649.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

February 28, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

K193029 Smith & Nephew, Inc. 1450 E Brooks Rd, Memphis, TN 38116 US Trade/Device Name: CONQUEST FN Contact Name: Thomas Fearnley Senior Regulatory Affairs Specialist

### Re: K193029

Trade/Device Name: CONQUEST FN Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDO, KTT Dated: January 24, 2020 Received: January 27, 2020

Dear Thomas Fearnley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Vesa Vuniqi, MS Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K193029

Device Name CONQUEST FN

Indications for Use (Describe)

The Smith & Nephew CONQUEST FN is indicated for displaced intracapsular femoral neck fractures.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995,

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the Smith & Nephew logo. The logo consists of an orange flower-like symbol on the left, followed by the text "We are smith&nephew" in black and orange. The words "We are" are in black, while "smith&nephew" is in orange.

| Submitted by:                                | Smith & Nephew, Inc.<br>Advance Surgical Devices Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Date of Summary:                             | October 29, 2019                                                                                               |
| Contact Person and Address:                  | Thomas Fearnley<br>Senior Regulatory Affairs Specialist<br>T (901) 399-1224<br>F (901) 566-7022                |
| Name of Device:                              | CONQUEST FN                                                                                                    |
| Common Name:                                 | Internal Fracture Fixation Device                                                                              |
| Device Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories<br>Class II     |
| Panel Code:                                  | Orthopedics/87                                                                                                 |
| Product Code:                                | JDO,KTT                                                                                                        |

### Device Description

The CONQUEST FN system is comprised of contoured locking bone plates and compatible locking and non-locking bone screws. The subject premarket notification describes additional sizes of the bone plates to the Smith & Nephew CONQUEST FN system. The subject devices are manufactured from the same implant-grade stainless steel (316L) and designed for singleuse. They will be provided in a sterile packaged option and will be sterilized via Gamma irradiation.

### Intended Use

The Smith & Nephew CONQUEST FN is intended for use in internal fixation of femoral neck fractures.

### Indications for Use

The Smith & Nephew CONQUEST FN is indicated for displaced and undisplaced intracapsular femoral neck fractures.

### Comparison to Technological Characteristics with Predicate Device

Device comparisons described in this premarket notification demonstrated that the proposed CONQUEST FN bone plates are substantially equivalent to the legally marketed predicate devices listed below with regard to intended use, indications for use, and performance characteristics.

The subject bone plates feature characteristics very similar to the CONQUEST FN system bone plates cleared via K152686, with the primary differences being the elimination of the plate shaft and distal screw holes. All other design aspects, indications for use, intended use, material and fundamental scientific technology remain the same as those in K152686.

{4}------------------------------------------------

## Substantial Equivalence Information

The substantial equivalence of the CONQUEST FN 3 Hole Small Stature Plate is based on its similarities in indications for use, design features, sterilization methods, and materials to the predicate systems listed in the following table.

| Manufacturer   | Description                                     | Submission<br>Number | Clearance<br>Date |
|----------------|-------------------------------------------------|----------------------|-------------------|
| Smith & Nephew | CONQUEST FN (Primary Predicate)                 | K152686              | 3/17/2016         |
| Smith & Nephew | Smith & Nephew Cannulated Screws and<br>Washers | K111994              | 10/11/2011        |

Table 5.1: Substantially Equivalent Predicates to CONQUEST FN 3 Hole Small Stature Plate

{5}------------------------------------------------

A comparison of the subject device to the predicate device is described in the following table. Table 5.2 Summary of Comparison to Predicate Devices

| Design Aspect<br>Reviewed                                   | CONQUEST FN 3<br>Hole Small Stature<br>Plate                                                                                     | CONQUEST FN<br>(primary predicate)                                                                                                | Cannulated Screws and<br>Washers                                                                                                                                              |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Type                                              | N/A                                                                                                                              | Primary                                                                                                                           | Reference                                                                                                                                                                     |
| 510(k) Number                                               | Subject 510(k)                                                                                                                   | K152686                                                                                                                           | K111994                                                                                                                                                                       |
| Same intended<br>use                                        | Yes                                                                                                                              | Yes                                                                                                                               | Yes                                                                                                                                                                           |
| Same<br>Indications for<br>Use                              | The Smith & Nephew<br>CONQUEST FN is<br>indicated for<br>displaced and<br>undisplaced<br>intracapsular femoral<br>neck fractures | The Smith & Nephew<br>CONQUEST FN is<br>indicated for<br>displaced and<br>undisplaced<br>intracapsular femoral<br>neck fractures  | The Smith & Nephew<br>Cannulated Screws and<br>Washers are used for<br>adult and pediatric<br>patients as indicated for<br>pelvic, small, and long<br>bone fracture fixation. |
| Similar<br>Operating<br>Principle                           | Yes                                                                                                                              | Yes                                                                                                                               | No                                                                                                                                                                            |
| Raw material<br>type                                        | 316L Stainless Steel                                                                                                             | 316L Stainless Steel                                                                                                              | 316L Stainless Steel                                                                                                                                                          |
| Similar design<br>features.<br>function and<br>intended use | Yes                                                                                                                              | Yes                                                                                                                               | No                                                                                                                                                                            |
| Anatomically<br>based Femoral<br>Locking Plate              | Yes                                                                                                                              | Yes                                                                                                                               | No                                                                                                                                                                            |
| Screws                                                      | 7.5 and 8.5mm<br>Locking Screws<br>(based on POGO<br>Screw design<br>K080649, up to 3)                                           | 7.5 and 8.5mm<br>Locking Screws<br>(based on POGO<br>Screw design<br>K080649, up to 3)<br>4.5mm locking and<br>non-locking screws | 4.0mm, 5.5mm,<br>6.5mm, 7.0mm and<br>8.0mm Cannulated<br>Screws                                                                                                               |

### Summary of Pre-Clinical Testing

- Construct Fatigue Testing was conducted on the CONQUEST FN 3 Hole Small Stature . Plate. Results of the test concluded that the CONQUEST FN 3 Hole Small Stature Plate met the acceptance criteria and is expected to have similar construct stability during bending fatigue as compared to the predicate.
Bacterial endotoxin testing was completed and met the acceptable endotoxin limits as stated in FDA Guidance , "Submission and Review of Sterility Information in Premarket Notification

{6}------------------------------------------------

(510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxin Testing: Questions and Answers," and ANSI/AAMI ST72.

### Conclusion

The proposed additional sizes of CONQUEST FN plates are substantially equivalent to the previously cleared CONQUEST FN plates cleared with the CONQUEST FN system in K152686 in that the indications for use and fundamental scientific technology for the subject devices are identical.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO/K193029](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO/K193029)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
