K032730 · Disc-O-Tech Medical Technologies, Ltd. · JDI · Mar 26, 2004 · Orthopedic
Device Facts
Record ID
K032730
Device Name
FIXION HIP SYSTEM
Applicant
Disc-O-Tech Medical Technologies, Ltd.
Product Code
JDI · Orthopedic
Decision Date
Mar 26, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Fixion® Hip System is intended for use as a hemi-hip or total hip replacement. The Fixion® Hip System is indicated as a unipolar hemi-hip replacement in cases of: - Femoral head and/or neck fractures or non-unions. - Aseptic necrosis of the femoral head and/or neck. - Osteo-, rheumatoid-, and/or post-traumatic arthritis of the hip, with minimal acetabular involvement. The Fixion® Hip System is indicated as a total hip replacement in cases of: - Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis. - Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. - Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results. The Fixion® Hip Stem is intended for cemented or non-cemented use.
Device Story
Fixion® Hip System is a modular hip prosthesis for hemi-hip or total hip replacement. System components include an expandable stainless steel femoral stem, head ball (stainless steel for hemi-hip, cobalt chromium for total hip), porous-coated cobalt chromium acetabular cup, UHMWPE liner, and cobalt chromium cancellous bone screws. The femoral stem is inserted into the femoral medullary canal and expanded in situ using a manual inflation device (pump). The device is used in clinical settings by surgeons. Expansion of the stem provides fixation within the femur. The system aims to restore joint function and alleviate pain in patients with disabling hip disease or fractures.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 7206-1, ISO 7206-4, ISO 7206-8, and FDA guidance documents for femoral stem prostheses and mechanically locked modular components. Materials testing conducted per ASTM F 138, ASTM F 1586, ISO 5832-1, and ISO 5832-9.
Technological Characteristics
Materials: Stainless steel (ASTM F 138, ISO 5832-1/9), cobalt chromium, UHMWPE. Design: Modular hip prosthesis with expandable femoral stem. Principle: Mechanical expansion of stem via manual pump. Connectivity: None. Sterilization: Not specified. Software: None.
Indications for Use
Indicated for patients requiring hemi-hip or total hip replacement due to femoral head/neck fractures, non-unions, aseptic necrosis, or degenerative, rheumatoid, or post-traumatic arthritis. Also indicated for revision of failed previous hip procedures or cases where alternative reconstructive techniques are less likely to succeed. Intended for cemented or non-cemented use.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Fixion® Unipolar Modular Hemi-Hip System (K014072, K030972)
Fixion® Interlocking Proximal Femoral Intramedullary Nailing System (K010988, K012967, K023437)
Submission Summary (Full Text)
{0}------------------------------------------------
#### 510(K) Summary MAR 2 6 2004
# Disc-O-Tech Medical Technologies Ltd. Fixion® Hip System
#### Company Name
Disc-O-Tech Medical Technologies Ltd.
3 Hasadnaot St.
Herzliya 46728, Israel
#### Submitter's Name and Contact Person
Hila Wachsler-Avrahami Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St. Herzliya 46728, Israel Tel: + 972-9-9511511 Fax: + 972-9-9548939
### Date Prepared
August 2003
#### Trade/Proprietary Name
Fixion® Hip System
#### Classification
Class II
{1}------------------------------------------------
#### Predicate Devices
公
Fixion® Unipolar Modular Hemi-Hip System (Fixion® MH, K014072, K030972) by Disc-O-Tech Medical Technologies Ltd.
- U1 Hip System (K994078) by United Orthopedic Corporation. ﺮ
- Elite® Modular Hip System (K871867) by Depuy Orthopeadics. ア
- S-ROM® Total Hip System (K851422, K913231, K954935, K961939) by Depuy A Orthopeadics.
- Link® MP® Hip System (K955296) by Link America Inc. 公
- Fixion® Interlocking Proximal Femoral Intramedullary Nailing System 公 (Fixion® PF, K010988, K012967, K023437) by Disc-O-Tech Medical Technologies Ltd.
#### Performance Standards
The following standards were referred:
- ン ISO 7206-1 (Draft, 1995): Implants for Surgery - Partial and Total Hip Joint Prostheses - Part 1: Classification and Description of Dimensions.
- ISO 7206-4 (Draft, 1999): Implants for Surgery Partial and Total Hip Joint 公 Prostheses - Part 4: Determination of Endurance Properties of Stemmed Femoral Components.
- ISO 7206-8 (1995): Implants for Surgery Partial and Total Hip Joint A Prostheses - Part 8: Endurance Performance of Stemmed Femoral Components with Application of Torsion.
- ア Guidance Document for for Femoral Stem Prostheses (Draft), ORDB/DGRD/CDRH/FDA, August 1, 1995.
- 公 Guidance Document for Testing Non-Articulating, "Mechanically Locked", Modular Implant Components (Draft), ORDB/DGRD/CDRH/FDA, May 1, 1995.
- ASTM F 138-2000: Standard Specification for Stainless Steel Bar and Wire ﺮ for Surgical Implants.
<sup>3 11</sup>asadnao! St., Herzliya 46728, Israel; Tel: 972-9511511; Fax: 972-9-9548939; F-mail: hila@disc-o-tech.com Page 19 of 23
{2}------------------------------------------------
-
- ASTM F 565-2000: Standard Practice for Care and Handling of تز Implants and Instruments.
- ASTM F 1586-95: Wrought Nitrogen Strengthened 21 Chromium 10 A Nickel - 3 Manganese - 2.5. Molybdenum Stainless Steel Bar for Surgical Implants.
- ISO 5832-1 (1997): Implants for Surgery Metallic Materials Part 1: ﺮ Wrought High Nitrogen Stainless Steel.
- ISO 5832-9 (1995): Implants for Surgery Metallic Materials Part 9: V Wrought High Nitrogen Stainless Steel.
- BS EN 12563: 1999 Non-active surgical implant Joint replacement ア implants - Specific requirements for hip joint replacement implants.
#### Intended Use
The Fixion® Hip System is intended for use as a hemi-hip or total hip replacement. The Fixion® Hip System is indicated as a unipolar hemi-hip replacement in cases of:
- ア Femoral head and/or neck fractures or non-unions.
- Aseptic necrosis of the femoral head and/or neck. ﺮ
- Osteo-, rheumatoid-, and/or post-traumatic arthritis of the hip, with V minimal acetabular involvement.
The Fixion® Hip System is indicated as a total hip replacement in cases of:
- Painful, disabling joint disease of the hip resulting from: degenerative A arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup ﺮ arthroplasty or other procedure.
- A management problems where arthrodesis or alternative Clinical reconstructive techniques are less likely to achieve satisfactory results.
The Fixion® Hip Stem is intended for cemented or non-cemented use.
<sup>3</sup> Fasadnaot St., Herzliya 46728, Jsrael, Tel: 972-9-9511511; Fax: 972-9-9548939; E-mail: hila@disc-o-tech.com Page 20 of 23
{3}------------------------------------------------
page 4 of 5
K032730
#### System Description
The Fixion® Hip System is composed of an implant, inflation device (pump) and an instrumentation set.
- Implant comprises femoral stem and head ball, and for total hip >> replacement - acetabular cup and liner and cancellous bone screws, as detailed below:
- A stainless steel Stem, including stem body and neck, which is inserted . into the femoral medullary canal and expanded following insertion.
- A Head Ball, assembled on the stem neck and articulates within the . acetabulum (in hemi-hip replacement) or acetabular cup and liner (in total hip replacement). A unipolar stainless steel head (for hemi- hip replacement), and a cobalt chromium head (for total hip replacement) are available.
- A porous coated, cobalt chromium Acetabular Cup, inserted and . fixated within the acetabulum.
- A UHMWPE Acetabular Cup Liner, which is seated inside the . acetabular cup.
- Cobalt chromium Cancellous Bone Screws, intended to fixate the . acetabular cup to the acetabulum, in cases additional stabilization of the cup is required.
- ﯩﺰ Instrumentation Set - a set of accessories to be used with the Fixion® Hip implants.
- > Inflation Device (Pump) - a manual pump used to expand the Fixion® stem.
#### Substantial Equivalence
Hip System intended use, design, materials, technological The Fixion® characteristics and principles of operation are substantially equivalent to those of Disc-O-Tech's Fixion® Unipolar Modular Hemi-Hip System (K014072, K030972), Fixion® Interlocking PF Intramedullary Nailing System (K010988, K012967,
<sup>3</sup> Hasadnaot St., Herzliya 46728, Israel; Tel: 972-9511511; Fax: 972-9-9548939; E-mail: hila@disc-o-tech.com; www.disc-o-tech.com Page 21 of 23
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the text "page 5 of 5" and "k032730" written in black ink. The text "page 5 of 5" is underlined. The text appears to be handwritten.
K023437), United Orthopedic Corporation' U1 Hip System (K994078 Orthopaedics' Elite® Modular Hip System (K871867) and S-ROM® Total Hip System (K851422, K913231, K954935, K961939 and others), and the Link America Inc.' Link® MP® Hip System (K955296).
3 Hasadnavt St., Herzliya 46728, Israel; Tel: 972-9-9511511; Fax: 972-9-9548939; E-mail: hila@disc-o-tech.com; www.disco-tech.com Page 22 of 23
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
==============================================================================================================================================================================
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with three stripes on its wings, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 2 6 2004
Ms. Hila Wachsler-Avrahami Regulatory Affairs Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot Street Herzliya 46728, Israel
Re: K032730 Trade/Device Name: Fixion® Hip System Regulation Numbers: 21 CFR 888.3350; 21 CFR 888.3358; 21 CFR 888.3360 Regulation Names: Hip joint metal/polymer semi-constrained cemented prosthesis; Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis; Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis. Regulatory Class: II
Product Codes: JDI, LPH, MBL, KWL, JDG, LWJ Dated: December 25, 2003 Received: December 29, 2003
Dear Ms. Wachsler-Avrahami:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Mark N Milliman
'elia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indication for Use Mark N. Milliner
Division Sign Off:
510(K) Number (if known): K032730
Device Name: Fixion® Hip System
#### Indication for Use:
Division of General, Restorative,
K032730
510(a) Number
The Fixion® Hip System is intended for use as a hemi-hip or total hip replacement.
The Fixion® Hip System is indicated as a unipolar hemi-hip replacement in cases of:
- A Femoral head and/or neck fractures or non-unions.
- A Aseptic necrosis of the femoral head and/or neck.
- A Osteo-, rheumatoid-, and/or post-traumatic arthritis of the hip, with minimal acetabular involvement.
The Fixion® Hip System is indicated as a total hip replacement in cases of:
- A Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup A arthroplasty or other procedure.
- A Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
The Fixion® Hip Stem is intended for cemented or non-cemented use.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) (Part 21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
<sup>3</sup> Hasadnaot St., Herzliya 46728, Israel; Tel: 972-9-9511511; Fax: 972-9-9548939; E-mail: hila@disc-o-tech.com Page 16 of 23
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.