K042664 · Depuy, Inc. · JDC · Dec 8, 2004 · Orthopedic
Device Facts
Record ID
K042664
Device Name
LPS UPPER EXTREMITY
Applicant
Depuy, Inc.
Product Code
JDC · Orthopedic
Decision Date
Dec 8, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LPS Upper Extremity is intended for use in replacement of the proximal, mid-shaft or intercalary portion, distal and/or total humerus. This system is especially designed for cases that require extensive resection and restoration. Specific diagnostic indications for use include: - Primary bone neoplasnms (e.g., osteosarcomas, chondrosarcomas, Ewing's sarcomas) requiring . extensive resection(s) and replacement(s) of the proximal and/or distal humerus; - Metastatic bone disease and pathologic fractures with extensive bone loss or where other forms . of treatment such as internal fixation are inadequate; - Patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular . necrosis, osteoarthritis, and inflammatory joint disease (JD), e.g., rheumatoid arthritis, requiring extensive resection and replacement of the proximal and/or distal humerus; - Patients suffering from congential or acquired deformity, such as post-traumatic deformity . and/or arthritis; - Communited fractures of the proximal, distal and/or humeral shaft where forms of fixation . are indadequate; - . Persistent humeral fracture non-union; - Patients suffering from severe cuff tear arthropathy that does not respond to any conservative . therapy or better alternative surgical treatment; - Revision shoulder or elbow arthroplasty cases requiring extensive resection(s) and . replacements of the proximal, distal or total humerus. - Severe trauma requiring extensive resection and replacement . The LPS Upper Extremity components are for CEMENTED USE ONLY.
Device Story
LPS Upper Extremity is a modular shoulder and elbow prosthesis system designed for oncology and complex revision cases requiring extensive humeral resection. System replaces proximal, mid-shaft, distal, or total humerus. Components are implanted via cement fixation. Device provides structural replacement for bone loss where standard fixation is inadequate. Used by orthopedic surgeons in clinical/surgical settings. Benefits include restoration of limb function and structural integrity in patients with severe bone loss or disease.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design and material similarity to legally marketed predicate devices.
Technological Characteristics
Modular shoulder and elbow prosthesis system; metal/polymer construction; cemented fixation; designed for extensive humeral resection and replacement.
Indications for Use
Indicated for patients requiring extensive resection and replacement of proximal, distal, or total humerus due to primary bone neoplasms, metastatic bone disease, pathologic fractures, NIDJD (avascular necrosis, osteoarthritis), inflammatory joint disease (rheumatoid arthritis), congenital/acquired deformity, comminuted fractures, persistent non-union, severe cuff tear arthropathy, or revision arthroplasty.
Regulatory Classification
Identification
An elbow joint metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace an elbow joint. It is made of alloys, such as cobalt-chromium-molybdenum, or of these alloys and of an ultra-high molecular weight polyethylene bushing. The device prevents dislocation in more than one anatomic plane and consists of two components that are linked together. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,”
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra High Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,”
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(viii) ISO 14630:1997 “Non-active Surgical Implants—General Requirements,”
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 981-93 “Practice for Assessment of Compatibility of Biomaterials (Nonporous) for Surgical Implant with Respect to Effect of Material on Muscle and Bone,”
(v) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(vi) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vii) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings, ” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
{0}------------------------------------------------
DEC - 8 2004
# 510(k) Summary LPS Upper Extremity
DePuy, Inc. 700 Orthopaedic Drive Warsaw, IN 46581
## A. Contact Person:
Tiffani Rogers Regulatory Affairs Associate (574) 371-4927
## B. Device Information:
| Proprietary Name: | LPS Upper Extremity |
|----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Shoulder prosthesis; Elbow prosthesis |
| Classification Name and<br>Regulatory Class: | Prosthesis, shoulder, non-constrained, metal/polymer<br>cemented: Class II per 21 CFR §888.3650<br>Elbow joint metal/polymer constrained cemented<br>prosthesis: Class II per 21 CFR §888.3150 |
87 KWT, 87 JDC Product Code:
### C. Indications for Use:
The LPS Upper Extremity is intended for use in replacement of the proximal, mid-shaft or intercalary portion, distal and/or total humerus. This system is especially designed for cases that require extensive resection and restoration. Specific diagnostic indications for use include:
- Primary bone neoplasnms (e.g., osteosarcomas, chondrosarcomas, Ewing's sarcomas) requiring . extensive resection(s) and replacement(s) of the proximal and/or distal humerus;
- Metastatic bone disease and pathologic fractures with extensive bone loss or where other forms . of treatment such as internal fixation are inadequate;
- Patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular . necrosis, osteoarthritis, and inflammatory joint disease (JD), e.g., rheumatoid arthritis, requiring extensive resection and replacement of the proximal and/or distal humerus;
- Patients suffering from congential or acquired deformity, such as post-traumatic deformity . and/or arthritis;
- Communited fractures of the proximal, distal and/or humeral shaft where forms of fixation . are indadequate;
- . Persistent humeral fracture non-union;
- Patients suffering from severe cuff tear arthropathy that does not respond to any conservative . therapy or better alternative surgical treatment;
- Revision shoulder or elbow arthroplasty cases requiring extensive resection(s) and . replacements of the proximal, distal or total humerus.
{1}------------------------------------------------
- Severe trauma requiring extensive resection and replacement .
The LPS Upper Extremity components are for CEMENTED USE ONLY.
#### D. Device Description:
The LPS Upper Extremity components are designed for the replacement of the mid-shaft or intercalary portion of the humerus, proximal, distal and/or total humerus. Unlike primary shoulder and elbow systems, this system is used when the amount of resection and restoration is extreme (e.g. in oncology cases, end-stage revision). A total humeral replacement is possible in those cases where no part of the humerus can be salvaged.
#### E. Substantial Equivalence:
The substantial equivalence of the LPS Upper Extremity is substantiated by its similarity in indications for use, design, materials, sterilization and packaging to the Global Advantage Total Shoulder system (K992065, cleared July 12, 1999), the Biomet Orthopedic 3-piece Proximal Humeral Replacement System (K020045, cleared January 30, 2002), the Biomet Orthopaedic Discovery Elbow - Mosaic Distal Humeral Replacement System (K033280, cleared December 19, 2003) and the Acclaim Elbow, cleared as the DePuy Total Elbow (K992656, cleared November 05, 1999).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC - 8 2004
Ms. Tiffani D. Rogers Regulatory Affairs Associate DePuy Orthopaedics, Inc. P.O. Box 988 700 Orthopaedic Drive Warsaw, Indiana 46581-0988
Re: K042664
Trade/Device Name: LPS Upper Extremity Regulation Numbers: 21 CFR 888.3150, 21 CFR 888.3650 Regulation Names: Elbow joint metal/polymer constrained cemented prosthesis, Shoulder joint metal/polymer non-constrained cemented prosthesis Regulatory Class: II Product Codes: JDC, KWT Dated: September 23, 2004 Received: September 28, 2004
Dear Ms. Rogers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
## Page 2 - Ms. Tiffani D. Rodgers
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use 510(k) Number (if known): _ KO 42664 Device Name: LPS Upper Extremity
Indications for Use:
The LPS Upper Extremity is intended for use in replacement of the proximal, mid-shaft or intercalary portion, distal and/or total humerus. This system is especially designed for cases that require extensive resection and restoration. Specific diagnostic indications for use include:
- Primary bone neoplasnms (e.g., osteosarcomas, chondrosarcomas, Ewing's sarcomas) . requiring extensive resection(s) and replacement(s) of the proximal and/or distal humerus;
- Metastatic bone disease and pathologic fractures with extensive bone loss or where other . forms of treatment such as internal fixation are inadequate;
Patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular . necrosis, osteoarthritis, and inflammatory joint disease (JD), e.g., rheumatoid arthritis, requiring extensive resection and replacement of the proximal and/or distal humerus;
Patients suffering from congential or acquired deformity, such as post-traumatic deformity . and/or arthritis;
Communited fractures of the proximal, distal and/or humeral shaft where forms of fixation . are indadequate;
Persistent humeral fracture non-union; �
Patients suffering from severe cuff tear arthropathy that does not respond to any conservative . therapy or better alternative surgical treatment;
Revision shoulder or elbow arthroplasty cases requiring extensive resection(s) and . replacements of the proximal, distal or total humerus.
Severe trauma requiring extensive resection and replacement .
The LPS Upper Extremity components are for CEMENTED USE ONLY.
| Prescription Use | X (Part 21 CFR 801 Subpart D) | AND/OR | Over-The-Counter Use | (21 CFR 807 Subpart C) |
|------------------|-------------------------------|--------|----------------------|------------------------|
|------------------|-------------------------------|--------|----------------------|------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of GDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
| Page | of |
|---------------|---------|
| 510(k) Number | K042664 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.