← Product Code [HWC](/productcode/HWC) · K994152

# THE LIGAMENT ANCHOR SOFT SCREW (K994152)

_Corin USA · HWC · Mar 6, 2000 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K994152

## Device Facts

- **Applicant:** Corin USA
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Mar 6, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The screws are used in anterior cruciate ligament and posterior cruciate ligament reconstruction using hamstrings. This device may also be used for bone patellar tendon bone fixation. The soft screw provides solid fixation of the ligament without damaging grafts or the need for pre-tapping, thus avoiding damage to the hamstring grafts. The ligament anchor soft screw can be used in both the femur and tibia or in conjunction with the Corifix Ligament Anchor which is already 510k approved.

## Device Story

Corifix Ligament Anchor Soft Screw is a headless, tapered, cannulated, threaded screw; manufactured from titanium alloy. Used in orthopedic surgery for anterior and posterior cruciate ligament reconstruction; provides fixation of hamstring tendon grafts to bone. Device features rounded threads to prevent graft damage; eliminates need for pre-tapping. Implanted by surgeons in clinical settings. Fixation strength tested to 691N, exceeding clinical loads of 150N (walking) and 450N (jogging). Benefits include reduced soft-tissue morbidity, reduced post-operative pain on kneeling, and improved cosmetic outcomes.

## Clinical Evidence

Bench testing only. Mechanical testing determined maximum clinical loads of 150N (walking) and 450N (jogging). Fixation strength of the device was measured at 691N.

## Technological Characteristics

Material: Titanium alloy 6Al4V (BS 7252/3, ASTM F136-96). Geometry: Headless, tapered, cannulated, rounded threads. Drive: 3.5mm hex. Sizes: 7mm, 9mm, 10mm diameters; 20mm, 25mm, 30mm, 35mm lengths. Energy source: None (mechanical).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Smith & Nephew RCI Screw ([K947134](/device/K947134.md))

## Reference Devices

- Corifix Ligament Anchor

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### 510(k) SUMMARY MAR - 6 2000

Corifix Name of Company:

The Corinium Centre Cirencester Gloucestershire GL7 1YJ England

Name of Device:

Ligament Anchor Soft Screw .

Device Description:

The Corifix ligament Anchor is a headless, tapered, round threaded, cannulated screw manufactured from titanium alloy (6A1 4V) which conforms to BS 7252/3 and ASTM 136/96. The screw has a 3.5mm hex drive.

The Ligament Anchor Soft Screw has been on the foreign (i.e. non-US) market since January 1997, since that date approximately 1,700 screws have been implanted.

## Intended Use.

The screws are used in anterior cruciate ligament and posterior cruciate ligament reconstruction using hamstrings. This device may also be used for bone patellar tendon bone fixation.

The soft screw provides solid fixation of the ligament without damaging grafts or the need for pre-tapping, thus avoiding damage to the hamstring grafts.

The ligament anchor soft screw can be used in both the femur and tibia or in conjunction with the Corifix Ligament Anchor which is already 510k approved. (

## Material.

The material used for the Ligament Anchor Soft Screw is Titanium Alloy 6A14V, certified to BS 7252:Pt. 3:1997 Metallic materials for surgical implants - Specification for wrought titanium 6-Aluminium 4-Vanadium alloy and ASTM F136-96 Wrought Titanium-6Aluminium-4Vanadium ELI.

## Device Geometry.

The Corifix Ligament Anchor Soft Screw has a rounded thread to prevent damage to the graft, is cannulated and tapered to aid insertion, and has a 3.5mm hex drive.

## Sizes

The Corifix Ligament Anchor Soft Screw is available in four diameters i.e 7mm, 9mm & 10 mm, for each diameter of screw four lengths of screw are available i.e. 20mm, 25mm, 30mm, & 35 mm.

{1}------------------------------------------------

#### Summary of Mechanical Testing (Further details provided on Page 14 of this dossier).

The titanium alloy 6A14V from which the Corifix Ligament Anchor Soft Screw is manufactured is tested and certified to BS 7252:Pt. 3:1997 Metallic Materials for Surgical Implant -Specification for wrought titanium 6-Aluminium 4-Vanadium alloy and ASTM F136-96 Wrought Titanium-6Aluminium-4Vanadium ELI.

Independent mechanical testing of the most popular sizes of screw determined that the maximum expected clinical loads applied to this device are 150N during walking and 450N during jogging. The fixation strength of the Corifix Ligament Anchor Soft Screw was found to be 691N.

#### Device to which Substantial Equivalence is Claimed and Device Specific Features

Substantial equivalence for the Corifix Ligament Anchor Soft Screw is claimed with the Smith & Nephew RCI Screw, which has been in clinical use within the USA for 5 years. (Predicate Device No. K947134).

The comparable features of the RCI Screw & Corifix Ligament Anchor Soft Screw are that both screws have a rounded thread, are cannulated, made of titanium alloy 6AI4V, have a 3.5mm hex. drive and are tapered to aid insertion. The advantage of the Corifix Ligament Anchor Soft Screw is that it is available in a range of sizes.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human figure, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 6 2000

Mr. Craig Corrance President Corin U.S.A. 10500 University Center Drive Suite 130 Tampa, Florida 33612

Re: K994152

Trade Name: Corafix Ligament Anchor Soft Screw Regulatory Class: II Product Code: HWC Dated: November 18, 1999 Received: December 8, 1999

Dear Mr. Corrance:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2 - Mr. Craig Corrance

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Russell Jager

& James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 510k Number (of known): Prov. - K994152

# Device Name :- Corifix Ligament Anchor Soft Screw.

### INDICATIONS FOR USE

The Corifix Ligament Anchor Soft Screw is used for anterior and posterior cruciate ligament reconstruction using hamstrings.

This technique provides reduced soft-tissue morbidity and reduced post operative pain on kneeling whilst improving the cosmetic appearance for the patient.

The hamstring surgical technique and the Corifix Ligament Anchor Soft Screw work together for fixation of a hamstring tendon graft, without bone block, to achieve anatomic aperture fixation of the hamstring tendon graft in the tibia or femur.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE).

Russell Rogers

e

Prescription Use (Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/device/K994152](https://fda.innolitics.com/device/K994152)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
