ALTA FULLY THREADED CROSS-LOCKING SCREWS
Device Facts
| Record ID | K963398 |
|---|---|
| Device Name | ALTA FULLY THREADED CROSS-LOCKING SCREWS |
| Applicant | Howmedica Corp. |
| Product Code | HWC · Orthopedic |
| Decision Date | Oct 10, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Alta® Fully Threaded Cross-Locking Screws are intended to be used in conjunction with the Alta® Femoral and Alta® Tibial/Humeral Intramedullary Rod Systems. Specifically, the screws are used to lock the IM rods in position after placement in the femoral canal.
Device Story
Alta® Fully Threaded Cross-Locking Screws; titanium alloy orthopedic fixation hardware. Used as accessory to Alta® Femoral and Alta® Tibial/Humeral Intramedullary Rod Systems. Function: lock IM rods in position within femoral canal post-placement. Operated by orthopedic surgeons in clinical/OR settings. Benefit: provides mechanical stability to intramedullary fixation constructs.
Clinical Evidence
Bench testing only.
Technological Characteristics
Material: Titanium alloy. Diameters: 5.0mm and 3.7mm. Lengths: 25mm to 100mm. Fully threaded design. Mechanical fixation component.
Indications for Use
Indicated for use with Alta® Femoral and Alta® Tibial/Humeral Intramedullary Rod Systems to lock IM rods in position within the femoral canal.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Alta® Femoral Locking IM Rod System (K850441, K922266)
- Alta® CFx Reconstruction Rod System (K926232, K935295, K960524)
- Alta® Tibial/Humeral IM Rod System (K884500, K890936, K954554)
- Biomet Uniflex Tibial Nail System
- Biomet Retrograde Femoral Nail System
- Zimmer ZMS Intramedullary Recon Nail
Related Devices
- K972108 — ALTA INTRAMEDULLARY ROD SYSTEM · Howmedica Corp. · Aug 12, 1997
- K972811 — HOWMEDICA FULLY THREADED SCREWS · Howmedica Corp. · Oct 6, 1997
- K970549 — TITANIUM ALLOY BONE SCREWS · Aesculap, Inc. · Apr 30, 1997
- K974720 — HOWMEDICA(R) FULLY THREADED SCREWS · Howmedica, Inc. · Mar 13, 1998
- K993004 — ALTA FULLY THREADED CROSS-LOCKING 5.0 MM SCREW · Howmedica Osteonics Corp. · Oct 6, 1999