SYNTHES STERILE 3.5 MM AND 4.0 MM CANNULATED SCREWS
Device Facts
| Record ID | K963192 |
|---|---|
| Device Name | SYNTHES STERILE 3.5 MM AND 4.0 MM CANNULATED SCREWS |
| Applicant | Synthes (Usa) |
| Product Code | HWC · Orthopedic |
| Decision Date | Nov 8, 1996 |
| Decision | SN |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
Synthes 3.5 mm and 4.0 mm Cannulated Screws are intended for fracture fixation of small bones and small bone fragments.
Device Story
Cannulated screws (3.5 mm and 4.0 mm) designed for small bone fracture fixation; used with guide wire for precise placement. 3.5 mm version features reverse cutting flutes and cortical threads; 4.0 mm version features self-drilling/self-tapping tips, reverse cutting flutes, and cancellous threads. Provided sterile via gamma radiation. Used by surgeons in clinical settings to stabilize bone fragments, facilitating healing.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Materials: Not specified. Dimensions: 3.5 mm and 4.0 mm diameters. Features: Cannulated design for guide wire, reverse cutting flutes, cortical/cancellous threads, self-drilling/self-tapping tips (4.0 mm). Sterilization: Gamma radiation.
Indications for Use
Indicated for fracture fixation of small bones and small bone fragments in patients requiring internal fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Synthes 2.7 mm Cortex Screws (K963192)
Reference Devices
- Richards 4.0 mm Cannulated Screw
- Alphatec Titanium Cannulated Mini Lag Screw System