EXTREMITY BONE SCREW
Device Facts
| Record ID | K963118 |
|---|---|
| Device Name | EXTREMITY BONE SCREW |
| Applicant | Acu Med, Inc. |
| Product Code | HWC · Orthopedic |
| Decision Date | Oct 10, 1996 |
| Decision | SN |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The 2.7mm Extremity Bone Screw is intended for fracture and osteotomy fixation of the upper and lower extremities. This screw is not intended for usage in the spine and is not for use with any available bone plates, washers, etc.
Device Story
2.7mm titanium alloy bone screw; used for fracture and osteotomy fixation in upper and lower extremities; driven by 2.5mm hex head; provided non-sterile; requires steam sterilization by user; intended for surgical use by physicians; provides mechanical fixation of bone fragments.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Material: Titanium alloy per ASTM F 136. Dimensions: 2.7mm diameter; lengths 12-24mm. Drive: 2.5mm hex head per ASTM F 543. Sterilization: Steam sterilization (SAL 10⁻⁶).
Indications for Use
Indicated for fracture and osteotomy fixation of upper and lower extremities. Contraindicated for use in the spine and use with bone plates or washers.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Synthes' 3mm Cannulated Screw System
Related Devices
- K013954 — LORENZ SELF DRILLING SCREWS · Walter Lorenz Surgical, Inc. · Feb 28, 2002
- K111230 — SYNTHES 3.5MM LOW PROFILE CORTICAL SCREW · Synthes (USA) Products, LLC · May 24, 2011
- K961498 — RESCH ARTHROSCOPIC AND PERCUTANEOUS SCREW FIXATION SYSTEM (A.S.F.) · Howmedica, Inc. · Feb 6, 1997
- K102343 — EMERGE MEDICAL CANNULATED SCREW FIXATION SYSTEM · Emerge Medical · Nov 23, 2010
- K970549 — TITANIUM ALLOY BONE SCREWS · Aesculap, Inc. · Apr 30, 1997