K961649 · Smith & Nephew Endoscopy, Inc. · HWC · Jul 16, 1996 · Orthopedic
Device Facts
Record ID
K961649
Device Name
ACUFEX TIBIAL ANCHOR SCREW & SPIKED WASHER
Applicant
Smith & Nephew Endoscopy, Inc.
Product Code
HWC · Orthopedic
Decision Date
Jul 16, 1996
Decision
SN
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Acufex® Tibial Anchor Screw & Spiked Washer is indicated for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures.
Device Story
Device consists of two-part tibial anchor screw and spiked washer; used for soft tissue-to-bone fixation in orthopedic procedures. Outer screw (9mm diameter, 20-30mm length) features cancellous threads for bone contact; inner screw (7mm diameter, 20-25mm length) secures spiked washer (14-20mm diameter) to outer screw. Spikes at washer periphery provide tissue grip. Device implanted by orthopedic surgeons in clinical/OR settings to secure tendons/ligaments. Provides mechanical fixation to facilitate healing. Materials include titanium and reinforced polyacetal polymer.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Materials: Titanium and reinforced polyacetal polymer. Design: Two-part screw system (outer 9mm, inner 7mm) with spiked washer (14-20mm). Mechanical fixation principle. Sterile, single-use.
Indications for Use
Indicated for fixation of soft tissue, including tendons and ligaments, to bone during orthopedic surgical procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
K962194 — MLI SOFT TISSUE SCREW & WASHER SYSTEM · Medicinelodge, Inc. · Jul 30, 1996
K012572 — SOFT TISSUE SCREW AND WASHER · Biomet, Inc. · Dec 3, 2001
K983906 — DEPUY ORTHOTECH SPIKED WASHER · DePuy Orthopaedics, Inc. · Dec 18, 1998
K083607 — APERFIX IMPLANT WITH INSERTER, 8, 9, 10, 11, MM X30 MM · Cayenne Medical, Inc. · Mar 4, 2009
K042460 — TI - SCREW ANCHOR SP · Arthrotek, Inc. · Oct 29, 2004
Submission Summary (Full Text)
{0}
JUL 16 1996
K 96/649
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
The assigned 510(k) number is :
This summary was prepared on April 26, 1996
## A. Submitter
Smith & Nephew Endoscopy
130 Forbes Boulevard
Mansfield, MA 02048
## B. Company Contact
Tedd Gosian
Clinical Regulatory Manager
## C. Device Name
Trade Name: Acufex® Tibial Anchor Screw & Spiked Washer
Common Name: Fixation Screw & Washer
Classification Name: Single/Multiple component metallic bone fixation appliances and accessories
## D. Predicate/Legally Marketed Devices
Concept® Soft Tissue Anchoring Implants (K871037)
Synthes® Spiked Washer
## E. Device Description
Acufex® Tibial Anchor Screw & Spiked Washer is a device used in soft tissue to bone fixation. The Tibial Anchor Screw is a two part screw. The outer larger part has cancellous threads, which come in direct contact with bone, and is 9 mm in diameter with a length of 20 - 30 mm. The inner part, 7 mm diameter and 20 - 25 mm length, is used to attach the Spiked Washer directly to the outer part. The spiked washer is 14 - 20 mm in diameter with spikes set at the periphery.
## F. Intended Use
The Acufex® Tibial Anchor Screw & Spiked Washer is indicated for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures.
{1}
104
# G. Substantial Equivalence
The AcufeX® Tibial Anchor Screw & Spiked Washer is intended for use for fixation of soft tissue, such as tendon and ligaments, to bone in orthopedic procedures. The AcufeX® Tibial Anchor Screw & Spiked Washer is similar in design, function, materials and intended use to other devices currently marketed and in commercial distribution, namely Concept® Soft Tissue Anchoring Implants (K871037) and Synthes® Spiked Washer.
Risks to health have been addressed through the specified materials, processing controls, quality assurance, and compliance to the Medical Device Good Manufacturing Practices regulations.
A summary comparison of the characteristics of the AcufeX® Tibial Anchor Screw & Spiked Washer and the substantially equivalent device is presented in the table below.
| | Current Product | Substantially Equivalent Product | |
| --- | --- | --- | --- |
| Attribute I | AcufeX® Tibial Anchor Screw & Spiked Washer | Concept® Soft Tissue Anchoring Implants | Synthes® Spiked Washer |
| Indication | Soft Tissue Fixation | Soft Tissue Fixation | Ligament Fixation |
| Length | Outer: 20 - 30 mm
Inner: 18 - 25 | Screw: 25 - 55 mm | |
| Diameter | Outer: 9 mm.
Inner: 7 mm
Washer: 14 - 20 mm O.D. | Screw: 6.5 mm -
Washer: 14 to 20 mm O.D. | Washer: 13.5 mm O.D. |
| Materials | Screw: Titanium
Washer: Titanium
Reinforced Polyacetal Polymer | Screw: Titanium
Washer: Titanium | Washer: Stainless Steel reinforced polyacetal |
| Labeling | Sterile, Single Use Only | Sterile, Single Use Only | Sterile, Single Use |
Applicant _Vicki R. Brown_ Date _4/26/96_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.