← Product Code [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC) · K230867

# Bone Bolt System (K230867)

_University of Utah, Department of Orthopaedics · HWC · Jun 26, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K230867

## Device Facts

- **Applicant:** University of Utah, Department of Orthopaedics
- **Product Code:** [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC.md)
- **Decision Date:** Jun 26, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

The 3.5mm, 4.0mm and 4.5mm Bone Bolt System is indicated for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bone fragments, long bones, and long bone fragments of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia, calcaneus, femur, and fibula, and the bones of the hand and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by Bone Bolt fixation. The 5.0mm, 6.0mm and 7.0mm Bone Bolt System is indicated for fixation of fractures, fusions, osteotomies, non-unions, and malunions of long bones and long bone fragments of humerus, radius, ulna, tibia, femur and fibula, pelvis, sacrum, and the bones of the foot in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by Bone Bolt fixation.

## Device Story

Bone Bolt System comprises cannulated bolts, nut-sleeves, and optional washers; implanted from opposing bone surfaces to stabilize bone fragments; used for fracture fixation, arthrodeses, and osteotomies. Cannulated design allows placement over guidewires for precision. Used in clinical settings by surgeons; provides mechanical stabilization to facilitate bone healing. Benefits include secure fixation of bone fragments.

## Clinical Evidence

No clinical or animal testing conducted. Performance characterized via bench testing in accordance with ASTM F543-17 and ASTM F1264-16 Annex A4, including torsional strength, driving torque, axial pull-out strength, and static/dynamic bending testing.

## Technological Characteristics

Implants: Ti-6Al-4V (ASTM F136). Accessories: stainless steel (ASTM F899), polypropylene, silicone. Cannulated design for guidewire placement. Mechanical fixation via bolts and nut-sleeves.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- DePuy Synthes Cortex and Cannulated Screws ([K161616](/device/K161616.md))
- Synthes 3.5mm Cortex Screws ([K043185](/device/K043185.md))

## Reference Devices

- Acumed Frag-Loc® System ([K130986](/device/K130986.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

June 26, 2023

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

University of Utah, Department of Orthopaedics % Hollace Rhodes Vice President MCRA, LLC 803 7th Street NW, 3rd Floor Washington, District of Columbia 20001

Re: K230867

Trade/Device Name: Bone Bolt System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC. HTN Dated: March 29, 2023 Received: March 29, 2023

Dear Hollace Rhodes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali-S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K230867

Device Name Bone Bolt System

# Indications for Use (Describe)

The 3.5mm, 4.0mm and 4.5mm Bone Bolt System is indicated for fixation of fractures, fusions, osteotomies, non-unions, and malunions of small bone fragments, long bones, and long bone fragments of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia, calcaneus, femur, and fibula, and the bones of the hand and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by Bone Bolt fixation.

The 5.0mm, 6.0mm and 7.0mm Bone Bolt System is indicated for fixation of fractures, fusions, osteotomies, non-unions, and malunions of long bones and long bone fragments of humerus, radius, ulna, tibia, femur and fibula, pelvis, sacrum, and the bones of the foot in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by Bone Bolt fixation.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for The University of Utah. On the left side of the logo is a large red block U. To the right of the U is the text "THE UNIVERSITY OF UTAH" stacked on top of each other.

# 510(k) Summary

| Device Trade Name: | Bone Bolt System                                                                             |
|--------------------|----------------------------------------------------------------------------------------------|
| Sponsor:           | University of Utah, Department of Orthopaedics<br>590 Wakara Way<br>Salt Lake City, UT 84108 |
| Contact:           | T. Wade Fallin, M.S.<br>Research Professor<br>Phone: (801) 587-2938<br>Fax: (801) 587-5411   |
| Prepared by:       | MCRA, LLC                                                                                    |
| Date Prepared:     | June 26, 2023                                                                                |
| Common Names:      | Screw, Fixation, Bone<br>Washer, Bolt Nut                                                    |
| Classifications:   | 21 CFR 888.3040<br>21 CFR 888.3030                                                           |
| Class:             | II                                                                                           |
| Product Codes:     | HWC<br>HTN                                                                                   |

#### Indications for Use:

The 3.5mm, 4.0mm and 4.5mm Bone Bolt System is indicated for fixation of fractures, fusions, osteotomies, nonunions, and malunions of small bone fragments, long bones, and long bone fragments of clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, tibia, calcaneus, femur, and fibula, and the bones of the hand and foot in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by Bone Bolt fixation.

The 5.0mm, 6.0mm and 7.0mm Bone Bolt System is indicated for fixation of fractures, fusions, osteotomies, nonunions, and malunions of long bone fragments of humerus, radius, ulna, tibia, femur and fibula, pelvis, sacrum, and the bones of the foot in adults and in both children (2-12 years) and adolescents (12-21 years) in which growth plates have fused or in which growth plates will not be crossed by Bone Bolt fixation.

#### Device Description:

The Bone Bolt System consists of various sizes and lengths of bolts and nut-sleeves, which are implanted from opposing bone surfaces to provide stabilization of bones and bone fragments for the fixation of fractures, arthrodeses, and osteotomies. The bolts and nut-sleeves are cannulated for use with guidewires for precise placement in bone. The Bone Bolt System includes optional washers of various sizes.

#### Predicate Devices:

Primary Predicate DePuy Synthes Cortex and Cannulated Screws (K161616)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for The University of Utah. On the left is a large red block U. To the right of the U is the text "THE UNIVERSITY OF UTAH" in a serif font.

Additional Predicate Synthes 3.5mm Cortex Screws (K043185)

Reference Device Acumed Frag-Loc® System (K130986)

## Materials:

The Bone Bolt implants are made from Ti-6Al-4V per ASTM F136. The Bone Bolt accessories are made from stainless steel per ASTM F899, polypropylene and silicone.

## Performance Data:

The Bone Bolt System performance was characterized through the following tests, in accordance with ASTM F543-17 and ASTM F1264-16 Annex A4:

- Torsional Strength Testing
- Driving Torque Testing
- Axial Pull-out Strength Testing
- Static and Dynamic Bending Testing

No clinical or animal testing was conducted.

## Substantial Equivalence:

The Bone Bolt System has the same intended use and uses the same design, materials, principles of operations, and similar range of sizes as the predicate devices, the DePuy Synthes Cortex and Cannulated Screws (K1616), and Synthes 3.5mm Cortex Screws (K043185). Side-by-side performance testing demonstrates that the Bone Bolt System has equivalent or statistically significantly better performance than its predicates. The Bone Bolt System's similarities in technological characteristics to its predicates and the reference device, the Acumed Frag-Loc® System (K130986), further support its substantial equivalence. Thus, the Bone Bolt System is substantially equivalent to the DePuy Synthes Cortex and Cannulated Screws (K161616), Synthes 3.5mm Cortex Screws (K043185).

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K230867](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K230867)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
