Treace Medical Concepts (TMC) Hammertoe Fixation System
K222564 · Treace Medical Concepts, Inc. · HWC · Nov 15, 2022 · Orthopedic
Device Facts
Record ID
K222564
Device Name
Treace Medical Concepts (TMC) Hammertoe Fixation System
Applicant
Treace Medical Concepts, Inc.
Product Code
HWC · Orthopedic
Decision Date
Nov 15, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The TMC Hammertoe Fixation System is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe, and mallet toe. Cannulated implants in the TMC Hammertoe Fixation System can be used with K-wires for the delivery of implants or the temporary stabilization of outlying joints (e.g. MTP Joint). The Implantable K-Wires are indicated for use in fixation of bone reconstructions, and as guide pins for insertion of other implants. Additionally, Implantable K-Wires are indicated for the fixation of osteotomies and reconstruction of the lesser toe following correction procedures for hammertoe, claw toe, and metatarsophalangeal joint instability.
Device Story
TMC Hammertoe Fixation System consists of cannulated PEEK implants and stainless steel K-wires; used by orthopedic surgeons in clinical settings for lesser toe reconstruction; implants provide bone fixation following osteotomies; K-wires serve as guide pins or temporary stabilization for outlying joints (e.g., MTP joint); system components delivered sterile via gamma irradiation for single use; device facilitates bone healing and structural correction; clinical benefit derived from mechanical stabilization of lesser toe deformities.
Clinical Evidence
Bench testing only. Evaluated via static and dynamic four-point bending (ASTM F382-17), axial pullout (ASTM F543-17), static torsion (ASTM F543-17), and driving insertion/removal torque (ASTM F543-17).
Technological Characteristics
Cannulated implants: PEEK (ASTM F2026). K-wires: Stainless steel (ASTM F138). System includes bone preparation instruments. Sterile, single-use, gamma-irradiated. Mechanical fixation via metallic/polymer fasteners.
Indications for Use
Indicated for patients undergoing lesser toe correction procedures (hammertoe, claw toe, mallet toe) requiring fixation of osteotomies, bone reconstruction, or stabilization of metatarsophalangeal joint instability. K-wires also indicated as guide pins for implant insertion.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
K183055 — Vector Hammertoe Correction System · Nvision Biomedical Technologies, LLC · Feb 19, 2019
K190926 — Hammertoe Truss System (HTS) · 4Web, Inc. · Jul 23, 2019
K150443 — ToeTac Hammertoe Fixation System · Restore Surgical, LLC Dba Instratek · Jun 4, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
November 15, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Treace Medical Concepts, Inc. % Danielle Besal Principal Consultant MRC Global 9085 E. Mineral Circle, Suite 110 Centennial, Colorado 80112
Re: K222564
Trade/Device Name: Treace Medical Concepts (TMC) Hammertoe Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: August 24, 2022 Received: August 24, 2022
Dear Danielle Besal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'neill -S
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222564
Device Name Treace Medical Concepts (TMC) Hammertoe Fixation System
Indications for Use (Describe)
The TMC Hammertoe Fixation System is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe, and mallet toe. Cannulated implants in the TMC Hammertoe Fixation System can be used with K-wires for the delivery of implants or the temporary stabilization of outlying joints (e.g. MTP Joint).
The Implantable K-Wires are indicated for use in fixation of bone reconstructions, and as guide pins for insertion of other implants. Additionally, Implantable K-Wires are indicated for the fixation of osteotomies and reconstruction of the lesser toe following correction procedures for hammertoe, claw toe, and metatarsophalangeal joint instability.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
Treace Medical Concepts (TMC) Plating System August 24, 2022
- Company: Treace Medical Concepts, Inc. 203 Fort Wade Rd., Suite 150 Ponte Vedra, FL 32081
- Primary Contact: Danielle Besal Principal Consultant, MRC Global Phone: 901-827-8670 Danielle.Besal@askmrcglobal.com
Company/Secondary Kristina Hall Contact: Director, Regulatory Affairs Treace Medical Concepts, LLC Phone: 904-373-5940 khall@treace.net
- Trade Name: Treace Medical Concepts (TMC) Hammertoe Fixation System
- Common Name: Screw, Fixation, Bone
- Classification: Class II
- 21 CFR 888.3040 Smooth or Threaded Metallic Bone Fixation Regulation Number: Fastener
- Panel: Orthopedic
- Product Code: HWC
- Primary Predicate: K132912 In2Bones SAS, Duafit® interphalangeal implant
- Additional Predicate: K140148, Wright Medical Technology Inc., PRO-TOE® Hammertoe Fixation System
{4}------------------------------------------------
## Device Description:
The Treace Medical Concepts (TMC) Hammertoe Fixation System is for the fixation of osteotomies and reconstruction of the lesser toes. The cannulated implant is made of PEEK (Polyetheretherketone) according to ASTM F2026 and is available in different sizes. The implants are delivered sterile packaged and for single use only.
The Implantable K-wire is made from stainless steel according to ASTM F138 and available in different sizes. The k-wire is delivered sterile packaged and for single use only.
All implantable components are provided sterile by gamma irradiation.
# Indications for Use:
The TMC Hammertoe Fixation System is indicated for the fixation of osteotomies and reconstruction of the lesser toes following correction procedures for hammertoe, claw toe, and mallet toe. Cannulated implants in the TMC Hammertoe Fixation System can be used with Kwires for the delivery of implants or the temporary stabilization of outlying joints (e.g., MTP Joint).
The Implantable K-Wires are indicated for use in fixation of bone fractures, for bone reconstructions, and as guide pins for insertion of other implants. Additionally, Implantable K-Wires are indicated for the fixation of osteotomies and reconstruction of the lesser toe following correction procedures for hammertoe, claw toe, mallet toe, and metatarsophalangeal ioint instability.
## Substantial Equivalence:
The subject TMC Hammertoe Fixation System is substantially equivalent to the predicate In2Bones SAS, Duafit® interphalangeal implant (K132912) and Wright Medical Technology Inc., PRO-TOE® Hammertoe Fixation System (K140148).
The subject device, TMC Hammertoe Fixation System, is comprised of polyetheretherketone (PEEK) implant, stainless steel k-wires for implantation, and stainless steel bone preparation instruments. The system is provided sterile by gamma irradiation. The PEEK implant is manufactured according to the surgical specification requirements of ASTM F2026. The implantable k-wires are made from stainless steel according to ASTM F138.
The components are equivalent in intended use, indication for use, geometry, design, materials, and mechanical performance. Thus, it can be concluded that the subject does not raise new questions about safety and effectiveness.
{5}------------------------------------------------
#### Performance Testing:
The TMC Hammertoe Fixation System geometry, materials, and mechanical properties (Static and Dynamic Four-Point Bending per ASTM F382-17, Annex A1 and A2, Axial Pullout per ASTM F543-17, Annex A3, Static Torsion per ASTM F543-17, Annex A1, and Driving Insertion and Removal Torque per ASTM F543-17, Annex A2,) have been evaluated to confirm substantial equivalence to the predicate devices.
#### Conclusion:
The subject TMC Plating System is substantially equivalent to the predicates, In2Bones SAS, Duafit® interphalangeal implant (K132912) and Wright Medical Technology Inc., PRO-TOE® Hammertoe Fixation System (K140148).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.