Small Cannulated Screws (2.5mm - 4.0mm diameter) Small Cannulated Screws (2.5mm - 4.0mm diameter) are intended for fixation of fractures and non-unions of small bones and small bone arthrodeses. Examples include, but are not limited to scaphoid and other carpal and phalangeal fusions, osteotomies, and bunionectomies. Large Cannulated Screws (4.5mm - 7.5mm diameter) Large Cannulated Screws (4.5mm – 7.5 mm diameter) are intended for fracture fixation of large bones and large bone fragments. Diameters 6.5 mm and larger are intended for large bone fragments such as slipped capital femoral epiphyses; pediatric femoral neck fractures; tibial plateau fractures; SI joint disruptions; intercondylar femur fractures; subtalar arthrodesis and fixation of pelvis and iliosacral joint.
Device Story
Cannulated screw system for bone fragment fixation; includes screws, washers, and guide wires. Cannulated design allows precise placement over guide wire. Used in surgical settings by orthopedic surgeons. Screws/washers provide mechanical stabilization of bone fractures or arthrodesis sites. Benefits include precise anatomical alignment and secure fixation of bone fragments.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including torque to failure, driving torque, and axial pullout per ASTM F543. Device evaluated as MR Conditional.
Technological Characteristics
Materials: 316L stainless steel (ASTM F138) for screws/washers; stainless steel (ASTM F899) for instruments; cobalt chrome (ASTM F562) for guide wires. Design: Cannulated screw system for guide wire placement. Provided non-sterile; requires sterilization prior to use. MR Conditional.
Indications for Use
Indicated for fixation of fractures, non-unions, and arthrodeses of small bones (e.g., scaphoid, carpal, phalangeal, osteotomies, bunionectomies) and large bones/fragments (e.g., slipped capital femoral epiphyses, pediatric femoral neck, tibial plateau, SI joint, intercondylar femur, subtalar, pelvis, iliosacral joint).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" in a smaller font below.
Orthopediatrics, Corp Jen Gregory Regulatory Affairs Manager 2850 Frontier Drive Warsaw, Indiana 46582
Re: K190324
Trade/Device Name: OrthoPediatrics Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: July 1, 2019 Received: July 3, 2019
Dear Jen Gregory:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
July 5, 2019
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For- Shumaya Ali, MPH Assistant Director DHT6C: Division of Stereotaxic, Trauma and Restorative Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K190324
Device Name OrthoPediatrics Cannulated Screw System
#### Indications for Use (Describe)
Small Cannulated Screws (2.5mm - 4.0mm diameter)
Small Cannulated Screws (2.5mm - 4.0mm diameter) are intended for fixation of fractures and non-unions of small bones and small bone arthrodeses. Examples include, but are not limited to scaphoid and other carpal and phalangeal fusions, osteotomies, and bunionectomies.
Large Cannulated Screws (4.5mm - 7.5mm diameter)
Large Cannulated Screws (4.5mm – 7.5 mm diameter) are intended for fracture fixation of large bones and large bone fragments. Diameters 6.5 mm and larger are intended for large bone fragments such as slipped capital femoral epiphyses; pediatric femoral neck fractures; tibial plateau fractures; SI joint disruptions; intercondylar femur fractures; subtalar arthrodesis and fixation of pelvis and iliosacral joint.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the OrthoPediatrics Cannulated Screw System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'The New 510(k) Paradigm – Alternate Approaches to Demonstration Substantial Equivalence in Premarket Notifications - Final Guidance', issued on March 20, 1998.
| Sponsor: | OrthoPediatrics, Corp.<br>2850 Frontier Drive<br>Warsaw, IN 46582<br>Establishment Registration Number:<br>9102640 Phone: (574) 267-0880<br>Fax: (574) 269-3692 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Jen Gregory<br>Regulatory Affairs Manager |
| Date: | February 11, 2019 |
| Subject Device: | Trade Name: OrthoPediatrics Cannulated Screw System |
| | Regulation Number: 888.3040 |
| | Regulation Name: Smooth or threaded metallic bone fixation fastener |
| | Product Code: HWC |
| | Common Name: Screw, Fixation, Bone |
#### Legally marketed devices to which substantial equivalence is claimed:
- Primary Predicate:
- . K140891 – Biomet Cannulated Screw System Secondary Predicates:
- K012945 – Synthes 2.4 mm Cannulated Screw – For the OrthoPediatrics 2.5-4.0 mm Cannulated Screws
- . K963172 – Synthes 4.5 mm Cannulated Screw – For the OrthoPediatrics 4.5-7.5 mm Cannulated Screws
#### Device Description
The OrthoPediatrics Cannulated Screw System is a cancellous screw system intended for the use in the fixation of bone fragments. The subject Cannulated Screw System is composed of cannulated screws offered in a variety of sizes and options to accommodate patient anatomy, as well as washers and associated instrumentation to facilitate or aid in placement of the subject screws. The cannulated design of these screws allows them to be used in combination with a guide wire, also included in this
{5}------------------------------------------------
system, for precise placement into bone. The included Bone Screw Washers provide optional increased stabilization between the screw head and cortical bone. All subject screws and washers are manufactured from medical grade stainless steel per ASTM F138. All instruments are manufactured from stainless steel per ASTM F899, with the exception of the guide wires, which are manufactured from cobalt chrome per ASTM F562.
#### Intended Use and Indications for Use
## Small Cannulated Screws (2.5mm – 4.0mm diameter)
Small Cannulated Screws (2.5mm - 4.0mm diameter) are intended for fixation of fractures and nonunions of small bones and small bone arthrodeses. Examples include, but are not limited to scaphoid and other carpal fractures, metacarpal and phalangeal fusions, osteotomies, and bunionectomies.
#### Large Cannulated Screws (4.5mm – 7.5mm diameter)
Large Cannulated Screws (4.5mm – 7.5 mm diameter) are intended for fracture fixation of large bones and large bone fragments. Diameters 6.5 mm and larger are intended for large bones and large bone fragments such as slipped capital femoral epiphyses; pediatric femoral neck fractures; tibial plateau fractures; SI joint disruptions; intercondylar femur fractures; subtalar arthrodesis and fixation of pelvis and iliosacral joint.
## Summary of Technological Characteristics
The technological characteristics (materials, design, sizing) of the subject OrthoPediatrics Cannulated Screw System are similar to the following predicate devices:
#### Primary Predicate:
- K140891 Biomet Cannulated Screw System o Secondary Predicates:
- K012945 Synthes 2.4 mm Cannulated Screw For the OrthoPediatrics 2.5-4.0 mm Cannulated . Screws
- K963172 – Synthes 4.5 mm Cannulated Screw – For the OrthoPediatrics 4.5-7.5 mm Cannulated Screws
The rationale for substantial equivalence is based on consideration of the following characteristics:
- Intended Use: The OrthoPediatrics Cannulated Screw System is intended for use in the fixation of small and large bone fragments, identical to the predicate devices.
- Indications for Use: Indications for Use are similar to the predicate Biomet Cannulated Screw System (K140891) and the predicate Synthes Cannulated Screws (K963172 and K012945).
- Materials: The proposed OrthoPediatrics Cannulated Screw System screws and washers are manufactured from 316L stainless steel conforming to ASTM F138 which is identical to the material of the predicate Synthes Cannulated Screws (K963172 and K012945)
- Design Features: The proposed OrthoPediatrics Cannulated Screw System incorporates similar design features and size ranges as the predicate devices.
- . Function: The OrthoPediatrics Cannulated Screw System is intended for the use in the fixation of small and large bone fragments, identical to the predicate devices.
{6}------------------------------------------------
- Sterilization: The proposed OrthoPediatrics Cannulated Screw System is provided non-sterile and requires sterilization prior to use which is the same sterilization method utilized for the predicate devices, which are provided both sterile and non-sterile.
## Performance Data
- . Non-Clinical Tests
- Torque to Failure per ASTM F543
- Driving Torque per ASTM F543
- Axial Pullout per ASTM F543
Testing demonstrated the subject device to be substantially equivalent to predicate Synthes Cannulated Screws (K963172 and K012945) devices.
- . The OrthoPediatrics Cannulated Screw System was evaluated for use in an MR Environment and were determined to be MR Conditional.
- . Clinical Tests None provided as a basis for substantial equivalence.
## Substantial Equivalence Conclusion
OrthoPediatrics believes that the subject device is substantially equivalent to the legally marketed predicate device based on intended use, technology, materials, as well as the mechanical testing and biocompatibility assessment.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.