The ICONN Campbell Interference Screw System is intended for fixation of tissue, including ligament or tendon to bone for the following indications: Shoulder: Rotator Cuff Repairs, Bankart Repair, Biceps Tenodesis. Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle. Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis. Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist, Lateral Epicondylitis repair.
Device Story
Implantable interference screw; secures soft tissue (ligament/tendon) to bone. Used by orthopedic surgeons in clinical/OR settings. Available in PEEK (Solvay Zeniva ZA-600L) or Titanium (Ti 6Al-4V ELI Grade 23). Features square-shaped driver interface and squared-off thread profile. Sterile, single-use device; gamma-irradiated. Surgeon inserts screw into bone tunnel to provide mechanical fixation of graft/tissue. Benefits patient by restoring joint stability and facilitating healing in reconstructive orthopedic procedures.
Clinical Evidence
Bench testing only. Evaluated axial pullout strength, cyclic loading, and displacement per ASTM F543-2013. Insertion characteristics (torsional properties and driving torque) analyzed. Biocompatibility testing (cytotoxicity, irritation, sensitization) performed per ISO 10993-1:2009/(R)2013. Endotoxin testing conducted per ANSI/AAMI ST72:2016 (results < 0.05 EU/device).
Technological Characteristics
Materials: Solvay Zeniva ZA-600L PEEK or Ti 6Al-4V ELI (Grade 23) titanium. Dimensions: 6mm-12mm diameters, 20mm-25mm lengths. Design: Threaded bone fastener with square driver interface. Sterilization: Gamma irradiation. Standards: ASTM F543-2013 (mechanical testing), ISO 10993-1 (biocompatibility), ANSI/AAMI ST72:2016 (endotoxin).
Indications for Use
Indicated for fixation of ligament or tendon to bone in shoulder, foot/ankle, knee, elbow, and hand/wrist surgical procedures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
December 7, 2018
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
ICONN Orthopedics, LLC Whitt Israel Chief Executive Officer 400 Union Hill Drive, Suite 150 Birmingham, Alabama 35209
Re: K180670
Trade/Device Name: ICONN Campbell Interference Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC, MBI Dated: November 2, 2018 Received: November 5, 2018
Dear Mr. Israel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180670
Device Name ICONN Campbell Interference Screw
#### Indications for Use (Describe)
The ICONN Campbell Interference Screw System is intended for fixation of tissue, including ligament or tendon to bone for the following indications:
Shoulder: Rotator Cuff Repairs, Bankart Repair, Biceps Tenodesis. Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle.
Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis.
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist, Lateral Epicondylitis repair.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "ICONN Orthopedics". The logo consists of the word "ICONN" in a sans-serif font, with the two "N"s being slightly larger than the other letters. The "O" in the middle is a blue circle made of three curved lines. Below the word "ICONN" is the word "ORTHOPEDICS" in a smaller, sans-serif font, with a line on either side of the word.
400 Union Hill Drive, Suite 150 Birmingham, AL 35209
## 510(k) SUMMARY
| Submitter's Name: | ICONN Orthopedics, LLC |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's Address: | 400 Union Hill Drive<br>Suite 150<br>Birmingham, AL 35209 |
| Submitter Contact Person: | Whitt Israel<br>CEO<br>(205) 913-2068 |
| Date Summary was Prepared: | March 12, 2018 |
| Trade Name or Proprietary<br>Name: | ICONN Campbell Interference Screw |
| Common or Usual Name: | Interference Screw |
| Classification: | Class II per 21 CFR §888.3040 |
| Product Code: | HWC - Screw, Fixation, Bone<br>MBI - Fastener, Fixation, Nondegradable, Soft Tissue |
| Classification Panel: | Division of Orthopedic Devices |
| Knee: | Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis. |
| Elbow: | Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction. |
| Hand/Wrist: | Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist, Lateral Epicondylitis repair. |
| Predicate Devices: | The predicate device was selected to help demonstrate substantial equivalence of several characteristics to include intended use, design, sizes, material, and sterility. The predicate device is one product but is available in two materials: Arthrex PEEK Interference Screw (K062466) Arthrex Titanium Interference Screw (K062466) |
| | The predicate device is intended for fixation of tissue, including ligament or tendon to bone for the same indications listed for the ICONN Campbell Interference Screw. |
| Technological Characteristics: | The ICONN Campbell Interference Screw System interference screws are of similar design to the predicate and are manufactured with the same basic material in similar sizes. The differences are in the geometric design of the thread profile and the driver/screw interface. The thread geometry of the ICONN Campbell Interference Screws have a squared off base, while the Arthrex screws have a rounded trough. In regard to driver interface, the ICONN Campbell screws have a square shaped driver interface, while the Arthrex screws have a star shaped driver interface. Due to the design differences, performance testing was conducted to characterize the performance of the ICONN Campbell Interference Screw in comparison to the predicate device. |
| Performance Data: | ASTM F543-2013 was followed in order to compare axial pullout strength (section A3 of the standard) after cyclic loading and displacement between the subject device and predicate device along with insertion characteristics of the subject device. The version of the screw made out of PEEK material was tested, as that material constitutes a worst case between the two materials offered. For cyclic loading and axial pullout comparison, the Arthrex PEEK Interference Screw was tested as it is made out of similar material. For insertion testing, torsional properties |
Device Description: The ICONN Campbell Interference Screws (various part numbers) are sterile (gamma irradiation), single use implantable devices available in two materials: Solvay's Zeniva®ZA-600L polyetheretherketone (PEEK) or Ti 6Al-4 ELI (Grade 23) and is intended for fixation of tissue, including ligament or tendon to bone. The screws are available in multiple sizes ranging from diameters of Ø6mm - Ø12mm in 1mm increments. The Ø6mm screw is available only in 20mm long, the Ø7mm - Ø10mm screws are available in 20mm and 25mm lengths, and the Ø11mm and Ø12mm screws are available in 25mm lengths only. Intended Use: The ICONN Campbell Interference Screw System is intended for fixation of tissue, including ligament or tendon to bone for the following indications: Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle.
{4}------------------------------------------------
{5}------------------------------------------------
(section A1 of the standard) and driving torque (section A2 of the standard) were analyzed and compared to demonstrate the device can be safely implanted.
Testing showed that the ICONN Campbell Interference Screw performed similarly or better than the predicate or within acceptable ranges for the intended use for axial pullout after cyclic loading and displacement during cyclic loading. For insertion, testing results showed that the device can be safely implanted.
Cytotoxicity, irritation, and sensitization, testing were carried in accordance with ISO 10993-1:2009/(R)2013 for the ICONN Campbell Interference Screws for both the PEEK and titanium screws. All testing met acceptance criteria.
In accordance with ANSI/AAMI ST72:2016 LAL testing was conducted using the Kinetic-Chromogenic assay for the ICONN Campbell PEEK and Titanium Interference Screws. One (1) each of the test articles were placed in 10 mL of Water for Injection and extracted at 37°C using a shaker for at least an hour. Positive and inhibition enhancement controls were included. Results of the PEEK and Titanium Interference Screws demonstrated the test articles were < 0.05 EU/device.
### Conclusion:
The ICONN Campbell Interference Screw is substantially equivalent to the predicate and does not raise different questions of safety and effectiveness.