K170698 · Providence Medical Technology, Inc. · HWC · May 31, 2017 · Orthopedic
Device Facts
Record ID
K170698
Device Name
ALLY(TM) Bone Screws
Applicant
Providence Medical Technology, Inc.
Product Code
HWC · Orthopedic
Decision Date
May 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Intended Use
ALLY™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Device Story
ALLY Bone Screws are titanium alloy, single-use, fully threaded cortical screws; available in headed, headless, or snap-off configurations with various diameters/lengths. Used by surgeons in clinical settings for bone reconstruction, osteotomy, arthrodesis, joint fusion, and fracture repair/fixation. Screws provide mechanical stabilization of bone segments to facilitate healing. Device selection is based on anatomical site and size requirements.
Clinical Evidence
Bench testing only. Evaluated for static torsion, static pull-out, and driving torque per ASTM F543. Engineering analysis performed for screw bending strength. Bacterial endotoxin testing (LAL) confirmed compliance with device limits.
Technological Characteristics
Titanium alloy bone screws; headed, headless, or snap-off configurations; various diameters and lengths. Testing performed per ASTM F543. Single-use.
Indications for Use
Indicated for bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation in patients requiring bone fixation appropriate to device size.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Orthohelix Surgical Designs, Inc.: CSS Cannulated Screw System (K060428)
Wright Medical Technology, Inc.: WRIGHT Compression Screws (K082320)
Related Devices
K121713 — PMT BONE SCREWS · Providence Medical Technology, Inc. · Sep 27, 2012
K130319 — MEDLINE CANNULATED SCREW · Medline Industries, Inc. · May 31, 2013
K102343 — EMERGE MEDICAL CANNULATED SCREW FIXATION SYSTEM · Emerge Medical · Nov 23, 2010
K203239 — Arthrex Low Profile Screws · Arthrex, Inc. · Dec 30, 2020
K082320 — DARCO SMALL SCREW · Wrightmedicaltechnologyinc · Nov 5, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or a sense of movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Providence Medical Technology, Inc. Edward Liou Chief Operating Officer 1331 N. California Blvd., Suite 320 Walnut Creek, California 94596
May 31, 2017
Re: K170698
Trade/Device Name: ALLYTM Bone Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: May 17, 2017 Received: May 18, 2017
Dear Mr. Liou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Statement
510(k) Number (if known):_K170698
Device Name: ALLY™ Bone Screws
Indications for Use:
ALLY™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation, appropriate for the size of the device.
Prescription Use ___ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) SUMMARY
## Providence Medical Technology's ALLYTM Bone Screws
| Date Prepared: | May 26, 2017 |
|-------------------------|------------------------------------------------------------------------------------------------------|
| Company: | Providence Medical Technology, Inc.<br>1331 N. California Blvd., Suite 320<br>Walnut Creek, CA 94596 |
| Contact Person: | Edward Liou<br>ed@providencemt.com<br>Phone: 415-923-9376<br>Fax: 415-923-9377 |
| Trade Name: | ALLY™ Bone Screws |
| Common Name: | Bone screw |
| Classification Name: | Screw, fixation, bone |
| Regulation Number: | 21 CFR Part 880.3040 |
| Product Code:<br>Class: | HWC<br>Class II |
| Predicate Devices: | |
| Primary Predicate: | PMT Bone Screws (K121713, cleared 09/27/2012) |
| Reference Devices: | Orthohelix Surgical Designs, Inc.: CSS Cannulated Screw System<br>(K060428, cleared 03/07/2006) |
# (K082320, cleared 11/05/2008)
## Device Description
The ALL Y™ Bone Screws are fully threaded cortical screws offered in various diameters and lengths. All screws are manufactured from titanium alloy. The implants are single use only.
#### Indications for Use
The ALLY™ Bone Screws are indicated for use in bone reconstruction, osteotomy, arthrodesis, joint fusion, fracture repair, and fracture fixation appropriate for the size of the device.
Wright Medical Technology, Inc.: WRIGHT™ Compression Screws
{4}------------------------------------------------
## Technological Characteristics
The ALLYTM Bone Screws have the following technological characteristics:
- o Manufactured from Titanium Alloy
- o Available headed, headless, or snap-off with various tip configurations
- 0 Available in various diameters and lengths
#### Performance Data
The ALLY™ Bone Screws are evaluated for static torsion, static pull out, and driving torque testing per ASTM F543 Standard Specification and Test Methods for Metallic Bone Screws.
Engineering analysis for screw bending strength has been completed.
Bacterial Endotoxin testing (LAL) has been completed and shown to meet the device limits.
#### Substantial Equivalence
The ALLY™ Bone Screws are as safe and effective as the predicate device. The ALLY™ Bone Screws have the same indications for use, technological characteristics, and principles of operation as its predicate device. The reference devices are available in the extended size ranges being added to the ALLYTM Bone Screws.
Comparison between ALLY™ Bone Screws and its predicate devices raises no new issues of safety or effectiveness. The ALLY™ Bone Screws are substantially equivalent to the PMT Bone Screws.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.