← Product Code [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC) · K151701

# Tenodesis screw system (K151701)

_Normed Medizin-Technik GmbH · HWC · Jan 26, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K151701

## Device Facts

- **Applicant:** Normed Medizin-Technik GmbH
- **Product Code:** [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC.md)
- **Decision Date:** Jan 26, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Tenodesis screws are intended for soft tissue reattachment procedures in foot surgery. The system is indicated for soft tissue reattachment procedures in foot surgery, such as tendon reconstructions and tendon transfers. The system can be used in adult patients. Additional information is provided in the corresponding surgical techniques.

## Device Story

Tenodesis screw system consists of threaded, fully cannulated titanium alloy (Ti-6Al-4V) anchors used for soft tissue reattachment in foot surgery; specifically tendon reconstructions and transfers. Implants are provided non-sterile for single-use; associated instruments are reusable. Device is used by surgeons in clinical settings to secure soft tissue to bone. Mechanical fixation is achieved via screw insertion; providing stability during healing. Benefits include reliable reattachment of tendons in foot procedures.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing (biocompatibility per ISO 10993-1, mechanical performance testing) and historical market presence of predicate devices.

## Technological Characteristics

Material: Titanium alloy Ti-6Al-4V (ASTM F136). Design: Threaded, fully cannulated bone fixation fastener. Energy source: None (mechanical). Connectivity: None. Sterilization: Non-sterile (user-sterilized).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Herbert/Whipple Bone Screw System, Herbert Bone Screw, Herbert Cannulated Bone Screw System, and Herbert Mini Bone Screw ([K143165](/device/K143165.md))
- NCB Cannulated Screw ([K061211](/device/K061211.md))
- G-Force Tenodesis Screw ([K083150](/device/K083150.md))
- Arthrex Tenodesis Screw Family ([K051726](/device/K051726.md))
- Arthrex Resorbable Interference Screw (Rattler Interference Screw) ([K041274](/device/K041274.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Normed Medizin-Technik GmbH Mr. Arne Briest RA/QA Manager Ulrichstrasse 7 Tuttlingen, D-78532 Germany

January 26, 2016

Re: K151701

Trade/Device Name: Tenodesis screw system Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC, MBI Dated: December 23, 2015 Received: December 28, 2015

Dear Mr. Arne Briest:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K151701

Device Name Tenodesis screw system

Indications for Use (Describe)

The Tenodesis screws are intended for soft tissue reattachment procedures in foot surgery.

The system is indicated for soft tissue reattachment procedures in foot surgery, such as tendon reconstructions and tendon transfers.

The system can be used in adult patients. Additional information is provided in the corresponding surgical techniques.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

### This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

Page 1 of 1

SC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for NORMED, a Zimmer company. The logo is black and white and features the word "NORMED" in a bold, sans-serif font. The "O" in NORMED is replaced by a globe. Below the word "NORMED" is the text "A ZIMMER COMPANY" in a smaller, sans-serif font.

# 510(k) Summary

# 1. Submission Sponsor and Correspondent

 Normed Medizin-Technik GmbH Ulrichstrasse 7 D-78532 Tuttlingen Germany

| Phone:                            | + 49 7461 93 43 0  |
|-----------------------------------|--------------------|
| Fax:                              | + 49 7461 93 43 20 |
| Contact:                          | Mr. Arne Briest    |
| FDA Establishment Registration #: | 9611091            |

#### 2. Date Prepared

January 26, 2016

#### 3. Device Identification

| Trade/Proprietary Name:    | Tenodesis screw system                                                       |
|----------------------------|------------------------------------------------------------------------------|
| Common/Usual Name:         | Screw (smooth or threaded metallic bone fixation fastener)                   |
| Classification Name:       | Screw, Fixation, Bone and<br>Fastener, Fixation, Non degradable, Soft Tissue |
| Classification Regulation: | 21CFR 888.3040                                                               |
| Product Code:              | HWC/MBI                                                                      |
| Device Class:              | Class II                                                                     |
| Classification Panel:      | Orthopedic                                                                   |

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for Normed, a Zimmer company. The logo is black and white and features the word "NORMED" in large, bold letters. Above the word "NORMED" is the text "The specialist for small bones". Below the word "NORMED" is the text "A ZIMMER COMPANY".

### 4. Legally Marketed Predicate Device

K143165 - Herbert/Whipple Bone Screw System, Herbert Bone Screw, Herbert Cannulated Bone Screw System, and Herbert Mini Bone Screw

K061211 - NCB Cannulated Screw

K083150 - G-Force Tenodesis Screw

K051726 - Arthrex Tenodesis Screw Family

K041274 - Arthrex Resorbable Interference Screw (Rattler Interference Screw)

# 5. Device Description

The Tenodesis screw system is intended for soft tissue reattachment procedures in foot surgery. The screws are made of titanium alloy Ti-6AI-4V (ASTM F136). They are threaded, fully cannulated anchors.

The implants are offered in various sizes. The implants are provided non-sterile and single-use only.

The Tenodesis instruments are non-sterile and reusable.

### 6. Indications for Use

The Tenodesis screws are intended for soft tissue reattachment procedures in foot surgery.

The system is indicated for soft tissue reattachment procedures in foot surgery, such as tendon reconstructions and tendon transfers.

The system can be used in adult patients. Additional information is provided in the corresponding surgical techniques.

# 7. Substantial Equivalence Discussion

The Tenodesis screw system has the same intended use as the G-Force Tenodesis Screw, the Arthrex Tenodesis Screw Family and the Arthrex Resorbable Interference Screw. The Tenodesis screw system has similar performance characteristics, is manufactured from similar materials and is similar in design to the all selected predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the logo for NORMED, a Zimmer company. The logo is black and white and features the word "NORMED" in large, bold letters. The "O" in NORMED contains a graphic of the earth with lines of longitude and latitude. Below the word "NORMED" is the text "A ZIMMER COMPANY" in a smaller font.

### 8. Non-Clinical Performance Data

- -Biocompatibility - Biocompatibility testing on the plates was conducted and evaluated per ISO 10993-1. All testing passed.
- -Self-tapping performance, driving, torsional properties and axial pullout strength of the Tenodesis screw system and the predicate devices resulted in similar mechanical properties and performance. The subject and predicate devices are substantially equivalent.

# 9. Clinical Performance Data

There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device.

# 10. Statement of Substantial Equivalence

By definition, a device is substantially equivalent to a predicate device has the same intended use (G-Force Tenodesis Screw, Arthrex Tenodesis Screw Family and the Arthrex Resorbable Interference Screw) and the same technological characteristics (all selected predicates) as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.

It has been shown in this 510(k) submission that the difference between the Tenodesis screws and the predicate devices do not raise any questions regarding its safety and effectiveness.

Performance testing and compliance with voluntary standards, demonstrate that the Tenodesis screw system is substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, biocompatibility, performance characteristics, and intended use (G-Force Tenodesis Screw, Arthrex Tenodesis Screw Family and the Arthrex Resorbable Interference Screw).

The Tenodesis screw system is determined to be substantially equivalent to the referenced predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K151701](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K151701)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
