← Product Code [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC) · K150211

# EzFuze Implant System (K150211)

_Vilex IN Tennessee, Inc. · HWC · Nov 10, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K150211

## Device Facts

- **Applicant:** Vilex IN Tennessee, Inc.
- **Product Code:** [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC.md)
- **Decision Date:** Nov 10, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Fixation of osteotomies and reconstruction of the lesser toes and lesser fingers following correction procedures for hammertoe, claw toe, mallet toe, and other deformities of the feet and hands.

## Device Story

EzFuze Implant System consists of single-piece metallic implants for PIP joint fixation in lesser toes and digits; available in two styles: barbed prongs on both ends or hybrid design with screw at one end and barbed prongs at other; offered straight or with 10° bend. Used by surgeons in clinical settings for bone fixation during reconstructive procedures. Implants provide mechanical stabilization of osteotomies to facilitate healing and correct deformities. Device functions as a permanent internal fixation fastener.

## Clinical Evidence

No clinical data provided; substantial equivalence supported by bench testing and comparative analysis of design, materials, and indications.

## Technological Characteristics

Single-piece metallic bone fixation fasteners; barbed prong or hybrid screw/barbed designs; straight or 10° bend configurations. Materials are identical to predicate devices. Class II device (21 CFR 888.3040).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Trident Fusion Implant by Vilex ([K141937](/device/K141937.md))
- SmartToe by Stryker ([K070598](/device/K070598.md))
- X-Fuse by Stryker ([K112197](/device/K112197.md))
- OSSplate Hammerlock by BioMedical Enterprises ([K091951](/device/K091951.md))
- OSSplate Hammerlock by BioMedical Enterprises ([K131640](/device/K131640.md))
- DigiFuse by Metasurg ([K111536](/device/K111536.md))
- eZ-Staple Superelastic Bone Fixation Staple by Vilex ([K112837](/device/K112837.md))
- Newdeal K-wire by Newdeal ([K022599](/device/K022599.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 10, 2015

Vilex (in Tennessee), Incorporated % Abraham Lavi, Ph.D. Consultant Vilex, Incorporated 8374 Market Street, #167 Lakewood Ranch, Florida 34202

Re: K150211

Trade/Device Name: EzFuze™ Implant System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: September 25, 2015 Received: September 28, 2015

Dear Dr. Lavi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for Vilex Inc. The logo consists of the word "VILEX" in red, with a red box around it. Below the logo, it says "Manufacturer of Vilex™ bone implants, Power equipment & surgical instruments".

Phone: www.vilex.com

Fax:

(931) 474-7550 (931) 474-7551

111 Moffitt Street McMinnville, TN 37110 USA E-mail: info@vilex.com

# INDICATIONS FOR USE

### 510(k) NUMBER:

DEVICE NAME: EzFuze™ Implant System

#### INDICATIONS FOR USE:

The EzFuze Implants have the following Indications for Use:

Fixation of osteotomies and reconstruction of the lesser fingers following correction procedures for hammertoe, claw toe, mallet toe, and other deformities of the feet and hands.

Prescription Use × (Per 21 CFR 801.109)

Concurrence of CDRH, Office of Device Evaluation (DOE)

Page 1 of 1

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Vilex Inc. The logo consists of the word "VILEX" in large, red, bold letters, with the letters connected to form a rectangular shape. Below the logo, the text reads "Manufacturer of Vilex™ bone implants, Power equipment & surgical instruments".

Phone:
Fax: www.vilex.com (931) 474-7550
(931) 474-7551

111 Moffitt Street
McMinnville, TN 37110 USA
E-mail: info@vilex.com

# 510(k) Summary EzFuze Implant System

| Sponsor:                          | Vilex in Tennessee, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                   | Abraham Lavi                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Date Prepared:                    | September 25, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Trade Name:                       | EzFuze Implant System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Common Name:                      | Threaded metallic bone fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification<br>Name:           | 21 CFR 888.3040 - Smooth or threaded metallic bone fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product Code:                     | HWC/ Orthopedics, Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Predicate Devices:                | Trident Fusion Implant by Vilex (K141937)<br>SmartToe and X-Fuse by Styker (K070598 and K112197)<br>OSSplate Hammerlock by BioMedical Enterprises (K091951 and K131640)<br>DigiFuse by Metasurg (K111536)<br>eZ-Staple Superelastic Bone Fixation Staple by Vilex (K112837)<br>Newdeal K-wire by Newdeal (K022599)                                                                                                                                                                                                                                                                                                                        |
| Description of<br>Device:         | The Vilex EzFuze Implant System includes single-piece devices intended for the fixation of<br>PIP joints in lesser toes and digits. Two styles are offered: EzFuze Implant with barbed<br>prongs on both ends and Hybrid EzFuze Implant with a screw design at one end and the other<br>end includes barbed prongs. The EzFuze and Hybrid EzFuze Implants are offered straight or<br>with a 10° bend.                                                                                                                                                                                                                                     |
| Indications for<br>Use:           | The EzFuze Implants have the following Indications for Use:<br>Fixation of osteotomies and reconstruction of the lesser toes and lesser fingers following<br>correction procedures for hammertoe, claw toe, mallet toe, and other deformities of the feet<br>and hands.                                                                                                                                                                                                                                                                                                                                                                   |
| Technological<br>Characteristics: | The technological characteristics for the EzFuze Implant System are the same as the<br>characteristics of the predicate devices. All of the sizes included in the Vilex EzFuze<br>Implant system are within the range of offerings of the predicate devices and the designs of<br>the EzFuze devices are similar to the predicate devices. The materials used to manufacture<br>the EzFuze Implants are the same as those used to manufacture the predicate devices.                                                                                                                                                                      |
| Substantial<br>Equivalence        | The design features of the EzFuze Implant System are substantially equivalent to the design<br>features of other predicate devices previously cleared for market. The methods used to<br>establish equivalence are indications for use, material of construction, sizes, and shapes. The<br>safety and effectiveness of the EzFuze Implant System are adequately supported by the<br>substantial equivalence information, material information and analysis data provided within<br>this Premarket Notification. Therefore, it is concluded that the EzFuze Implant System is<br>substantially equivalent to the noted predicate devices. |
| Conclusions:                      | While the EzFuze Implant System is not identical to the predicate devices, any differences that may exist do not significantly affect device safety and effectiveness. In addition, the differences do not add new or increased risks and complications. Therefore, it is concluded that the EzFuze Implant System is substantially equivalent to the predicate devices as outlined previously and should not render the subject device NSE.                                                                                                                                                                                              |

{4}------------------------------------------------

# 510(k) Summary EzFuze Implant System

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K150211](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K150211)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
