← Product Code [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC) · K111265

# METASURG SUBTALAR IMPLANT (K111265)

_Metasurg · HWC · Aug 9, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K111265

## Device Facts

- **Applicant:** Metasurg
- **Product Code:** [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC.md)
- **Decision Date:** Aug 9, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Metasurg subtalar implant is indicated for use in the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is designed to block forward, downward and medial displacement of the talus, thus allowing normal subtalar joint motion but blocking excessive pronation and the resulting sequela. Indications include: • Severe pronation • Calcaneal valgus deformity • Plantarflexed talus • Failed correction with long term orthotic treatment • Congenital and painful flatfoot deformity • Repair of tarsal coalitions • Subtalar instability • Posterior tibial tendon dysfunction • Paralytic flat foot deformity The Metasurg subtalar implants are intended for single use only.

## Device Story

Metasurg Subtalar Implant; one-piece titanium device; implanted into sinus tarsi of foot. Function: blocks forward, downward, and medial displacement of talus; restricts excessive pronation while permitting normal subtalar joint motion. Used by surgeons for stabilization of subtalar joint in patients with flatfoot deformities or instability. Single-use device.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Material: Titanium Alloy (Ti 6Al-4V ELI). Form factor: One-piece implant, 8mm to 12mm diameter range. Intended for single use.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K111265

/1,

-

# 510(k) Summary

:

AUG - - 9 9 2011

| Date Prepared:           | May 3, 2011                                                                                                                                                                                                                                                                                                                                        |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                 | Metasurg<br>16350 Park Ten Place, Suite 101<br>Houston, TX 77084                                                                                                                                                                                                                                                                                   |
| Company Contact:         | Joshua Scott<br>Phone: (281) 398-5656<br>Fax: (281) 398-5660                                                                                                                                                                                                                                                                                       |
| Device Trade Name:       | Metasurg Subtalar Implant                                                                                                                                                                                                                                                                                                                          |
| Classification Name:     | Smooth & threaded metallic bone fixation fasteners (21<br>CFR 888.3040, Product Code HWC, Class II)                                                                                                                                                                                                                                                |
| Common Name:             | Subtalar Arthrorisis Implant                                                                                                                                                                                                                                                                                                                       |
| Substantial Equivalence: | Documentation is provided which demonstrates the Metasurg<br>Subtalar Implant to be substantially equivalent to other legally<br>marketed devices.                                                                                                                                                                                                 |
| Device Description:      | The Metasurg Subtalar Implant is a one-piece device made of<br>titanium intended to be implanted into the sinus tarsi of the foot.<br>The implant is offered in 5 sizes ranging from 8mm to 12mm in<br>diameter. The implant is used in the treatment of the talus<br>relative to the calcaneus.                                                   |
| Intended Usage:          | The Metasurg subtalar implant is indicated for use in the<br>treatment of the hyperpronated foot and stabilization of the<br>subtalar joint. It is designed to block forward, downward and<br>medial displacement of the talus, thus allowing normal subtalar<br>joint motion but blocking excessive pronation and the resulting<br>sequela.       |
|                          | Indications include:<br>• Severe pronation<br>• Calcaneal valgus deformity<br>• Plantarflexed talus<br>• Failed correction with long term orthotic treatment<br>• Congenital and painful flatfoot deformity<br>• Repair of tarsal coalitions<br>• Subtalar instability<br>• Posterior tibial tendon dysfunction<br>• Paralytic flat foot deformity |
|                          | The Metasurg subtalar implants are intended for single use only.                                                                                                                                                                                                                                                                                   |
| Material:                | Titanium Alloy (Ti 6Al-4V ELI)                                                                                                                                                                                                                                                                                                                     |

Section 5 - Page 1 of 1

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Metasurg % Mr. Joshua Scott 16350 Park Ten Place Suite 101 Houston, Texas 77084

- 9 2011 માંદિ

Re: K111265

Trade/Device Name: Metasurg Subtalar Implant Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: July 19, 2011 Received: July 20, 2011

Dear Mr. Scott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{2}------------------------------------------------

Page 2 - Mr. Joshua Scott

CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutlTDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson

Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Indications for Use

## 510(k) Number: Pending_ K11/ 265

### Device Name: Metasurg Subtalar Implant

#### Indications for Use:

The Metasurg subtalar implant is indicated for use in the treatment of the hyperpronated foot and stabilization of the subtalar joint. It is designed to block forward, downward and medial displacement of the talus, thus allowing normal subtalar joint motion but blocking excessive pronation and the resulting sequela.

Indications include:

- Severe pronation .
- Calcaneal valgus deformity .
- Plantarflexed talus .
- Failed correction with long term orthotic treatment .
- Congenital and painful flatfoot deformity .
- . Repair of tarsal coalitions
- Subtalar instability .
- Posterior tibial tendon dysfunction .
- Paralytic flat foot deformity .

The Metasurg subtalar implants are intended for single use only.

Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WIRTE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices

510(k) Number. 12111256

Section 4 - Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K111265](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K111265)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
