← Product Code [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC) · K111000

# PARCUS 2.0MM / 2.5MM / 3.5MM MITI TITANIUM SUTURE ANCHOR (K111000)

_Parcus Medical, LLC · HWC · Jul 28, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K111000

## Device Facts

- **Applicant:** Parcus Medical, LLC
- **Product Code:** [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC.md)
- **Decision Date:** Jul 28, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Parcus MiTi Suture Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications: Shoulder Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair. Knee Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair. Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair. Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Foot/Ankle Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair. Elbow Tennis Elbow Repair, Biceps Tendon Reattachment. Hand/Wrist Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.

## Device Story

Parcus MiTi Suture Anchor; threaded, tapered titanium alloy (Ti-6Al-4V ELI) fastener; used for soft tissue-to-bone attachment. Available in 2.0mm, 2.5mm, and 3.5mm diameters; preloaded with sutures (#3-0 to #2) and driver. Operated by surgeons in clinical settings. Device provides mechanical fixation of soft tissue to bone; facilitates healing in orthopedic repairs (e.g., rotator cuff, ligament reconstruction). Performance verified via pull-out strength and insertion torque testing.

## Clinical Evidence

Bench testing only. Evaluated pull-out strength and insertion torque. Side-by-side comparisons with predicate devices demonstrated no significant differences in performance metrics.

## Technological Characteristics

Material: Titanium alloy Ti-6Al-4V ELI (ASTM F136). Design: Threaded, tapered fastener. Diameters: 2.0mm, 2.5mm, 3.5mm. Configuration: Preloaded with sutures (#3-0 to #2) and driver. Energy source: None (mechanical).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Parcus V-LoX Titanium Suture Anchors ([K090075](/device/K090075.md))
- Smith & Nephew Twinfix Ti 2.8 Suture Anchor ([K053344](/device/K053344.md))
- Smith & Nephew MINITAC Ti 2.0 Suture Anchor ([K000797](/device/K000797.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image displays the logo for PARCUS Medical, LLC, accompanied by the phrase "SURGICAL INNOVATION VALUE DRIVEN" beneath the logo. Above the logo, the text "K111000 JUL 28 2011 P.1/2" is present. The logo is in bold, black font, and the text below is in a smaller, lighter font.

# 510(k) Summarv

| Submitter:           | Parcus Medical, LLC<br>839 South Neenah Ave.<br>Sturgeon Bay, WI 54234                        |
|----------------------|-----------------------------------------------------------------------------------------------|
| Company Contact:     | Barton Bracy<br>Phone: (920) 746-2972<br>Fax: (920) 746-2974                                  |
| Date Prepared:       | April 8, 2011                                                                                 |
| Trade Name:          | Parcus MiTi Suture Anchor                                                                     |
| Common Name:         | Suture Anchor                                                                                 |
| Classification Name: | Fastener, Fixation, Non-Degradable, Soft Tissue<br>21 CFR 888.3040 - Product Code HWC and MBI |

## Predicate Devices:

- Parcus V-LoX Titanium Suture Anchors (K090075) .
- Smith & Nephew Twinfix Ti 2.8 Suture Anchor (K053344) .
- Smith & Nephew MINITAC Ti 2.0 Suture Anchor (K000797) .

# Device Description:

The Parcus MiTi Suture Anchor is a threaded, tapered fastener for use in attachment of soft tissue to bone. The device is made from a Titanium alloy, Ti-6Al-4V ELI (ASTM F136). The product family includes devices that come preloaded with either one or two sutures between #3-0 and #2 in size either with or without attached needles, and a driver, and is available in three different diameters, 2.0mm, 2.5mm and 3.5mm. The 2.0mm anchor is offered in the following suture configurations: two #3-0 and one or two #2-0 sutures. The 2.5mm anchor is offered in the following suture configurations: two #3-0, two #2-0 and one #2 sutures. The 3.5mm anchor is offered in the following suture configurations: two #3-0, two #1 and two #2 sutures. All configurations are offered in multiple color variations and with or without attached needles.

### Intended Use:

The Parcus MiTi Suture Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications:

- Shoulder Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.

{1}------------------------------------------------

# k 111 000

| PARCUS                              |                                                                                                                                                                                                                         |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SURGICAL INNOVATION<br>VALUE DRIVEN |                                                                                                                                                                                                                         |
| Knee                                | Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair,<br>Posterior Oblique Ligament Repair, Extra Capsular Reconstruction,<br>Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair. |
| Foot/Ankle                          | Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles<br>Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament<br>Repair.                                                            |
| Elbow                               | Tennis Elbow Repair, Biceps Tendon Reattachment.                                                                                                                                                                        |
| Hand/Wrist                          | Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral<br>Ligament Reconstruction, TFCC.                                                                                                                      |

# Substantial Equivalence Summary:

The Parcus MiTi Suture Anchor is substantially equivalent to the predicate devices listed above in which the basic features and intended uses are very similar. Any differences between the MiTi Suture Anchor and the predicate devices are considered minor and do not raise questions concerning safety and effectiveness.

# Summary Performance Data:

The pull out strength and insertion torque was measured for the Parcus MiTi Suture Anchors. The published literature was reviewed and side by side comparisons were done with the Parcus Medical and Smith & Nephew predicate devices. The results of the insertion torque testing, pullout force, and the literature review demonstrated that there were no significant differences between the Parcus MiTi Suture Anchors and the predicate devices.

Fax 1.920.746.2974

PARCUSMEDICAL.COM

Call 1.877.746.2972

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services-USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread, with three horizontal lines running through the body of the eagle. The image is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Parcus Medical, LLC % Mr. Barton Bracy Vice President Marketing and Product Development 839 South Neenah Avenue Sturgeon Bay, Wisconsin 54235

JUL 28 2011

Re: K111000

Trade/Device Name: Parcus MiTi Suture Anchors Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC, MBI Dated: July 19, 2011 Received: July 20, 2011

Dear Mr. Bracy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

# Page 2 - Mr. Barton Bracy

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,
Mark M. Milliman

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name: _Parcus MiTi Suture Anchors

### Indications for Use:

The Parcus MiTi Suture Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications:

- Shoulder Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.
- Knee Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair. Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.
- Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Foot/Ankle Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair.
- Elbow Tennis Elbow Repair, Biceps Tendon Reattachment.
- Hand/Wrist Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.

FCC.
Muh M. Mulkun

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number

< IIII 000

Prescription Use X (Part 21 CFR 801 Subpart D) Over the Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K111000](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K111000)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
