APERFIX FEMORAL IMPLANT WITH INSERTER, 9,10,11 MM X 35MM AND X 29MM

K083612 · Cayenne Medical, Inc. · HWC · Mar 4, 2009 · Orthopedic

Device Facts

Record IDK083612
Device NameAPERFIX FEMORAL IMPLANT WITH INSERTER, 9,10,11 MM X 35MM AND X 29MM
ApplicantCayenne Medical, Inc.
Product CodeHWC · Orthopedic
Decision DateMar 4, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The AperFix® Femoral Implant with Inserter is intended for use in tenodesis procedures with soft tissue grafts, utilizing either arthroscopic or open techniques during Anterior Cruciate Ligament (ACL), Posterior Cruciate Ligament (PCL), Medial Collateral Ligament (MCL), Lateral Collateral Ligament (LCL), and Medial Patellofemoral Ligament (MPFL) reconstruction.

Device Story

AperFix Femoral Implant is a non-absorbable internal fixation device for tendon-to-bone attachment. Device consists of body, wedge, two deployable arms, and compression pads; pre-loaded on sterile inserter. Used by surgeons in OR during ligament reconstruction (ACL, PCL, MCL, LCL, MPFL). Implant provides mechanical fixation of soft tissue grafts to bone. Single-use device. Benefits include secure graft fixation during healing. Mechanical testing demonstrates superior pull-out strength compared to predicate.

Clinical Evidence

Bench testing only. Mechanical testing performed to evaluate pull-out strength; results showed pull-out strength is significantly higher than that of a predicate device.

Technological Characteristics

Non-absorbable internal fixation device. Components: body, wedge, two deployable arms, compression pads. Diameters: 9, 10, 11 mm; lengths: 29, 35 mm. Sterilization: Co gamma irradiation (25-40 kGy). Single-use. Mechanical fixation principle.

Indications for Use

Indicated for patients undergoing tenodesis procedures with soft tissue grafts for ACL, PCL, MCL, LCL, or MPFL reconstruction via arthroscopic or open surgical techniques.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K083612 (pg. 1 510(k) Summary AperFix® Femoral Implant with Inserter # 510(k) Summary # Cayenne Medical, Inc. AperFix Femoral Implant with Inserter ## ADMINISTRATIVE INFORMATION Manufacturer Name: Cayenne Medical, Inc. 16597 N. 92nd St., Suite 101 Scottsdale, AZ 85260 Telephone (480) 520-3661 FAX (480) 520-3670 Official Contact: Kereshmeh Shahriari DEVICE NAME Classification Names: Trade/Proprietary Name: Screw, fixation, bone AperFix® Femoral Implant with Inserter Common Name: Bone screw # DEVICE CLASSIFICATION FDA has classified bone screws as Class II devices (21 CFR 888.3040). The product code for screw, fixation, bone is HWC. These devices are reviewed by the Orthopedic Joint Devices Branch ### INTENDED USE The AperFix® Femoral Implant with Inserter is intended for use in tenodesis procedures with soft tissue grafts, utilizing either arthroscopic or open techniques during Anterior Cruciate Ligament (ACL), Posterior Cruciate Ligament (PCL), Medial Collateral Ligament (MCL), Lateral Collateral Ligament (LCL), and Medial Patellofemoral Ligament (MPFL) reconstruction. Revised Page 17 of 316 {1}------------------------------------------------ 510(k) Summary AperFix® Femoral Implant with Inserter #### DEVICE DESCRIPTION The Cayenne Medical AperFix® Femoral Implant with Inserter is a non-absorbable internal fixation device intended to provide tendon to bone fixation in arthroscopic or open procedures such as Anterior Cruciate Ligament (ACL), Posterior Cruciate Ligament (PCL), Medial Collateral Ligament (MCL), Lateral Collatcral Ligament (LCL), and Medial Patellofemoral Ligament (MPFL) reconstruction. It includes a body, wedge, two deployable arms, and compression pads. AperFix Femoral Implants are available in diameters of 9, 10 and 11 mm with two different lengths of 29 mm and 35 mm. Each device is intended for single use and is pre-loaded on a sterile inserter. The AperFix Femoral Implant with Inserter will be supplied sterile. Sterilization is accomplished by means of Co gamma irradiation at a dose of 25-40 kGy. Mechanical testing was performed on the AperFix Femoral Implant. It was shown that pull-out strength is significantly higher than that of a predicate device. #### EQUIVALENCE TO MARKETED PRODUCT Cayenne Medical, Inc. demonstrated that, for the purposes of FDA's regulation of medical devices, the AperFix Femoral Implant with Inserter is substantially equivalent in indications and design principles to predicate devices, each of which has been determined by FDA to be substantially equivalent to preamendment devices. Page 18 of 316 {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the eagle. The eagle is black, and the text is also black. Public Health Service MAR - 4 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Cayenne Medical, Inc. % Ms. Kereshmeh Shahriari 16597 N. 92md St. Suite 101 Scottsdale, Arizona 85260 Re: K083612 Trade/Device Name: AperFix® Femoral Implant with Inserter Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: December 3, 2008 Received: December 5, 2008 Dear Ms. Shahriari: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ## Page 2 -- Ms. Kereshmeh Shahriari This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark H. Milburn Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: AperFix® Femoral Implant with Inserter Indications for Use: The AperFix Femoral Implant with Inserter is intended for use in tenodesis procedures with soft tissue grafts, utilizing cither arthroscopic or open techniques during Anterior Cruciate Ligament (ACL), Posterior Cruciate Ligament (PCL), Medial Collateral Ligament (MCL), Lateral Collateral Ligament (LCL), and Medial Patellofemoral Ligament (MPFL) reconstruction. Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative and Neurological Devices Revised Page 15 of 316 510(k) Number/14083617
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...