← Product Code [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC) · K080649

# POGO SCREW (K080649)

_Fxdevices · HWC · Nov 25, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K080649

## Device Facts

- **Applicant:** Fxdevices
- **Product Code:** [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC.md)
- **Decision Date:** Nov 25, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The POGO Screw is indicated for use in the general management of fractures and reconstructive surgery.

## Device Story

POGO Screw is a cannulated metallic bone fixation fastener used for compression fixation between bone fragments. Device is manufactured from 316 surgical stainless steel; available in lengths from 55mm to 130mm. Provided sterile or non-sterile for user sterilization. Used by orthopedic surgeons in clinical settings for fracture management and reconstructive procedures. Provides mechanical stabilization of bone segments to facilitate healing.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Material: 316 surgical stainless steel. Form factor: Cannulated screw, 55mm-130mm length. Principle: Mechanical compression fixation. Sterilization: Provided sterile or user-sterilized.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Smith and Nephew Cannulated Screws ([K060736](/device/K060736.md))
- Stryker Medical Asnis III ([K071092](/device/K071092.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

10/9/2008

## 510K Summary of Safety and Effectiveness Fx Devices POGO Screw Oct 10. 2008

- 1. Sponsor Name FxDevices One South Ocean Blvd., Suite 324 Boca Raton, FL 33432
NOV 2 5 2008

| 2.                  | Device Name:                                       | POGO Screw |
|---------------------|----------------------------------------------------|------------|
| Panel               | Orthopaedic                                        |            |
| Classification Name | Smooth or Threaded Metallic Bone Fixation Fastener |            |
| CFR Number          | Class II (per 21 CFR 888.3040)                     |            |
| Product Code        | HWC                                                |            |

- 3. Identification of Predicate or Legally Marketed Device The POGO Screw is substantially equivalent to the Smith and Nephew Cannulated Screws cleared under K060736 and the Stryker Medical Asnis III cleared under K071092.
- 4. Device Description

The POGO Screws are comprised of various size cannulated screws for the fixation of bone fractures. The components are manufactured from 316 surgical stainless steel. The screws are available in various sizes from 55mm to 130mm in length. They are provided sterile and also non sterile to be sterilized by the user prior to use.

- 5. Intended Use The POGO Screw is indicated for use in the general management of fractures and reconstructive surgery.
- 6. Comparison of Technological Characteristics The POGO Screw and the predicate device accomplish the same function of providing compression fixation between a base bone and a bone fragment. Both devices accommodate a range of total lengths within each product design.
- 7 Performance Testing Bench testing was conducted to support equivalency
- 8 Statement of Equivalency The POGO Screw is substantially equivalent in design, materials, construction and intended use as those of the predicate. Since the POGO Screw is the same in intended use and technological characteristics as the predicate devices, the POGO Screw does not raise any new safety and efficacy concerns when compared to these similar legally marketed devices.

{1}------------------------------------------------

FxDevices Premarket Notification -- POGO Screw 10/9/2008

The test results demonstrate that the POGO Screw is substantially equivalent to the predicate device and is capable of safely and accurately performing the stated intended use.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "Public" and "Health" on the first line and "Service" on the second line. The text appears to be part of a document or sign.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FxDevices, Inc. % Mr. Rich Lipschutz Operations Manager One South Ocean Blvd., Suite 324 Boca Raton, Florida 33432

NOV 2 5 2008

Re: K080649

Trade/Device Name: POGO Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: October 10, 2008 Received: October 16, 2008

Dear Mr. Lipschutz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Image /page/2/Picture/11 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle symbol on the right and text arranged in a circle on the left. The text reads "DEPARTMENT OF HEALTH * HUMAN SERVICES * USA" around the circle.

{3}------------------------------------------------

Page 2 - Mr. Rich Lipschutz

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M. Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

FxDevices Premarket Notification - POGO Screw

## Indications for Use

510(k) Number (if known): Ko80649

Device Name:

Indications For Use:

The POGO Screw is indicated for use in the general management of fractures and reconstructive surgery.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark A. McKerss

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number

Page 1 of

Kool

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K080649](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K080649)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).