SPIRALOK ANCHOR

K060271 · Depuy Mitek, A Johnson & Johnson Company · HWC · May 4, 2006 · Orthopedic

Device Facts

Record IDK060271
Device NameSPIRALOK ANCHOR
ApplicantDepuy Mitek, A Johnson & Johnson Company
Product CodeHWC · Orthopedic
Decision DateMay 4, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SpiraLok Anchor is intended for: Shoulder: Rotor Cuff Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair; Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair; Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis; Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

Device Story

SpiraLok Anchor is a PLA threaded suture anchor preloaded on a disposable inserter assembly. Device facilitates fixation of two strands of suture to bone to reattach soft tissue. Suture options include Ethibond (non-absorbable), Panacryl (absorbable), and Orthocord (composite), potentially with tapered needles. Used by surgeons in clinical settings to secure soft tissue to bone, allowing natural healing. Device provides mechanical fixation; suture holds tissue in place during the healing process.

Clinical Evidence

Bench testing only.

Technological Characteristics

Material: PLA (polylactic acid). Design: Threaded suture anchor. Delivery: Preloaded on a disposable inserter assembly. Function: Mechanical fixation of soft tissue to bone via suture. Sterilization: Not specified.

Indications for Use

Indicated for soft tissue-to-bone fixation in shoulder (rotator cuff, biceps, AC joint), foot/ankle (stabilization, Achilles), knee (collateral ligaments, IT band), and elbow (biceps, collateral ligaments) repairs.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY - 4 (2005 # K060271 SECTION 6 – 510(k) SUMMARY ### Page 1 of 2 | Submitter's Name and Address: | DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive,<br>Raynham, MA 02767 USA | |-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Ruth Forstadt, RAC<br>Project Management Lead, Regulatory Affairs<br>DePuy Mitek<br>a Johnson & Johnson company<br>325 Paramount Drive,<br>Raynham, MA 02767 USA<br>Telephone: (508) 977-3988<br>Facsimile: (508) 828-3750<br>e-mail: rforstad@dpyus.jnj.comT | | Name of Medical Device | UUUUUUUUUUUUDevice Regulation:<br>Fastener, Fixation, Biodegradable, Soft Tissue<br>(21 CFR 888.3030)<br>Product code: 87 MAI<br><br>Common/Usual Name:<br>Biodegradable Fixation Fastener<br><br>Proprietary Name:<br>SpiraLok Anchor | | Device Classification | In accordance with per 21 CFR 888.3030 suture anchors are classified<br>by the FDA as Class II Medical Devices. | | Indications for Use | The SpiraLok Anchor is intended for: Shoulder: Rotor Cuff Repair,<br>Biceps Tenodesis, Acromio-Clavicular Separation Repair; Foot/Ankle:<br>Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair;<br>Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament<br>Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis;<br>Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral<br>Ligament Reconstruction. | | Device Description | The SpiraLok Anchor is a PLA threaded suture anchor preloaded on a<br>disposable inserter assembly intended for fixation of two strands of<br>suture to bone. Ethibond non-absorbable suture, the Panacryl<br>absorbable suture and the Orthocord composite suture options may<br>include tapered needles to facilitate suture passage through tissue. The<br>attached suture is then used to reattach soft tissue back to bone where it<br>reconnects through the healing process. | | Substantial Equivalence | The SpiraLok Anchor is a commercially marketed device that was<br>subject of K041069 (cleared November 9, 2004). When used for the<br>proposed indications the device is substantially equivalent to the<br>Arthrex Bio-Corkscrew Suture Anchor (K003227, cleared 1/8/2001). | | Safety and Performance | The determination of substantial equivalence for this device was based<br>on a detailed device description. Non-clinical laboratory testing was<br>performed demonstrating that the device is safe and can be considered<br>substantially equivalent to the predicate Arthrex Bio-Corkscrew Suture<br>Anchor for the proposed new intended uses. | {1}------------------------------------------------ ## K06 0271 510(k) SUMMARY ### Page 2 of 2 {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle. The logo is simple and monochromatic. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 . . . . . DePuy Mitek c/o Ms. Ruth Forstadt, RAC Project Management Lead, Regulatory Affairs 325 Paramount Drive Raynham, Massachusetts 02767 MAY - 4 2006 Re: K060271 Trade/Device Name: SpiraLok Absorbable Threaded Suture Anchor Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HWC, MAI Dated: January 31, 2006 Received: February 1, 2006 Dear Ms. Forstadt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 -- Ms. Ruth Forstadt, RAC forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html . Sincerely yours, Hukut Lemevios Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use 510(k) Number (if known): K060271 Device Name:___SpiraLok Absorbable Threaded Suture Anchor Indications For Use: The Spiralok Anchor is intended for: - Shoulder: Rotator Cuff Repair, Biceps Tenodesis, Acromio-Clavicular Separation . Repair - Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair . - Knee: Medial Collateral Ligamet Repair, Lateral Collateral Ligament Repair, . Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis - Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament . Reconstruction. Prescription Use 7 (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Helbert Lerner (Divisio Division of General, Restorative, and Neurological Devices Page 1 of 1 **510(k) Number** K060271
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...