← Product Code [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC) · K031751

# RAPIDFIRE SELF DRILLING SCREWS (K031751)

_Walter Lorenz Surgical, Inc. · HWC · Jun 26, 2003 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K031751

## Device Facts

- **Applicant:** Walter Lorenz Surgical, Inc.
- **Product Code:** [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC.md)
- **Decision Date:** Jun 26, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

Internal fixation devices to aid the surgeon in the stabilization of oral craniomaxillofacial skeletal bone, small bones of the hand, feet, wrist, ankles, fingers and toes, for the fixation of fractures, osteotomies, revision procedures, joint fusion and reconstructive procedures.

## Device Story

Rapidfire Self Drilling Screws are metallic bone fixation fasteners; 1.5 mm to 2.0 mm diameter; thread lengths up to 7 mm; center or cross drive head. Designed for self-drilling into bone without mandatory predrilling. System includes disposable cartridge containing five screws and disposable screwdriver blades. Used by surgeons in clinical settings for stabilization of craniomaxillofacial and small extremity bones. Cartridges delivered sterile; loaded into driver system for patient application. Provides mechanical fixation to support bone healing in fractures, osteotomies, and reconstructive procedures.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Metallic bone fixation screws; 1.5-2.0 mm diameter; up to 7 mm thread length; center or cross drive head; self-drilling design; sterile, single-use disposable cartridge system.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Lorenz Self Drilling Screws ([K013954](/device/K013954.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 26 2003

031751
page 1 of 2

## Summary of Safety and Effectiveness

| Device Name:         | Lorenz Self Drilling Screw |
|----------------------|----------------------------|
| Classification Name: | Screw, Fixation, Bone      |
| Device Product Code: | 87HWC (21 CFR 888.3040)    |

### Intended Use:

Internal fixation devices to aid the surgeon in the stabilization of oral craniomaxillofacial skeletal bone, small bones of the hand, feet, wrist, ankles, fingers and toes, for the fixation of fractures, osteotomies, revision procedures, joint fusion and reconstructive procedures.

### Description:

The self drilling screws are 1.5 mm to 2.0 mm in diameter and the thread lengths may range up to 7mm. The head may be center drive or cross drive. The screw is designed so that a predrilled hole is not required, but may be used. Five screws are placed within a disposable cartridge A disposable screwdriver blade is loaded between each screw. The cartridges are delivered to the user sterile. The cartridge is loaded into a driver blade system for delivery to the patient.

### Sterility Information:

The Rapidfire Self Drilling Screws will be marketed as sterile, single use devices.

### Substantial Equivalence:

Walter Lorenz considers the Rapidfire Self Drilling Screws equivalent to Lorenz Self Drilling Screws (K013954).

### Possible risks:

- 1. Poor bone formation, Osteoporosis, Osteomyelitis, inhibited revascularization, or infection can cause loosening of the device.
- 2. Nonunion or delayed union which may lead to breakage of the implant.
- 3. Migration, bending, fracture or loosening of the implant
- 4. Metal sensitivity, or allergic reaction to a foreign body.
- 5. Decrease in bone density due to stress shielding.
- 6. Pain, discomfort, abnormal sensation, or palpability due to the presence of the device.
- 7. Increased fibrous tissue response around the fracture site and/or the implant.
- 8. Necrosis of bone.

{1}------------------------------------------------

9. Inadequate healing.

31751
page 2 of 2

Apart from these adverse effects there are always possible complications of any surgical procedure such as, but not limited to, infection, nerve damage, and pain which may not be related to the implant.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a symbol of three stylized human figures, represented by curved lines, facing to the right.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 6 2003

Ms. Kim Reed Senior Regulatory Specialist Walter Lorenz Surgical, Inc. 1520 Tradeport Drive Jacksonville, Florida 32218

Re: K031751

Trade/Device Name: Rapidfire Self Drilling Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: HWC Dated: June 4, 2003 Received: June 6, 2003

Dear Ms. Reed:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good many:facturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Kim Reed

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
R. Mark N. Milliken

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure ·

{4}------------------------------------------------

### STATEMENT OF INDICATIONS FOR USE

510(k) Number: K031751

Device Name: Rapidfire self drilling screws

Indications For Use:

Internal fixation devices to aid the surgeon in the stabilization of oral craniomaxillofacial skeletal bone, small bones of the hand, feet, wrist, ankles, fingers and toes, for the fixation of fractures, osteotomies, revision procedures, joint fusion and reconstructive procedures.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter-Use (Optional Format 1-2-96)

Mark A. Milliman

Division of General, Restorative and Neurological Devices

510(k) Number -

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K031751](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K031751)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
