TITANIUM COMPRESSION ANCHOR SYSTEM MODEL VERSION 2

K012280 · Triage Medical, Inc. · HWC · Aug 17, 2001 · Orthopedic

Device Facts

Record IDK012280
Device NameTITANIUM COMPRESSION ANCHOR SYSTEM MODEL VERSION 2
ApplicantTriage Medical, Inc.
Product CodeHWC · Orthopedic
Decision DateAug 17, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is used for the correction of hallux valgus (bunion) and repair of metatarsal, metacarpal and phalangeal fractures.

Device Story

Titanium Compression Anchor System; orthopedic fixation device. Used for hallux valgus correction and fracture repair in metatarsal, metacarpal, and phalangeal bones. Implanted by surgeons in clinical/OR settings. Provides mechanical stabilization of bone segments to facilitate healing. Device design utilizes compression to maintain bone apposition.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Titanium construction; mechanical compression anchor system; intended for orthopedic fixation.

Indications for Use

Indicated for patients requiring correction of hallux valgus (bunion) or repair of metatarsal, metacarpal, and phalangeal fractures.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. Public Health Service AUG 1 7 2001 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Triage Medical, Inc. c/o Mr. Greg Holland Regulatory Specialists, Inc. 3722 Ave. Sausalito Irvine, California 92606 Re: K012280 Trade Name: Titanium Compression Anchor System Regulation Number: 888.3040 Regulatory Class: II Product Code: HWC Dated: July 19, 2001 Received: July 19, 2001 Dear Mr. Holland: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1}------------------------------------------------ Page 2 - Mr. Greg Holland This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely vours. Mark Mckesson Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Devices Evaluation Center for Devices and Radiological Devices Enclosure {2}------------------------------------------------ Indications for Use Statement Page_1_of_1 510(k) Number (if known): KO12280 Device Name: Titanium Compression Anchor System Indications For Use: This device is used for the correction of hallux valgus (bunion) and repair of metatarsal, metacarpal and phalangeal fractures. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) for Mark H. Millenson (Division Sign-Off) Division of General, Restorative and Neurological Devices Prescripti0(k) Number (Per 21 CFR 801.109) K012280 OR Over-The-Counter Use No (Optional Format 1-2-96) REGULATORY SPECIALISTS
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...