K191150 · Bodycad Laboratories, Inc. · HSX · Oct 8, 2019 · Orthopedic
Device Facts
Record ID
K191150
Device Name
Bodycad Unicompartmental Knee System
Applicant
Bodycad Laboratories, Inc.
Product Code
HSX · Orthopedic
Decision Date
Oct 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3520
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The patient-specific Bodycad Unicompartmental Knee System (Bodycad UKS) is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with: · joint impairment due to osteoarthritis or traumatic arthritis of the knee. · varus deformity of the knee, and · as an alternative to tibial osteotomy in patients with unicompartmental OA. The patient-specific Bodycad UKS components fit within an envelope of dimensions that are specific to each patient. The Bodycad UKS femoral component and tibial baseplate are intended fixation. A Bodycad screw must be used for fixation of the femoral component.
Device Story
Patient-specific unicompartmental knee prosthesis; replaces medial tibiofemoral compartment. Inputs: patient MRI/CT images; surgeon input. Processing: Bodycad software generates CAD/CAM files for custom femoral/tibial components and single-use cutting guides. Output: patient-specific implants and surgical guides. Used in clinical setting by orthopedic surgeons. Implants require cemented fixation; femoral component requires Bodycad screw. System provides personalized fit to patient anatomy; intended to restore joint function and alleviate pain from osteoarthritis or traumatic arthritis.
Clinical Evidence
Bench testing only. Performed fatigue strength testing per ASTM F3140-17 and ASTM F1800-12, insert dislodgement resistance, and locking pin pull-out force (pre/post-fatigue). Surgeon-user validation conducted via simulated surgeries.
Technological Characteristics
Materials: CoCrMo (ASTM F1537-11) for femoral component/screw; Ti6Al4V ELI (ASTM F136-13) for tibial baseplate/locking pin/screw; UHMWPE (ASTM F648-14) for tibial insert. Patient-specific geometry derived from CAD/CAM. Cemented fixation. Gamma radiation sterilization.
Indications for Use
Indicated for unicompartmental knee arthroplasty in patients with advanced medial compartment knee osteoarthritis, traumatic arthritis, or varus deformity requiring joint replacement, provided adequate healthy bone exists to support implants.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two main elements: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the acronym "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" stacked in two lines.
October 8, 2019
% Robert Poggie President BioVera Inc. 65 Promenade Saint Louis Notre-Dame-de-l'Ile-Perrot, QC, J7V 7P2 Canada
Re: K191150
Trade/Device Name: Bodycad Unicompartmental Knee System Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX Dated: September 4, 2019 Received: September 9, 2019
Dear Robert Poggie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Peter G. Peter G. Allen -S Date: 2019.10.08 Allen -S 11:44:53 -04'00'
For Ting Song Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K191150
Device Name
Bodycad Unicompartmental Knee System
#### Indications for Use (Describe)
The patient-specific Bodycad Unicompartmental Knee System (Bodycad UKS) is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with:
· joint impairment due to osteoarthritis or traumatic arthritis of the knee.
· varus deformity of the knee, and
· as an alternative to tibial osteotomy in patients with unicompartmental OA.
The patient-specific Bodycad UKS components fit within an envelope of dimensions that are specific to each patient. The Bodycad UKS femoral component and tibial baseplate are intended fixation. A Bodycad screw must be used for fixation of the femoral component.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
### Bodycad Unicompartmental Knee System
In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness of the Bodycad Unicompartmental Knee System.
# A. SUBMITTERS INFORMATION
| Submitter Name: | BioVera, Inc. |
|---------------------|----------------------------------------------------------------------------------|
| Submitter Address: | 65 Promenade Saint-Louis, Notre-Dame-De-L'Ile-Perrot, Quebec,<br>J7V 7P2, CANADA |
| Contact Person: | Robert A Poggie, PhD |
| Phone Number: | (514) 901-0796 |
| Fax Number: | (514) 901-0796 |
| Date of Submission: | April 29, 2019 |
# B. DEVICE IDENTIFICATION & MANUFACTURER
| Manufacturer Name: | Bodycad Laboratories Inc. |
|-----------------------|-----------------------------------------------------------------|
| Manufacturer Address: | 2035 rue du Haut-Bord, Quebec, Quebec, G1N 4R7, Canada |
| Registration Number: | 3012086398 |
| Contact Name: | Guy Sevigny |
| Title: | Director, Regulatory Affairs |
| Device Trade Name: | Bodycad Unicompartmental Knee System |
| Device Common Name: | Unicondylar knee device |
| Classification Name: | knee joint femorotibial metal/polymer non-constrained, cemented |
| Classification Code: | HSX - Class II |
| Classification Panel: | Orthopedic |
| Regulation Number: | 21 CFR section 888.3520 |
# C1. PREDICATE DEVICE
| K163700 | Bodycad Unicompartmental Knee System (BUKS) |
|---------|---------------------------------------------|
| K181302 | Bodycad Unicompartmental Knee System (BUKS) |
| C2. REFERENCE DEVICE | |
|----------------------|-------------------------------------------|
| K033363 | Zimmer Unicompartmental Knee System (ZUK) |
{4}------------------------------------------------
### D. DEVICE DESCRIPTION
The Bodycad Unicompartmental Knee System (BUKS) is a patient-specific prosthesis that consists of femoral and tibial implants for replacement of the medial tibiofemoral compartment of the knee. The patient-specific femoral and tibial components and single-use cutting quides are manufactured from CAD and CAM files generated from Bodycad software, which are based on MRI or CT images of the patient's knee and input from the surgeon. The BUKS is for cemented use only and is sterilized by gamma radiation.
The subject of this 510(k) notification is a line extension and design update that includes a new tibial baseplate that is compatible with the BUKS system (K163700, K181302) and instruments that are compatible with the new tibial baseplate.
Materials: Wrought Cobalt-28Chromium-6Molybdenum Alloy (CoCrMo; ASTM F1537-11) for the femoral component and screw, wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (Ti6Al4V ELI; ASTM F136-13) for the tibial baseplate, locking pin, and screw, Ultra-High-Molecular-Weight Polyethylene (UHMWPE; F648-14) for the tibial insert.
# E. INTENDED USE
The patient-specific Bodycad Unicompartmental Knee System (Bodycad UKS) is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with:
- joint impairment due to osteoarthritis or traumatic arthritis of the knee,
- varus deformity of the knee, and ●
- as an alternative to tibial osteotomy in patients with unicompartmental OA. .
The patient-specific Bodycad UKS components fit within an envelope of dimensions that are specific to each patient. The Bodycad UKS femoral component and tibial baseplate are intended for cemented fixation. A Bodycad screw must be used for fixation of the femoral component.
# F. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The BUKS described in this 510(k) notification has the same intended use and general technological features as the predicate devices cleared by FDA in K163700 and K181302 (BUKS). The design modifications to the tibial baseplate cleared in K163700 include: Reduced thickness of the base plate (UHMWPE inset sliding and upper rails, cement pocket, and baseplate), reduced diameter of the locking pin and corresponding reduction of the diameter of hole for locking pin, increased diameter and redesign of fixation posts, addition of a medial fixation fin, and elimination of the flange and screw-hole for ancillary fixation. Other design changes relative to BUKS components cleared in K163700 is a single-use instrument for preparation of the holes and slot for the new fixation posts and fin. Comparison of the design features of the new BUKS tibial baseplate with those of the original BUKS baseplate (K163700, K181302) and the ZUK reference device, and results of performance testing of the subject tibial base plate and tibial insert assembly have established substantial equivalence of the subject, predicate, and reference devices.
{5}------------------------------------------------
#### G. PERFORMANCE DATA
The following tests were performed for establishing substantial equivalence of subject, predicate, and reference devices:
- . Tibial baseplate fatigue strength per 3-point bend method as described in ASTM F3140-17, Standard Test Method for Cyclic Fatique Testing of Metal Tibial Tray Components of Unicondylar Knee Joint Replacements, and ASTM F1800-12, Standard Practice for Cyclic Fatigue Testing of Metal Tibial Tray Components of Total Knee Joint Replacements.
- Resistance to dislodgement of the polyethylene insert in both the as-manufactured and . post cyclic fatigue conditions.
- . Pull-out force for the locking pin at time-zero, as manufactured, and post cyclic fatigue.
- Surgeon-user validation, simulated surgeries with the new tibial baseplate, locking pin, . insert, and associated instruments.
The results of performance testing demonstrated substantial equivalence of the subject, predicate, and reference devices.
#### H. CONCLUSION
The subject components of this line extension to the BUKS are substantially equivalent and share the same intended use as the predicate BUKS cleared in K163700 and K181302, and the reference device (K033363).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.