← Product Code [HSX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX) · K170344

# Tahoe Uni Knee System (K170344)

_Consensus Orthopedics, Inc. · HSX · Oct 5, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX/K170344

## Device Facts

- **Applicant:** Consensus Orthopedics, Inc.
- **Product Code:** [HSX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX.md)
- **Decision Date:** Oct 5, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3520
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Tahoe Unicompartmental Knee System (TUKS) is designed as a system and is not intended for substitution of components from other systems. The indications for use are as follows: Primary medial or lateral compartmental intervention of (1) primary non-inflammatory degenerative disease, including osteoarthritis, traumatic arthritis, or osteonecrosis; (2) posttraumatic degenerative disease; (3) varus or valgus deformities; and (4) damage due to previous surgical intervention when the opposite compartment is preserved and when the anterior cruciate, posterior cruciate, medial collateral, and lateral collateral ligaments are present and functional. All TUKS implants are single use only, and are intended for implantation only with bone cement.

## Device Story

Fixed-bearing, round-on-flat unicompartmental knee system; restores function in medial or lateral tibial compartments. Components: cast CoCr femoral component; wrought CoCr tibial baseplate; UHMWPE or Vitamin-E polyethylene tibial insert. Insert snaps into baseplate via anterior-posterior locking mechanism. Used in orthopedic surgery; implanted by surgeons using bone cement. System design prohibits component substitution from other systems. Provides mechanical joint replacement; improves patient mobility; reduces pain associated with degenerative knee conditions.

## Clinical Evidence

Bench testing only. Included FEA simulations for femoral component backside geometry, 3-point bending fatigue testing of tibial baseplate (per WK45235), and tibiofemoral contact area/pressure testing (per ASTM F2083-12). Locking mechanism strength and manual insertion verified. Pyrogenicity testing confirmed <20 Endotoxin Units per construct. Wear testing waived based on poly thickness (6mm) meeting ISO 21536 requirements.

## Technological Characteristics

Materials: Cast CoCr alloy (ASTM F75), wrought CoCr alloy (ASTM F1537), UHMWPE (ASTM F648), Vitamin-E polyethylene (ASTM F2695). Design: Fixed-bearing, round-on-flat, unicompartmental, cemented. Locking mechanism: Anterior-posterior snap-fit. Standards: ASTM F2083-12 (contact), ISO 21536 (poly thickness), WK45235 (fatigue).

## Regulatory Identification

A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Uniglide Femoral component ([K050764](/device/K050764.md))
- Repicci II ([K971938](/device/K971938.md))
- Vanguard M Series ([K021621](/device/K021621.md), [K042093](/device/K042093.md))
- EPIK Metal-backed tibial component ([K022437](/device/K022437.md), [K020741](/device/K020741.md))
- Stride Metal-backed tibial component ([K123380](/device/K123380.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Consensus Orthopedics, Inc Zac Johnson Regulatory Affairs Specialist 1115 Windfield Wav El Dorado Hills, California 95762 October 5, 2017

### Re: K170344

Trade/Device Name: Tahoe Uni Knee System Regulation Number: 21 CFR 888.3520 Regulation Name: Knee joint femorotibial metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: HSX Dated: September 7, 2017 Received: September 7, 2017

Dear Mr. Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

### Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# UDICATIONS FOO USE STATEN

### 510(k) Number: K170344

Device Name: Tahoe Uni Knee Syste

## INDICATIONS AND USAGE

:swollot se are asu rof suoitesibui ad T :smetsys redto mort strenodimos i və qalında və bəbinə qalınmış və sə bənşisən si (SXUT) mətriy əsərlərin əsasən inil ədədi ədə

collaters), and lateral collateral liigaments are pressut and functional. ibem , and services on is not is no the not no the not member of institution is also do นอนพ แอฟเซอร์เท่ ไธวเป็นทรางงาน อา อนุม อริธเทพ (b) but : รอยู่ใหม่เข้าอุประเทศ เป็นเทพ (c) :อรออรib ຣາອາອົງອີງ ການປະກວດ (S) ; ເຂດາວອາດອາຣາ ເດີຍນາ ເປັນນາງງາມ ເປັນເປັນເອກະສາປີ ເ əvizesən vəzirlərindən yaranmışdır. İstinadlar Həmçinin baxında baxında bir və bir və bir və bir və bir və bir və

unun ən qalını və qalında və başların ərə bura şirkət ərazi və bir sənələri ilə bir və bir sənələri ilə bir və

IZ) and and and and and and and and

IO/ANA

Over the Counter Use CFR Part 801 Suppart

аасаан эг ээлч язнтома по эпитиог гэмт гэнт молав этгийн том оа

וחבות של CDPH, Office of Device Evaluation (ODF

{3}------------------------------------------------

### 2. 510(k) SUMMARY

| Sponsor Name:        | Consensus Orthopedics, Inc.<br>1115 Windfield Way, Suite 100<br>El Dorado Hills, CA 95762                                         |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Contact:      | Name: Zac Johnson<br>Phone: (916) 355-7154<br>Fax: (916) 355-7190<br>Email: zjohnson@consensusortho.com                           |
| Date Prepared:       | 01 February 2017                                                                                                                  |
| Trade Name:          | Tahoe Uni Knee System                                                                                                             |
| Common Name:         | Unicompartmental Knee System                                                                                                      |
| Classification Name: | Knee joint femorotibial metal/polymer non-constrained cemented<br>prosthesis; Class II device; 21 CFR 888.3520; Product Code: HSX |

### Device Description:

The Tahoe Uni Knee System (TUKS) is a primary fixed-bearing, round-on-flat unicompartmental knee system offering flexibility to restore function in the affected medial or lateral tibial compartments. The femoral and tibial components are intended use only. The femoral component is made from cast CoCr alloy (ASTM F75), whereas the tibial baseplate is made from wrought CoCr alloy (ASTM F1537). The tibial insert is made from UHMWPE (ASTM F648) or Vitamin-E polyethylene (VitalitE, ASTM F2695). The tibial insert is designed to snap into the baseplate via an anterior-posterior locking mechanism.

### Indications for Use:

The Tahoe Unicompartmental Knee System (TUKS) is designed as a system and is not intended for substitution of components from other systems. The indications for use are as follows:

Primary medial or lateral compartmental intervention of (1) primary non-inflammatory degenerative disease, including osteoarthritis, traumatic arthritis, or osteonecrosis; (2) posttraumatic degenerative disease; (3) varus or valgus deformities; and (4) damage due to previous surgical intervention when the opposite compartment is preserved and when the anterior cruciate, posterior cruciate, medial collateral, and lateral collateral ligaments are present and functional.

All TUKS implants are single use only, and are intended for implantation only with bone cement.

{4}------------------------------------------------

### Substantial Equivalence:

### Technological Characteristics/ Substantial Equivalence:

The Tahoe Uni Knee System (TUKS) implant components are substantially equivalent to legally marketed devices regarding their indications, technology, and performance (Table 4.1). The TUKS is substantially equivalent to the following unicompartmental knee systems intended for cemented use:

- Uniglide Femoral component used with all-poly tibial component (K050764). ●
- Oxford Femoral component used with Repicci II (K971938) or Vanguard M Series ● (K021621, K042093) metal-backed tibial components.
- EPIK Metal-backed tibial component (K022437; K020741). ●
- Stride - Metal-backed tibial component (K123380).

### Non-Clinical Performance Data:

Bench testing, FEA simulations, and engineering justifications were carried out on TUKS implant components to ensure their safety and effectiveness for clinical use. The femoral component was analyzed via FEA simulation to ensure its backside geometry (tensile side) would not fail when deflecting the posterior facet away from the central peg. The tibial baseplate was fatigue tested in 3-point bending per WK45235 to ensure the tray region would not fail under the absence of bone support on the tibial plateau mid-section. Joint stability testing was deemed unnecessary due to the unconstrained round-on-flat articulation. Joint contact characteristics were tested per ASTM F2083-12 to ensure adequate tibiofemoral contact area and contact pressure distribution under physiologic load. The tibial insert locking mechanism was tested to ensure ease of manual insertion and adequate connection strength. Wear testing was deemed unnecessary due to meeting the minimum requirement for poly thickness in the load bearing region (6mm) per ISO 21536. Pyrogenicity testing was completed and was deemed to be within 20 Endotoxin Units per surgical construct per the FDA Guidance Document Pyrogen and Endotoxins Testing.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX/K170344](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX/K170344)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
