CONTOURED ARTICULAR PROSTHESIS (CAP) HUMERAL HEAD RESURFACING SYSTEM
K023096 · Std Mfg., Inc. · HSD · Apr 10, 2003 · Orthopedic
Device Facts
| Record ID | K023096 |
| Device Name | CONTOURED ARTICULAR PROSTHESIS (CAP) HUMERAL HEAD RESURFACING SYSTEM |
| Applicant | Std Mfg., Inc. |
| Product Code | HSD · Orthopedic |
| Decision Date | Apr 10, 2003 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3690 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Contoured Articular Prosthetic (CAP) System is intended for the reconstruction of painful and/or severly disabled shoulder joints resulting from post-traumatic degenerative disease or avascular necrosis. The humeral head and neck should be of sufficient bone stock to support loading. The rotator cuff should be intact or reconstructable. The device is a single use implant intended to be used with bone cement.
Device Story
The Contoured Articular Prosthetic (CAP) is a humeral head resurfacing system for shoulder joint reconstruction. It consists of a Ti-6Al-4V ELI alloy taper post and a Co-Cr-Mo alloy articular component with a CP Titanium coating. The surgeon uses a drill guide to establish a working axis, drills a pilot hole, and inserts the taper post. A contact probe measures the topography of the surrounding articular surface to select an appropriately sized articular component, which is then impacted onto the taper post. The device is used in a clinical setting by an orthopedic surgeon. By minimizing bone resection and allowing for precise sizing, the device aims to restore joint function and reduce pain, while preserving bone for potential future arthroplasty.
Clinical Evidence
bench testing only
Technological Characteristics
Components: Taper post (Ti-6Al-4V ELI, ASTM F136), Articular component (Co-Cr-Mo, ASTM F799/F1537) with CP Titanium coating. Mechanical fixation via taper interlock and bone cement. Dimensions: Taper post available in 25mm and 30mm lengths; articular components available in 0.5mm offset increments. Non-software device.
Indications for Use
Indicated for patients with painful and/or severely disabled shoulder joints due to post-traumatic degenerative disease or avascular necrosis. Requires sufficient humeral bone stock and intact or reconstructable rotator cuff. Contraindicated if bone stock is insufficient or rotator cuff is irreparable.
Regulatory Classification
Identification
A shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis is a device made of alloys, such as cobalt-chromium-molybdenum. It has an intramedullary stem and is intended to be implanted to replace the articular surface of the proximal end of the humerus and to be fixed without bone cement (§ 888.3027). This device is not intended for biological fixation.
Predicate Devices
- Copeland Resurfacing Heads (K003044, K010664)
- Buechel-Pappas Humeral Head Resurfacing Component (K992394)
Related Devices
- K031971 — GLOBAL CAP RESURFACING REPLACEMENT SHOULDER · DePuy Orthopaedics, Inc. · Sep 24, 2003
- K131298 — EXACTECH EQUINOXE RESURFACING HUMERAL HEAD SYSTEM · Exactech, Inc. · Nov 26, 2013
- K071064 — ASCENSION HUMERAL RESURFACING ARTHOPLASTY (HRA) SYSTEM TPS/HA · Ascension Orthopedics, Inc. · Jun 22, 2007
- K142942 — HemiCAP Humeral Head XL (HHXL) Articular Resurfacing System · Arthrosurface, Inc. · Dec 19, 2014
- K062861 — ASCENSION HUMERAL RESURFACING ARTHROPLASTY SYSTEM · Ascension Orthopedics, Inc. · Oct 31, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(K) Summary of Safety and Effectiveness
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the Contoured Articular Prosthetic (CAP) Humeral Head Resurfacing Prosthesis.
| Submitted By: | STD Manufacturing, Inc.<br>1063 Turnpike Street<br>Stoughton, MA 02072<br>(781) 828-4400 |
|----------------------|-----------------------------------------------------------------------------------------------|
| Date: | September 16, 2002 |
| Contact Person: | Steven W. Ek<br>VP Engineering |
| Proprietary Name: | Contoured Articular Prosthetic (CAP)<br>Humeral Head Resurfacing System |
| Common Name: | Hemi-shoulder resurfacing prosthesis |
| Classification Name: | Prosthesis, Shoulder, Hemi-, Humeral, Metallic<br>Orthopedic<br>21 CFR § 888.3690<br>Class II |
| Product Code: | HSD |
## Intended Use:
The Contoured Articular Prosthetic (CAP) System is intended for the reconstruction of painful and/or severly disabled shoulder joints resulting from post-traumatic degenerative disease or avascular necrosis. The humeral head and neck should be of sufficient bone stock to support loading. The rotator cuff should be intact or reconstructable. The device is a single use implant intended to be used with bone cement.
Reprinted Page 011
{1}------------------------------------------------
023096
p. 2/3
### Device Description:
The Contoured Articular Prosthetic (CAP) consists of two components, a taper post component and an articular component, that mate together via a taper interlock to provide stable and immobile fixation of the implant and stress bearing contact at the bone/ prosthetic interface.
The taper post component is manufactured of a Ti-6Al-4V ELI alloy per ASTM F136. The taper post has a tapering distal tip, a full-length cannulation, and a proximal female taper bore. The taper post is available in a 25mm and 30mm length to provide for variation in humeral head size.
The articular component is a dome shaped component manufactured of a Cobalt-Chromium-Molybdenum alloy per ASTM F799 and ASTM F1537. The articular component has a bone contact surface that is coated with a CP Titanium coating and a polished articular bearing surface.
Utilizing the drill guide provided within the CAP instrumentation set, a surgeon is able to define a working axis that is normal to the articular cartilage surface at the site of the defect. After drilling a pilot hole, the taper post is driven into place using the trial cap to ensure that the surface of the articular component will be tangent and congruent to the existing cartilage surface when seated. Using the contact probe instrument corresponding to the implant diameter, offset measurements are taken to define the topography of the patients surrounding articular surface by revolving the probe around a centering shaft coaxial to the working axis of the taper post. With these offset measurements, the surgeon is able to select the articular component from a range of sizes that will allow it to seat flush to the surrounding articular surface. Offset increments in .5mm sizes will allow for an optimal fit to the existing articular cartilage.
A reamer, which matches the articular component internal geometry is used to prepare the site for the prosthetic to be implanted. This allows for a precise fit of the implant to the prepared site and minimizes bone resection, so as to provide minimal impact to any future arthroplasty procedure. The articular component is then impacted to seat the taper interlock between the two components.
Reprinted Page 012
{2}------------------------------------------------
023096
## Substantial Equivalence Information:
The intended use, materials, and application of the candidate device are substantially equivalent to those of the following predicate devices:
Copeland Resurfacing Heads (K003044, K010664) Buechel-Pappas Humeral Head Resurfacing Component (K992394)
Potential risks associated with the candidate device are the same as with other ioint prosthetic devices. These include, but are not limited to:
- Reaction to the bone cement .
- Reaction to the implant materials .
- Nerve palsy ●
- Embolus
- Implant loosening/ migration
- Infection
- . Delayed wound healing
- Damage to the implants
- . Excessive wear
- Hematoma .
- Need for Revision
- Incomplete resolution of symptoms ●
The device is technically easy to implant, and offers the surgeon a high degree of precision and flexibility in sizing and fitting the articular component to the existing anatomy. A minimal amount of bone and articular cartilage resection is also offered by the device.
Additional materials, manufacturing, and performance data to support the safety and effectiveness of the CAP System are provided within this Premarket Notification.
> Revised Page 013 (Page 014 Omitted)
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the snake and staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 0 2003
Mr. Steven W. Ek Vice President of Engineering STD Manufacturing, Inc. 1063 Turnpike Street Stoughton, Massachusetts 02072
Re: K023096
Trade/Device Name: Contoured Articular Prosthetic (CAP) Humeral Head Resurfacing System Regulation Number: 21 CFR 888.3690 Regulation Name: Shoulder Joint Humeral (hemi-shoulder) Metallic Uncemented Prosthesis Regulatory Class: Class II Product Code: HSD Dated: January 9, 2003 Received: January 10, 2003
Dear Mr. Ek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Steven W. Ek
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Mark A Millhuser
Celia M. Witten, Ph.D., MD Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number: K023096
Device Name: Humeral Head Resurfacing Prosthesis
Indications for Use:
For the reconstruction of painful and/or severely disabled shoulder joints resulting from post-traumatic degenerative disease or avascular necrosis. The humeral head and neck should be of sufficient bone stock to support loading. The rotator cuff should be intact or reconstructable. The device is a single use implant intended to be used with bone cement.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109 Over-The Counter Use
(Optional Format 1-2-96)
Mark N. Milligan
Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K023096