← Product Code [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB) · K992350

# PYLON INTRAMEDULLAARY NAIL SYSTEM (K992350)

_Alphatec Mfg., Inc. · HSB · Sep 22, 1999 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K992350

## Device Facts

- **Applicant:** Alphatec Mfg., Inc.
- **Product Code:** [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB.md)
- **Decision Date:** Sep 22, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Pylon Intramedullary Nail System is indicated for internal fixation of tibial or femoral fractures including low subtrochanteric fractures, transverse fractures, oblique and spiral fractures, segmental fractures, comminuted fractures, high supracondylar fractures, fractures with bone loss, stabilization for bone transport or lengthening and shortening, open fractures, corrective osteotomies, pathologic fractures, pseudoarthrosis of the bone shaft, nonunions, malunions, metaphyseal junction fractures, and acute bone lengthening and shortening.

## Device Story

Pylon Intramedullary Nail System consists of titanium alloy (Ti-6Al-4V) intramedullary rods and bone screws designed for internal fixation of tibial or femoral fractures. Rods are available in solid or cannulated constructs with varying diameters and lengths; proximal and distal holes allow for bone screw fixation. An end cap is provided to prevent bone ingrowth into the proximal end, facilitating future extraction. The system is used by surgeons in a clinical/surgical setting to stabilize bone fractures, nonunions, or malunions, and to support bone transport or lengthening procedures. By providing rigid internal fixation, the device maintains bone alignment, promoting healing and restoring structural integrity to the affected limb.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Constructed from titanium alloy (Ti-6Al-4V). System includes cylindrically shaped intramedullary rods (solid or cannulated) and bone screws. Features proximal/distal fixation holes and proximal end caps. Mechanical fixation device; no energy source or software components.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- ACE AIM Titanium Nails (Depuy ACE)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a logo for a company called "ALPHATEC Manufacturing, Inc." The logo consists of a circle with the Greek letter alpha inside it. Below the circle, the company name is printed in a bold, sans-serif font. The text is black and the background is white.

Premarket Notification slon Intramedullary Nail System Iphatec Manufacturing, Inc.

# 510(k) Summary

**510(k) Number**

510(k) Number K992350

### Manufacturer Identification

| Submitted By:          | Alphatec Manufacturing, Inc.<br>42-160 State Street<br>Palm Desert, CA 92211<br>phone: (760) 779-8250<br>fax: (760) 776-5527 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:        | Jason Blain<br>Manager of Product Development                                                                                |
| Date Summary Prepared: | July 12, 1999                                                                                                                |
| Device Identification  |                                                                                                                              |
| Proprietary Name:      | Pylon Intramedullary Nail System                                                                                             |
| Common Name:           | Intramedullary Fixation System                                                                                               |
| Classification:        | 21 CFR 888.3020: Rod, Fixation, Intramedullary and Accessories                                                               |

#### Device Description

The Pylon Intramedullary Nail System consists of intramedullary rods and screws for fixation inside the canal of the tibia or femur. The rods are cylindrically shaped and either made of a solid or cannulated construct, depending on the outer diameter of the rod. Rods are available in a variety of diameters and lengths and have holes located in the proximal and distal ends for fixation to bone by means of bone screws. Bone screws are also available in a multitude of diameters and lengths. An end cap is available that screws into the proximal end of the nails to prevent bone ingrowth which may hamper the attachment of extraction instrumentation.

All components of the Pylon Intramedullary Nail System are manufactured from titanium alloy (Ti-6Al-4V).

### Intended Use of the Device

The Pylon Intramedullary Nail System is indicated for internal fixation of tibial or femoral fractures including low subtrochanteric fractures, transverse fractures, oblique and spiral fractures, segmental fractures, comminuted fractures, high supracondylar fractures, fractures with bone loss, stabilization for bone transport or lengthening and

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo for a company called ALPHATEC Manufacturing, Inc. The logo features a circular emblem with a stylized "O" in the center, possibly representing the company's initial or a symbol of quality. Below the emblem, the company name "ALPHATEC Manufacturing, Inc." is printed in a bold, sans-serif font, emphasizing the company's focus on manufacturing.

20F2

shortening, open fractures, corrective osteotomies, pathologic fractures, pseudoarthrosis of the bone shaft, nonunions, malunions, metaphyseal junction fractures, and acute bone lengthening and shortening.

### Substantial Equivalence

The Pylon Intramedullary Nail System is substantially equivalent to ACE AIM Titanium Nails (Depuy ACE, a Johnson and Johnson company). The Pylon Intramedullary Nail System is similar to the listed predicate device in design, function, materials used, and indications for use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three stripes extending from its head, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 2 1999

Mr. Jason Blain Manager of Product Development Alphatec Manufacturing, Inc. 42-160 State Street Palm Desert, California 92211

Re: K992350 Trade Name: Pylon Intramedullary Nail System Regulatory Class: II Product Code: HSB Dated: July 12, 1999 Received: July 14, 1999

Dear Mr. Blain:

We have reviewed your Section 510(k) notification of intent to market the device referenced above, and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General Regulation (21 CFR Part 820), and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Jason Blain

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Sincerely yours,

ia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo for a company called "ALPHATEC Manufacturing, Inc." The logo consists of a circular emblem with the letters "OA" inside. Below the emblem, the company name is written in a bold, sans-serif font. The word "ALPHATEC" is on the first line, and "Manufacturing, Inc." is on the second line in a smaller font size.

Premarket Notification Pylon Intramedullary Nail System Alphatec Manufacturing, Inc.

## Intended Uses/Indications

K9923350 510(k) Number:

Device Name: Pylon Intramedullary Nail System

Indications for Use:

The Pylon Intramedullary Nail System is indicated for internal fixation of tibial or femoral fractures including low subtrochanteric fractures, transverse fractures, oblique and spiral fractures, segmental fractures, comminuted fractures, high supracondylar and spiral fractures with bone loss, stabilization for bone transport or lengthening and shortening, open fractures, corrective osteotomies, pathologic fractures, pseudoarthrosis of the bone shaft, nonunions, malunions, metaphyseal/diaphyseal junction fractures, and acute bone lengthening and shortening.

bodles

(Division Sigh, Off) · · sion of General Re 5 | Ü(K) Number

455

Prescription Use
(Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K992350](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K992350)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
