← Product Code [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB) · K232169

# FITBONE® Transport and Lengthening System (K232169)

_Orthofix Srl · HSB · Mar 22, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K232169

## Device Facts

- **Applicant:** Orthofix Srl
- **Product Code:** [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB.md)
- **Decision Date:** Mar 22, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

FITBONE Transport and Lengthening system is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, malunions, non-unions, or bone transport of the long bones. The FITBONE Transport and Lengthening system is indicated for adult only.

## Device Story

Implantable intramedullary transport/lengthening nail; implanted in femur or tibia medullary canal; connected to subcutaneous receiver via bipolar feed line. External control set (identical to K203399) includes control electronics station and transmitter. Distraction process powered by hermetically enclosed motor; electromagnetic field from external transmitter induces DC-voltage in subcutaneous receiver to drive motor. Nail anchored to bone via locking screws. No transcutaneous contact between nail and body surface. Used in clinical setting by surgeons. Output is mechanical distraction of bone; enables limb lengthening and fracture fixation. Benefits include controlled bone transport and lengthening.

## Clinical Evidence

Bench testing only. Testing included static and fatigue 4-point bending (ASTM F1264-16), torsional testing (ASTM F1264-16), and motor current consumption analysis under load. Fatigue limit calculations performed per ISO 12107:2012.

## Technological Characteristics

Implantable intramedullary nail; materials: stainless steel 1.4441 (ASTM F138-13) and silicone (NuSil MED-4870, MED-1511, MED-4750). Distraction via internal motor; electromagnetic energy transfer from external transmitter to subcutaneous receiver. Dimensions: 290-490mm length, 11-13mm diameter. Sterilization: Gas Plasma.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Precice® Bone Transport System ([K201567](/device/K201567.md))
- FITBONE® TAA ([K203399](/device/K203399.md))

## Reference Devices

- Orthofix Femoral Nailing system ([K973944](/device/K973944.md))
- Orthofix Tibial Nailing system ([K961027](/device/K961027.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 22, 2024

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Orthofix S.r.l. Elvira Taccarelli Regulatory Affairs Manager Via Delle Nazioni, 9 Bussolengo (VR), IT 37012 Italy

Re: K232169

Trade/Device Name: FITBONE® Transport and Lengthening System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: March 13, 2024 Received: March 13, 2024

Dear Elvira Taccarelli:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Digitally signed by Farzana Farzana Sharmin -S Date: 2024.03.22 15:20:00 Sharmin -S

Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Submission Number (if known)

K232169

Device Name

## FITBONE® TRANSPORT AND LENGTHENING SYSTEM

Indications for Use (Describe)

"FITBONE Transport and Lengthening system" is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, malunions, non-unions, or bone transport of the long bones. The "FITBONE Transport and Lengthening system" is indicated for adult only.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, bold letters. A small "R" in a circle is located to the right of the "X" in "ORTHOFIX", indicating a registered trademark.

## 510(K) SUMMARY

# ORTHOFIX SRL FITBONE® Transport and Lengthening System

## Submitter information

| Company Name:  | Orthofix S.r.l.                                       |
|----------------|-------------------------------------------------------|
| Address        | Via Delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy |
| Telephone      | +39 045 6719000                                       |
| Fax            | +39 045 6719380                                       |
| Contact Person | Elvira Taccarelli<br>Regulatory Affairs Manager       |
| Address        | Via Delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy |
| Telephone      | +39 045 6719000                                       |
| Fax            | +39 045 6719380                                       |
| Email address  | elvirataccarelli@orthofix.it                          |
| Date prepared  | 2024, January 31                                      |

### Trade Name, Common Name, Classification

| Trade Name:          | FITBONE® Transport and Lengthening System |
|----------------------|-------------------------------------------|
| Common Name:         | Intramedullary lengthening nail           |
| Classification Name: | Intramedullary fixation rod               |
| Regulation Number:   | 21 CFR 888.3020                           |
| Product Code:        | HSB                                       |
| Classification:      | Class II                                  |
| Panel code:          | Orthopedic                                |

## Primary Predicate and additional predicate device

| Primary Predicate              | 510(k) Number | Manufacturer                             |
|--------------------------------|---------------|------------------------------------------|
| Precice® Bone Transport System | K201567       | NuVasive Specialized<br>Orthopedics Inc. |
| Additional predicate           | 510(k) Number | Manufacturer                             |
| FITBONE® TAA                   | K203399       | Orthofix s.r.l.                          |

#### Reference devices

| Reference Devices               | 510(k) Number | Manufacturer    |
|---------------------------------|---------------|-----------------|
| Orthofix Femoral Nailing system | K973944       | Orthofix s.r.l. |
| Orthofix Tibial Nailing system  | K961027       | Orthofix s.r.l. |

() Orthofix Srl | Via delle Nazioni, 9 | 37012 Bussolengo (VR) Italy \ \ +39 045 6719380 ✈ orthofix.com

DIREZIONE E COORDINAMENTO: Orthofix Medical Inc. (Delaware)

SEDE LEGAL: VA VITOR RIAN 1 L . P. P.VA 097669150 CE VA NUMER TI097660150 R.E., M 125769 Cap. So. E., M 127769 Cap. So. Eu. 1.56.000 in. ves.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, sans-serif font. A registered trademark symbol is located to the right of the "X" in "ORTHOFIX".

| Device description               | The subject “FITBONE® Transport and Lengthening System” is<br>indicated for limb lengthening, open and closed fracture fixation,<br>pseudoarthrosis, malunions, non-unions, or bone transport of the long<br>bones. The Fitbone Transport and Lengthening System is indicated for<br>adult only.<br>The subject “FITBONE® Transport and Lengthening System” consists<br>of the implantable intramedullary transport or lengthening nail and its<br>trial nail accessories. The Subject device is implanted into the<br>medullary canal of the femur or tibia and connected to the additional<br>predicate intracutaneous Receiver by a bipolar feed line. The external<br>FITBONE Control Set is identical to that previously cleared for the<br>additional predicate Fitbone TAA device (K203399) and consists of a<br>control electronics station and transmitter. There are no changes to this<br>previously cleared Control Sets and Receiver as a result of this<br>submission.<br>The power required for the distraction process is controlled by<br>hermetically enclosed motor which draws the telescope apart. The<br>electro-magnetic field sent from the Transmitter to the Receiver is<br>converted in the Receiver into DC-Voltage to supply the motor of the<br>FITBONE Transport and Lengthening Nail with voltage, when actioned.<br>The subject nail is anchored to the bone by locking screws through<br>medial-lateral and AP holes in the nail depending on the configuration<br>holes in the nail. The energy needed for the distraction process is<br>transmitted from the outside by placing the external transmitter over the<br>implanted receiver, which is placed in the subcutaneous tissue during<br>surgery.<br>There is no transcutaneous contact between the implanted<br>intramedullary nail and the outer surface of the patient's body.<br>The subject trial nails accessories are available for each size model of<br>the FITBONE Transport (TN) and FITBONE Transport or Lengthening<br>(TLN) nails and are used to simulate the shape of the implant.<br>The subject nails and trial nails are provided in sterile conditions only<br>and are made from, as follows:<br>• Nail: implant grade stainless steel 1.4441, according to ASTM F138-<br>13 "Standard Specification for Wrought 18Chromium-14Nickel-<br>2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants<br>(UNS S31673)", and Silicone Nusilmed (NuSil MED-4870, NuSil MED-<br>1511, Nusil MED 4750).<br>• Trial nail: implant grade stainless steel 1.4441, according to ASTM<br>F138-13 "Standard Specification for Wrought 18Chromium-14Nickel-<br>2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants<br>(UNS S31673)" and ASTM F899-20 Standard Specification for<br>Wrought Stainless Steels for Surgical Instruments<br>For the implantation and removal of the subject device the same<br>instruments of the additional predicate Fitbone TAA (K203399) may be<br>used. |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use              | "FITBONE Transport and Lengthening system" is indicated for limb<br>lengthening, open and closed fracture fixation, pseudoarthrosis,<br>malunions, non-unions, or bone transport of the long bones.<br>The "FITBONE Transport and Lengthening system" is indicated for<br>adult only.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Technological<br>Characteristics | The following table provides a comparison and assessment of<br>fundamental scientific principles and technological characteristics of<br>the subject, primary predicate and additional predicate devices. Any<br>differences have been demonstrated to not raise new issues of safety<br>or performance by virtue of objective evidence (e.g., bench testing,<br>process validations, etc.). Therefore, the technological characteristics<br>of the subject FITBONE® Transport and Lengthening system are<br>equivalent to the primary predicate Precice® Bone Transport System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

() Orthofix Srl | Via delle Nazioni, 9 | 37012 Bussolengo (VR) Italy ↑ 29 045 6719380 曲 orthofix.com

DIREZIONE E COORDINAMENTO: Orthofix Medical Inc. (Delaware)

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "ORTHOFIX" in a bold, sans-serif font. To the left of the word is a blue, abstract shape that resembles a stylized "O". A small "R" in a circle is located to the upper right of the word "ORTHOFIX", indicating that it is a registered trademark.

|                                                                                                                                                                                                                                                                               | (K201567) and the additional predicate device FITBONE® TAA (K203399).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                                                                                                                                                                                                               | Technological Characteristic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Subject Device (Fitbone Bone Transport and Lengthening System)                                                                                                                                                             | Primary Predicate Device (Precice Bone Transport System, K201567)                                                                               | Additional Predicate Device (Fitbone TAA, K203399)                                                                                         |
| 1.                                                                                                                                                                                                                                                                            | Indications for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, malunions, non-unions, or bone transport of the long bones. The Fitbone Transport and Lengthening System is indicated for adults only. | Indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, malunions, non-unions, or bone transport of the long bones. | Intended for limb lengthening of the femur and tibia. Indicated for adult and pediatric (greater than 12 through 21 years of age) patients |
|                                                                                                                                                                                                                                                                               | Assessment: The subject indications are identical to the primary predicate and carry the same lengthening indication as the additional predicate. Equivalent - no significant new issues have been raised.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
| 2.                                                                                                                                                                                                                                                                            | Anatomical Sites                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Long bones                                                                                                                                                                                                                 | Long bones                                                                                                                                      | Femur, tibia                                                                                                                               |
|                                                                                                                                                                                                                                                                               | Assessment: The anatomical sites for use are identical to primary predicate and similar to additional predicate (femur and tibia are long bones). Equivalent - no significant new issues raised.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
| 3.                                                                                                                                                                                                                                                                            | Nail Material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Implant Grade Stainless Steel (1.4441, 316LVM)                                                                                                                                                                             | Implant Grade Stainless Steel (ASTM F2229, Biodur 108)                                                                                          | Implant Grade Stainless Steel (1.4441, 316LVM)                                                                                             |
|                                                                                                                                                                                                                                                                               | Assessment: Subject material is similar to primary predicate and identical to additional predicate. Both have been demonstrated to be biocompatible implant grade stainless steels with unique corrosion properties. Equivalent - no significant new issues raised.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
| 4.                                                                                                                                                                                                                                                                            | Nail Size Range                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 290-490mm in length; 11 and 13mm diameters in varying configurations                                                                                                                                                       | 280-400mm in length; 10, 11.5, and 13mm diameters in varying configurations                                                                     | 200-245mm in length; 9, 11, and 13mm diameters in varying configurations                                                                   |
|                                                                                                                                                                                                                                                                               | Assessment: Size ranges of subject are similar to primary and additional predicate. Potential performance differences arising from variations in size offerings have been demonstrated through bench testing to not raise significant new issues. Further mitigation is provided through weight bearing precautions in labeling. The ranges are therefore equivalent.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
| 5.                                                                                                                                                                                                                                                                            | Maximum Stroke Length Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 80mm                                                                                                                                                                                                                       | 100mm                                                                                                                                           | 80mm                                                                                                                                       |
|                                                                                                                                                                                                                                                                               | Assessment: The subject stroke length is within that of primary predicate and identical to the additional predicate. Equivalent - no significant new issues raised.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
| 6.                                                                                                                                                                                                                                                                            | Method of Distraction/Energy Source                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Internal motor electro-magnetically induced by an external transmitter with signal received through a receiver placed just under                                                                                           | Internal magnetically powered telescopic feature with two rotating magnets induced by an external controller                                    | Internal motor electro-magnetically induced by an external transmitter with signal received through a receiver placed just under           |
| Assessment: Subject distraction methodology is similar to primary predicate and<br>identical to additional predicate. Differences in technology have been demonstrated<br>through bench testing to not raise significant new issues. The methods are therefore<br>equivalent. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
| 7.                                                                                                                                                                                                                                                                            | Sterilization Method                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Gas Plasma                                                                                                                                                                                                                 | Gamma                                                                                                                                           | Gas Plasma                                                                                                                                 |
| Assessment: Subject is similar to primary predicate and identical to additional<br>predicate. Supporting process validation demonstrates that there are no new issues<br>raised by the differences in methodology. The sterilization methods are therefore<br>equivalent.     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
| Performance Analysis                                                                                                                                                                                                                                                          | Mechanical testing of the subject devices to demonstrate equivalence<br>in function and safety for the intended use of the FITBONE® Transport<br>and Lengthening nails and trial nails in respect to its additional<br>predicate FITBONE® TAA (K203399) have been performed for the<br>following functional attributes and are included in section "Bench<br>Testing" of this eSTAR application form, as below:<br>Implantable rod: Static 4-point bending test (with Subject<br>FITBONE® Transport and Lengthening nail) performed according<br>to ASTM F1264-16 - Standard Specification and Test Methods for<br>Intramedullary Fixation Devices, Implantable rod: Fatigue 4-point bending test (with Subject<br>FITBONE® Transport and Lengthening nail) performed according<br>to ASTM F1264-16 - Standard Specification and Test Methods for<br>Intramedullary Fixation Devices. The Fatigue limit calculation are,<br>instead, performed according to ISO 12107:2012 - Metallic<br>materials — Fatigue testing — Statistical planning and analysis of<br>data, Implantable rod: Torsional test (with Subject FITBONE® Transport<br>and Lengthening nail) performed according to ASTM F1264-16 -<br>Standard Specification and Test Methods for Intramedullary<br>Fixation Devices, Implantable rod: Absorbed current by the nail's motor during usage<br>when different forces are applied (with Subject FITBONE®<br>Transport and Lengthening nail). The test performed does not refer<br>to any standard since it is designed for this specific component.<br>The test requires the measurement of the current consumption of<br>the motor while running to obtain a displacement when different<br>forces are applied to the actuator. |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |
| Conclusion                                                                                                                                                                                                                                                                    | Based upon equivalences in: intended use, patient population, site of<br>application, conditions of use, operating principles, and the non-clinical<br>performance data, the subject FITBONE® Transport and Lengthening<br>System have been shown to be substantially equivalent to the legally<br>marketed predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                            |                                                                                                                                                 |                                                                                                                                            |

() Orthofix Srl | Via delle Nazioni, 9 | 37012 Bussolengo (VR) Italy ↑ 2 +39 045 6719380   冊 orthofix.com

DIREZIONE E COORDINAMENTO: Orthofix Medical Inc. (Delaware)

SEDE LEGAL: VA VITOR RIANI 6 L. P. P.VA 0897660150 CE VA NUMER TOSP660150 Registo Millero97680010 R.E. M 125769 Cap. So. E.o. 1.56.000 in. ves.
 SED 0.00 in. VA DELL MADON,

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Orthofix. The logo consists of a blue abstract shape on the left, followed by the word "ORTHOFIX" in black, sans-serif font. A registered trademark symbol is located to the right of the "X" in "ORTHOFIX".

() Orthofix Srl | Via delle Nazioni, 9 | 37012 Bussolengo (VR) Italy ↑ 29 045 6719380 曲 orthofix.com

DIREZIONE E COORDINAMENTO: Orthofix Medical Inc. (Delaware)

SED LEGAL: VA VITOR PIANI 6 L D. A. A. S. P. A 0897660150 CE VAT NUMB: RT097660150 R.E., Mi 125769 Cap. So. E.E. 1.560.00 in. ves.
 SED. SECONDRIA: VA DELL MAZON, 9 37012 BUS

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K232169](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K232169)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
