← Product Code [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB) · K201567

# Precice® Bone Transport System (K201567)

_Nuvasive Specialized Orthopedics, Inc. · HSB · Aug 10, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K201567

## Device Facts

- **Applicant:** Nuvasive Specialized Orthopedics, Inc.
- **Product Code:** [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB.md)
- **Decision Date:** Aug 10, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Precice® Bone Transport System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

## Device Story

System comprises intramedullary nail, locking screws, end caps, surgical instruments, and External Remote Controller (ERC). Nail contains enclosed rare earth magnet, distraction rod, and planetary gearing. Implanted in long bones by surgeons; provides internal fixation. ERC used non-invasively to adjust distraction rod extension via magnetic coupling; enables gradual limb lengthening or compression of intercalary bone segments. System available in various diameters/lengths to accommodate patient anatomy. Benefits include non-invasive adjustment of bone segment position post-implantation.

## Clinical Evidence

Bench testing only. Performance verified via static/dynamic compression, bending strength, torsion, torque resistance, axial pullout, retention force, and distraction force testing per ASTM F1264 and ASTM F543 standards.

## Technological Characteristics

Intramedullary fixation rod system. Materials: metallic components (implants). Sensing/Actuation: magnetic coupling between external controller and internal planetary gearing/distraction rod. Connectivity: non-invasive magnetic interface. Sterilization: gamma radiation for implants; instruments supplied non-sterile. Design: various diameters/lengths/screw configurations.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Precice® Bone Transport System ([K193016](/device/K193016.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION" in a smaller font size.

NuVasive Specialized Orthopedics, Inc. Madison Heffron Regulatory Affairs Specialist 101 Enterprise, Suite 100 Aliso Viejo, California 92656

# Re: K201567

Trade/Device Name: Precice® Bone Transport System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, HWC Dated: June 9, 2020 Received: June 11, 2020

# Dear Madison Heffron:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

August 10, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number K201567

Device Name

Precice® Bone Transport System

Indications for Use (Describe)

The Precice® Bone Transport System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

| Type of Use (Select one or both, as applicable)                                                                                                                |  |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="text-decoration: overline;">X</span>                     Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> |  |
| <div> <span>                     Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div>                                                    |  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and silver graphic to the left of the company name, "NUVASIVE," in silver lettering. Below the name, in smaller purple letters, is the text "SPECIALIZED ORTHOPEDICS, INC."

# Precice® Bone Transport System 510(k) Summary

#### A. Submitted by:

Madison Heffron Regulatory Affairs Specialist NuVasive Specialized Orthopedics, Inc. 101 Enterprise, Suite 100 Aliso Viejo, CA 92656 Telephone: (949) 532-7868

Date Prepared: June 8, 2020

#### B. Device Name

| Trade or Proprietary Name: | Precice® Bone Transport System |
|----------------------------|--------------------------------|
| Common or Usual Name:      | Intramedullary Fixation Rod    |
| Classification Name:       | Intramedullary Fixation Rod    |

| Device Class:              | Class II          |
|----------------------------|-------------------|
| Classification:            | 21 CFR § 888.3020 |
| Product Code (primary):    | HSB               |
| Product Code (additional): | HWC               |

#### C. Predicate Devices

The subject Precice® Bone Transport System is substantially equivalent to the primary predicate device, Precice® Bone Transport System (K193016).

#### D. Device Description

The Precice® Bone Transport System includes the Precice® Bone Transport Nail, locking screws, end caps, surgical instruments, and external remote controller (ERC). The Precice ® Bone Transport nails, screws and end caps are supplied sterile by gamma radiation while the instruments are supplied non-sterile and must be sterilized prior to use. The system is designed to achieve limb correction through gradual lengthening or compression of the intercalary bone segment and providing internal fixation for fractures of long bones. The Precice® Bone Transport intramedullary nail is implanted using locking screws, end caps, and reusable surgical instruments. The Precice® Bone Transport nail contains an enclosed rare earth magnet, distraction rod, and planetary gearing which allows the extension of the distraction rod to be adjusted noninvasively by the External Remote Controller (ERC). The Precice ® Bone Transport nail is available in various diameters, lengths and screw hole configurations to accommodate a variety of patient anatomies and implantation methods. The locking screws are also available in a variety of diameters, lengths, and thread styles. The ERC is available in several compatible models.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic to the left of the company name. The company name is in gray, with the words "SPECIALIZED ORTHOPEDICS, INC." in smaller purple font below.

The purpose of this submission is to introduce design modifications to the distal portion of the Precice ® Bone Transport device.

#### E. Indications for Use

The Precice® Bone Transport System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.

#### F. Comparison of Technological Characteristics with the Predicate Device

As was established in this submission, the subject Precice® Bone Transport System is substantially equivalent to the predicate Precice ® Bone Transport System (K193016) previously cleared by the FDA for commercial distribution in the United States. The subject device has been shown to be substantially equivalent and have equivalent technological characteristics to its predicate device through comparison in areas including design, labeling/intended use, material composition, and function.

#### G. Performance Data

Nonclinical performance verification testing was performed to demonstrate that the subject Precice® Bone Transport System is substantially equivalent to the predicate device. The following testing was performed:

| Testing Description                     | Applicable Standard                                                                            |
|-----------------------------------------|------------------------------------------------------------------------------------------------|
| Static Compression<br>Bending Strength  | ASTM F1264 - Standard Specification and<br>Test Methods for Intramedullary Fixation<br>Devices |
| Dynamic Compression<br>Bending strength |                                                                                                |
| Torsion                                 |                                                                                                |
| Torque Resistance                       | ASTM F543 - Standard Specification and Test<br>Methods for Metallic Medical Bone Screws        |
| Axial Pullout                           |                                                                                                |
| Retention Force Testing                 | N/A                                                                                            |
| Distraction Force                       | N/A                                                                                            |

The results demonstrate that the subject Precice® Bone Transport System is substantially equivalent to the predicate.

#### H. Conclusions

The subject Precice® Bone Transport System has been shown to be substantially equivalent to the legally marketed predicate device for its intended use.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K201567](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K201567)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
