← Product Code [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB) · K173458

# ORTHOFIX CHIMAERA Hip Fracture System - Trochanteric Nailing System (K173458)

_Orthofix Srl · HSB · Dec 7, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K173458

## Device Facts

- **Applicant:** Orthofix Srl
- **Product Code:** [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB.md)
- **Decision Date:** Dec 7, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The ORTHOFIX CHIMAERA Hip Fracture System – trochanteric nailing system is intended for insertion into the medullary canal of a femur for the alignment, stabilization of various types of fractures or deformities. The ORTHOFIX CHIMAERA Hip Fracture System - trochanteric nailing system is indicated for treatment of stable and unstable pertrochanteric, intertrochanteric and subtrochanteric fractures of the femur alone or when these fractures occur in combination with shaft fractures extending distally to a point approximal to the intercondylar notch. These includes traumatic fractures, refractures, non-union, reconstruction, malalignment, pathological fractures and impending pathological fractures.

## Device Story

System consists of intramedullary nails, end caps, and screws; used for internal fixation of femoral fractures. Surgeon inserts nail into femoral medullary canal using instrumentation (guide wire, awl, reamer, impactor). Proximal nail features threaded bore for connection to targeting handle via cannulated bolt. Device provides mechanical stabilization of bone fragments to facilitate healing. Used in clinical/surgical settings by orthopedic surgeons. Benefits include restoration of bone alignment and structural support for fracture recovery.

## Clinical Evidence

No clinical data. Safety and effectiveness supported by bench testing, including mechanical performance evaluation per ASTM F1264-16.

## Technological Characteristics

Intramedullary fixation system comprising nails, end caps, and screws. Mechanical testing performed per ASTM F1264-16 (Standard Specifications and Test Methods for Intramedullary Fixation Devices).

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- ORTHOFIX CHIMAERA Hip Fracture System - Trochanteric Nailing System ([K161466](/device/K161466.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

December 7, 2017

Orthofix Srl % Cheryl Wagoner Consultant Wagoner Consulting LLC PO Box 15729 Wilmington, North Carolina 28408

Re: K173458

Trade/Device Name: ORTHOFIX CHIMAERA Hip Fracture System - Trochanteric Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: November 6, 2017 Received: November 7, 2017

Dear Cheryl Wagoner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Katherine D. Kavlock -5 for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173458

Device Name

ORTHOFIX CHIMAERA Hip Fracture System - Trochanteric Nailing System

Indications for Use (Describe)

The ORTHOFIX CHIMAERA Hip Fracture System – trochanteric nailing system is intended for insertion into the medullary canal of a femur for the alignment, stabilization of various types of fractures or deformities.

The ORTHOFIX CHIMAERA Hip Fracture System - trochanteric nailing system is indicated for treatment of stable and unstable pertrochanteric, intertrochanteric and subtrochanteric fractures of the femur alone or when these fractures occur in combination with shaft fractures extending distally to a point approximal to the intercondylar notch. These includes traumatic fractures, refractures, non-union, reconstruction, malalignment, pathological fractures and impending pathological fractures.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Special 510(k) Premarket Notification ORTHOFIX CHIMAERA Hip Fracture System - Trochanteric Nailing System

# 510(k) Summary

(as required by 21 CFR 807.92)

| Submitter                      | Orthofix Srl                                      |                                                                                                                     |  |  |
|--------------------------------|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|--|--|
|                                | Via delle Nazioni, 9                              |                                                                                                                     |  |  |
|                                | 37012 Bussolengo (VR) - Italy                     |                                                                                                                     |  |  |
| Telephone                      | + 39 045 6719.000                                 |                                                                                                                     |  |  |
| Fax                            | + 39 045 6719.380                                 |                                                                                                                     |  |  |
| Contact Person                 | Gianluca Ricadona                                 |                                                                                                                     |  |  |
|                                | Sr. Quality & Regulatory Affairs Manager          |                                                                                                                     |  |  |
| Address                        |                                                   | Via delle Nazioni, 9                                                                                                |  |  |
|                                | 37012 Bussolengo (VR) - Italy                     |                                                                                                                     |  |  |
| Telephone                      | + 39 045 6719.000                                 |                                                                                                                     |  |  |
| Fax                            | + 39 045 6719.380                                 |                                                                                                                     |  |  |
| email                          | GianlucaRicadona@orthofix.it                      |                                                                                                                     |  |  |
|                                |                                                   |                                                                                                                     |  |  |
| Submission date                | 11/06/2017                                        |                                                                                                                     |  |  |
|                                |                                                   |                                                                                                                     |  |  |
| Trade Name                     | ORTHOFIX<br>CHIMAERA<br>Hip<br>Fracture<br>System |                                                                                                                     |  |  |
|                                | Trochanteric Nailing System                       |                                                                                                                     |  |  |
| Common Name                    | Rod, fixation, intramedullary and accessories     |                                                                                                                     |  |  |
| Panel Code                     | 87 / Orthopaedic                                  |                                                                                                                     |  |  |
| Classification Name            | Intramedullary fixation rod.                      |                                                                                                                     |  |  |
| Class                          | Class II                                          |                                                                                                                     |  |  |
| Regulation Number              | 21 CFR 888.3020                                   |                                                                                                                     |  |  |
| Product Code                   | HSB                                               |                                                                                                                     |  |  |
|                                |                                                   |                                                                                                                     |  |  |
| Name of Predicate Device       | 510(k) #                                          | Manufacturer                                                                                                        |  |  |
| ORTHOFIX CHIMAERA Hip          | K161466                                           | Orthofix Srl                                                                                                        |  |  |
|                                |                                                   |                                                                                                                     |  |  |
| Fracture System - Trochanteric |                                                   |                                                                                                                     |  |  |
| Nailing System                 |                                                   |                                                                                                                     |  |  |
|                                |                                                   |                                                                                                                     |  |  |
| Description                    |                                                   | The ORTHOFIX CHIMAERA Hip Fracture System -                                                                         |  |  |
|                                |                                                   | Trochanteric Nailing System consists of implantable                                                                 |  |  |
|                                | components                                        | (nails, end caps and screws)<br>and                                                                                 |  |  |
|                                |                                                   | instrumentation. The proximal part of the nail features a                                                           |  |  |
|                                |                                                   | threaded bore to connect the nail to the targeting handle                                                           |  |  |
|                                |                                                   | by means of a cannulated bolt. The insertion of the nail                                                            |  |  |
|                                |                                                   | into the femur medullary canal is typically performed by                                                            |  |  |
|                                |                                                   | some instruments, including: guide wire, awl, reamer                                                                |  |  |
|                                | and impactor elements.                            |                                                                                                                     |  |  |
| Intended Use                   |                                                   | The ORTHOFIX CHIMAERA Hip Fracture System -                                                                         |  |  |
| Indications for use            |                                                   | Trochanteric Nailing System is intended for insertion into                                                          |  |  |
|                                |                                                   | the medullary canal of a femur for the alignment,                                                                   |  |  |
|                                |                                                   | stabilization and fixation of various types of fractures or                                                         |  |  |
|                                | deformities.                                      |                                                                                                                     |  |  |
|                                |                                                   | The ORTHOFIX CHIMAERA Hip Fracture System -                                                                         |  |  |
|                                |                                                   | Trochanteric Nailing System is indicated for treatment of                                                           |  |  |
|                                |                                                   | stable and unstable pertrochanteric, intertrochanteric                                                              |  |  |
|                                |                                                   | and subtrochanteric fractures of the femur alone or when                                                            |  |  |
|                                |                                                   | these fractures occur in combination with shaft fractures,                                                          |  |  |
|                                |                                                   | or when these fractures occur in combination with shaft<br>fractures extending distally to a point approximately 10 |  |  |

{4}------------------------------------------------

# Special 510(k) Premarket Notification

ORTHOFIX CHIMAERA Hip Fracture System - Trochanteric Nailing System

|                                   | These includes traumatic fractures, re-fractures, non-<br>reconstruction,<br>malunion,<br>malalignment.<br>union.<br>fractures and impending<br>pathological<br>pathological<br>fractures. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Characteristics and | Documentation provided<br>demonstrates<br>that<br>the                                                                                                                                      |
|                                   |                                                                                                                                                                                            |
| Substantial Equivalence           | ORTHOFIX CHIMAERA Hip Fracture<br>System                                                                                                                                                   |
|                                   | Trochanteric Nailing System is substantially equivalent                                                                                                                                    |
|                                   | to the legally marketed predicate.                                                                                                                                                         |
|                                   | The components and instruments included in the Subject                                                                                                                                     |
|                                   | System and the predicate are all internal fracture fixation                                                                                                                                |
|                                   | systems, as defined in 21 CFR 888.3020.                                                                                                                                                    |
|                                   | The ORTHOFIX CHIMAERA Hip Fracture System -                                                                                                                                                |
|                                   |                                                                                                                                                                                            |
|                                   | Trochanteric Nailing System is substantially equivalent                                                                                                                                    |
|                                   | to the predicate device in: intended use and Indications                                                                                                                                   |
|                                   | for use, site of application, patient population,  condition                                                                                                                               |
|                                   | of use, basic design, technology, materials, implantable                                                                                                                                   |
|                                   | components and mechanical performances.                                                                                                                                                    |
| Performance Data                  | potential hazards have been evaluated<br>The<br>and                                                                                                                                        |
|                                   | controlled through a Risk Management Plan.                                                                                                                                                 |
|                                   | All testing met or exceeded the requirements, as                                                                                                                                           |
|                                   | established by the test protocols and applicable                                                                                                                                           |
|                                   | standard.                                                                                                                                                                                  |
|                                   |                                                                                                                                                                                            |
|                                   | A review of the mechanical data indicates that the                                                                                                                                         |
|                                   | components of the Subject device continues to be                                                                                                                                           |
|                                   | capable of withstanding expected loads without failure.                                                                                                                                    |
|                                   | Subject device was therefore found to be<br>The                                                                                                                                            |
|                                   | substantially equivalent to the Predicate device and has                                                                                                                                   |
|                                   | the mechanical properties to perform its indications                                                                                                                                       |
|                                   | safely.                                                                                                                                                                                    |
|                                   | Clinical data was not needed to support the safety and                                                                                                                                     |
|                                   | effectiveness of the Subject Device.                                                                                                                                                       |
|                                   |                                                                                                                                                                                            |
|                                   | The following standard has been followed to perform                                                                                                                                        |
|                                   | mechanical test on the System configuration:                                                                                                                                               |
|                                   | · ASTM F1264 -16 Standard Specifications and Test                                                                                                                                          |
|                                   | Methods for Intramedullary Fixation Devices.                                                                                                                                               |
| Conclusion                        | Based upon similarities in: intended use and Indications                                                                                                                                   |
|                                   | for use, site of application, patient population,  condition                                                                                                                               |
|                                   | of use, basic design, technology, materials, implantable                                                                                                                                   |
|                                   | components and mechanical performances, the Subject                                                                                                                                        |
|                                   | System has been shown to be substantially equivalent to                                                                                                                                    |
|                                   |                                                                                                                                                                                            |
|                                   | the legally marketed predicate device and to be safe and                                                                                                                                   |
|                                   | effective for its intended use.                                                                                                                                                            |

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K173458](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K173458)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
