Renovo Life Small Bone IM Nail System

K161254 · Renovo Life, LLC · HSB · Nov 7, 2016 · Orthopedic

Device Facts

Record IDK161254
Device NameRenovo Life Small Bone IM Nail System
ApplicantRenovo Life, LLC
Product CodeHSB · Orthopedic
Decision DateNov 7, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The Renovo Life Small Bone IM Nail System is indicated for fixation of fractures where the canal is narrow or flexibility of the implant is paramount, including the fibula, radius, and ulna, upper extremity fractures in all patients and lower extremity fractures in small-stature patients. This system is also intended to treat metaphyseal and epiphyseal fractures, such as radial neck fractures and is intended for fixation of small long bones, such as carpal and tarsal bones.

Device Story

Intramedullary fixation system for small bone fractures; consists of nails (2.5-4.5mm diameter, 150mm length, solid or cannulated) and 3.5mm screws. Used by orthopedic surgeons in clinical settings for fracture stabilization. Provides mechanical support to bone segments to facilitate healing. Non-sterile, single-use device.

Clinical Evidence

No clinical testing was performed. Evidence is based on bench testing (static/dynamic four-point bend, static torsion, compliance) per ASTM 1264-14, cadaveric implant trials, and rotational resistance tests.

Technological Characteristics

Materials: surgical grade stainless steel (ASTM F138) or titanium alloy (ASTM F136). Form factor: intramedullary nails (2.5-4.5mm diameter, 150mm length) and screws (3.5mm diameter, 10-34mm length). Mechanical fixation principle. Provided non-sterile, single-use.

Indications for Use

Indicated for fixation of fractures in patients with narrow canals or where implant flexibility is required. Includes fibula, radius, and ulna fractures; upper extremity fractures in all patients; lower extremity fractures in small-stature patients; metaphyseal and epiphyseal fractures (e.g., radial neck); and fixation of small long bones (carpal and tarsal).

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, arranged in a way that resembles a bird in flight. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 7, 2016 Renovo Life LLC % Mr. Donald Guthner Regulatory Consultant Orgenix, LLC 111 Hill Road Douglassville, Pennsylvania 19518 Re: K161254 Trade/Device Name: Renovo Life Small Bone IM Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: October 11, 2016 Received: October 12, 2016 Dear Mr. Guthner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K161254 Device Name Renovo Life Small Bone IM Nail System #### Indications for Use (Describe) The Renovo Life Small Bone IM Nail System is indicated for fixation of fractures where the canal is narrow or flexibility of the implant is paramount, including the fibula, radius, and ulna, upper extremity fractures in all patients and lower extremity fractures in small-stature patients. This system is also intended to treat metaphyseal fractures, such as radial neck fractures and is intended for fixation of small long bones, such as carpal and tarsal bones. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------|-------------------------------------------------------------------------------| | <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary | Manufacturer: | Renovo Life LLC<br>1104 Spruce Street<br>Belmont, NC 28012 | |--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date: | October 10, 2016 | | Submitted by: | Renovo Life LLC | | Company Contact | Harold Crowder<br>+1(704) 224-4136<br><a href="mailto:hcrowder@renovolife.com">hcrowder@renovolife.com</a> | | US Agent Information<br>Official Correspondent | Orgenix LLC<br>Mr. Donald W. Guthner<br>111 Hill Road<br>Douglassville, PA 19518<br>+1-646-460-2984<br>+1-484-363-5879 (FAX)<br><a href="mailto:dg@orgenix.com">dg@orgenix.com</a> | | Proprietary Name: | Renovo Life Small Bone IM Nail System | | Performance standards: | The Renovo Life Small Bone IM Nail System is made of surgical grade<br>stainless steel conforming to ASTM F138 or surgical grade Titanium<br>alloy conforming to ASTM F136. | | Regulation:<br>Common/Usual Name:<br>Classification name:<br>Review Panel:<br>Product Code:<br>Device Class: | 21 CFR 888.3020<br>Rod, Fixation, Intramedullary and Accessories<br>Intramedullary Fixation Rod<br>Orthopedic<br>HSB<br>Class II | | Substantial Equivalence: | Substantial equivalence for the Renovo Life Small Bone IM Nail<br>Systems is based on its similarities in indications for use, design features,<br>operational principles and material composition when compared to the<br>predicate devices. | | Predicate Devices: | <ul><li>K043185 Synthes Cortex Screw</li><li>K081452 Synthes Stainless Steel Elastic Intramedullary Nail System</li><li>K143276 Accumed Small Bone IM Nail System</li></ul> | {4}------------------------------------------------ The subject device is substantially equivalent to similar previously cleared devices. | Device Description: | The Renovo Life Small Bone IM Nail System nails are available in<br>diameters ranging from 2.5 mm to 4.5 mm with a length of 150 mm and<br>either solid or cannulated and materials are surgical grade stainless steel<br>or titanium.<br>Renovo Life Screws are 3.5 mm in diameter and are available in lengths<br>from 10mm to 34 mm in materials of surgical grade stainless steel or<br>titanium.<br>The Renovo Life Small Bone IM Nail System is provided non-sterile and<br>are intended for single use only. | |---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use: | The Renovo Life Small Bone IM Nail System is indicated for fixation of<br>fractures where the canal is narrow or flexibility of the implant is<br>paramount, including the fibula, radius, and ulna, upper extremity<br>fractures in all patients and lower extremity fractures in small-stature<br>patients. This system is also intended to treat metaphyseal and epiphyseal<br>fractures, such as radial neck fractures and is intended for fixation of small<br>long bones, such as carpal and tarsal bones. | | Summary of Technological<br>Characteristics | The Renovo Life Small Bone IM Nail System designed for fixation of<br>stable and unstable fractures in the fibula, radius and ulna. | | Non-Clinical Testing | The Renovo Life Small Bone IM Nail System has undergone comparative<br>testing with a predicate device conducted as per ASTM 1264-14, a<br>cadaveric implant trial and a rotational resistance test. The test data<br>showed the Renovo Life Small Bone IM Nail was substantially equivalent<br>to the predicate device in a static and dynamic four-point bend tests, static<br>torsion test and compliance tests, as described herein. | | Clinical Testing | No clinical testing was performed. | | Conclusion | The Renovo Life Small Bone IM Nail Systems have the same intended use<br>and similar indications, principles of operation, and technological<br>characteristics as the predicate devices.<br>Thus, the Renovo Life Small Bone IM Nail Systems are substantially<br>equivalent to the predicate devices. |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...