The Ellipse PRECICE Trauma Nail System is indicated for open and closed fracture fixation, pseudoarthrosis, or mal-unions and nonunions of long bones.
Device Story
Intramedullary fixation system for long bone fractures; comprises titanium nail with internal rare earth magnet, telescoping lead screw/nut assembly, and planetary gearing; includes locking screws and external remote controller (ERC). Device inserted into intramedullary canal by surgeon; secured with locking screws. Non-invasive adjustment performed via ERC; magnet-driven mechanism enables controlled distraction or compression for fracture reduction. Benefits include precise, non-invasive limb adjustment and stabilization. Used in clinical settings; operated by physicians.
Clinical Evidence
Bench testing only. Included design verification for retraction/distraction forces and cadaveric design validation for humeral insertion/fixation. Relies on previously submitted mechanical testing (ASTM F1264-03), sterilization validation (ISO 11137-2), shelf-life/packaging testing, O-ring seal performance, and biocompatibility (ISO 10993-1).
Technological Characteristics
Titanium intramedullary nail; telescoping lead screw/nut assembly; planetary gearing; rare earth magnet. Dimensions/lengths vary by anatomy (femur, tibia, humerus). Sterilization: Gamma radiation (ISO 11137-2). Connectivity: Non-invasive magnetic coupling with external remote controller. Mechanical testing per ASTM F1264-03.
Indications for Use
Indicated for patients requiring fixation of open or closed fractures, pseudoarthrosis, mal-unions, or non-unions of long bones (femur, tibia, humerus).
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring a symbol that resembles a human figure or a caduceus, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2015
Ellipse Technologies Incorporated John McIntyre Vice President, Regulatory, Clinical, Quality Affairs 13900 Alton Parkway Suite 123 Irvine, California 92618
Re: K152370
Trade/Device Name: PRECICE Trauma Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: August 20, 2015 Received: August 21, 2015
Dear John McIntyre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Casey Hanley -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K152370
Device Name
PRECICE® Trauma Nail System
Indications for Use (Describe)
The Ellipse PRECICE Trauma Nail System is 15270indicated for open and closed fracture fixation, pseudoarthrosis, or mal-unions and nonunions of long bones.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Ellipse. The logo consists of three overlapping circles on the left, with the word "ELLIPSE" in blue to the right of the circles. The circles are colored in shades of blue and green. The word "ELLIPSE" is in a sans-serif font.
101 Enterprise, Suite 100 Aliso Viejo, CA 92656-2604 Office: 949-837-3600 949-837-3664 Fax" www.ellispe-tech.com
## PRECICE Trauma Nail System 510(k) Summary August 2015
- 1. Company: Ellipse Technologies, Incorporated 101 Enterprise Suite 100 Aliso Viejo, CA 92656
- Contact: John McIntyre VP, Regulatory, Clinical and Quality Affairs Phone: (949) 837-3600 x203 Fax: (949) 837-3664
Date of Submission: August 19, 2015
- 2. Proprietary Trade Name: PRECICE Trauma Nail System
- Classification Name: Intramedullary Fixation Rod (21 CFR 888.3020) 3.
- 4. Product Code: HSB (Rod, Fixation, Intramedullary and Accessories)
5. Product Description: The PRECICE Trauma Nail System is composed of the PRECICE Trauma Nail (supplied sterile), locking screws, surgical instruments and an external remote controller (ERC or ERC 2P). The Nail is available in various diameters, lengths and screw-hole configurations to accommodate a variety of patient anatomies. The locking screws are also available in a variety of diameters and lengths. The PRECICE Trauma Nail is supplied sterile by gamma radiation while the locking screws and instruments are supplied non-sterilized prior to use. The Nail contains an enclosed rare earth magnet, telescoping lead screw/nut assembly, and planetary gearing. The Nail is supplied pre-distracted by 10 mm (femur and tibia models), 15 mm and 20 mm (humeral model) to allow for compression fracture reduction techniques.
Indications: The Ellipse PRECICE Trauma Nail System is indicated for open and closed fracture 6. fixation, pseudoarthrosis, or mal-unions and non-unions of long bones.
7. Substantial equivalence: A detailed comparison to the predicate device demonstrates that the Ellipse PRECICE Trauma Nail System is substantially equivalent to the following 510(k) cleared device:
| Trade Name: | PRECICE® Trauma Nail System |
|--------------------------|-----------------------------|
| Common Name: | Intramedullary Fixation Rod |
| 510(k) Clearance Number: | K142599 |
{5}------------------------------------------------
K152370 - Page 2 of 2
Image /page/4/Picture/1 description: The image shows the logo for Ellipse. On the left side of the logo are three overlapping circles in shades of blue and green. To the right of the circles is the word "ELLIPSE" in a bold, sans-serif font. The word is in a dark blue color.
101 Enterorise, Suite 100 Aliso Vieio. CA 92656-26 949-837-3600 949-837-366 www.ellispe-tech.com
Substantial equivalence is based on same indications for use, technological characteristics, principles of operation, designs, and on in vitro testing performed.
The modified PRECICE Trauma Nail System subject to this Special 510(k) and the predicate system have the same indications for use. Specifically, both systems are indicated for open and closed fracture fixation, pseudoarthrosis, or mal-unions and non-unions of long bones. This Special 510(k) is intended to include models specific for the humerus.
The modified Ellipse PRECICE Trauma Nail System has similar technological characteristics and the same principles of operation as that of the predicate. Both the current PRECICE Trauma Nails and the additional Humeral Trauma Nail models subject of this Special 510(k) are titanium intramedullary nails with a telescoping portion that can adjust the limb using principles of distraction osteogenesis. Both devices are inserted into the intramedullary canal of the long bone and secured with PRECICE locking screws. Both devices are adjusted non-invasively by the Ellipse External Remote Controller (ERC). The differences between the two devices are dimensional to better fit the anatomy and fixation requirements of the humerus, and the pre-distracted length of the device.
Non-clinical testing of the PRECICE Trauma Nail System included design verification testing to ensure the device meets the retraction (compression) force and distraction force required for the humerus and cadaveric design validation testing to ensure the device can be inserted into the medullary canal and fixed to the humerus using the humeral specific instruments. The following specific performance tests were completed on the PRECICE Trauma Nail System in order to establish equivalence to the predicate device:
| Test Description | Applicable Test Standard |
|---------------------------------------|--------------------------|
| Device Functionality and Verification | N/A |
| Design Validation | N/A |
Non-clinical testing that was submitted for the predicate PRECICE Trauma Nail System which are applicable to the subject device include mechanical testing according to the methods outlined in the standard ASTM F1264-03, validation of the gamma radiation sterilization cycle in accordance with the VDmas methodology as given in ISO 11137-2 to verify that the gamma radiation sterilization process provides a sterility assurance level of 10-, shelf life testing for the packaging after accelerated aging, Oring seal performance testing and biocompatibility in accordance with ISO 10993-1 for the intended use of the device.
There are no changes to the design of the ERC being made as a result of this submission, therefore all testing that was submitted for the ERC in the predicate PRECICE Trauma Nail System is applicable to the modified PRECICE Trauma Nail System.
Conclusions can be drawn from these tests that the modified Trauma Nail System with the Humeral Trauma Nail line extension is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.