PICCOLO COMPOSITE NAILING SYSTEM

K132774 · CarboFix Orthopedics , Ltd. · HSB · Oct 3, 2013 · Orthopedic

Device Facts

Record IDK132774
Device NamePICCOLO COMPOSITE NAILING SYSTEM
ApplicantCarboFix Orthopedics , Ltd.
Product CodeHSB · Orthopedic
Decision DateOct 3, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Piccolo Composite Humeral, Proximal Humerus, Tibial and Femoral Nails Indications include simple fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathological fractures; reconstruction, following tumor resection and grafting. The nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated. Piccolo Composite Ankle Arthrodesis Nails Indicated for tibiotalocalcaneal arthrodesis (fusion). Specific indications include: Avascular necrosis of the talus; Failed total ankle arthroplasty; Trauma (malunited tibial pilon fracture); Severe deformity or instability as a result of talipes equinovarus, cerebral vascular accident, paralysis or other neuromuscular disease; Revision ankle arthrodesis; Neuroarthropathy; Rheumatoid arthritis; Osteoarthritis; Pseudoarthrosis; Post-traumatic arthrosis; Previously infected arthrosis; Charcot foot; Severe endstage degenerative arthritis; Severe defects after tumor resection; Pantalar arthrodesis

Device Story

Intramedullary nailing system for orthopedic fracture fixation and ankle arthrodesis. System comprises carbon fiber reinforced polyetheretherketone (CFR-PEEK) nails and titanium-alloy interlocking screws. Nails feature tantalum markers for fluoroscopic visualization. Used by orthopedic surgeons in clinical/surgical settings. Device provides mechanical stabilization of long bone fractures or fusion of ankle joint. Surgeon inserts nail into medullary canal and secures with screws. Output is physical stabilization of bone segments; aids healing and patient mobility. No software or electronic processing involved.

Clinical Evidence

No clinical data provided. Substantial equivalence based on bench testing and comparison to predicate devices.

Technological Characteristics

Materials: Carbon fiber reinforced polyetheretherketone (CFR-PEEK) nails, titanium-alloy screws. Features: Tantalum markers for fluoroscopic visualization. Form factor: Intramedullary nails with proximal/distal screw holes. Mechanical fixation principle. No energy source or software.

Indications for Use

Indicated for patients requiring fixation of long bone fractures (humeral, proximal humerus, tibial, femoral) including simple, comminuted, spiral, oblique, segmental, nonunions, malunions, polytrauma, pathological fractures, and tumor resection reconstruction. Also indicated for tibiotalocalcaneal arthrodesis in patients with avascular necrosis, failed arthroplasty, trauma, severe deformity/instability (e.g., talipes equinovarus, paralysis), revision arthrodesis, neuroarthropathy, arthritis, pseudoarthrosis, or post-tumor resection defects.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132774 # 510(K) Summary CarboFix Orthopedics Ltd. ## Piccolo Composite® Nailing Systems ## Applicant Name CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 46724, Israel ## Contact Person Yael Rubin CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzliya 46724, Israel Tel: +972 9 9511511, Fax: +972 9 9548939 # Date Prepared September 2013 # Trade/Proprietary Name Piccolo Composite® Nailing System ## Common Name Intramedullary Nailing System ### Classification Name Rod, Fixation, Intramedullary and Accessories (Class II, per 21 CFR §888.3020; Product Codc HSB) ОСТ 0 3 2013 {1}------------------------------------------------ #### Predicate Devices - Piccolo Composite® Nailing Systems (CarboFix Orthopedics Ltd.; K091425, . K100497, K102369, K111056, K123810) - Piccolo Composite Plate Systems (CarboFix Orthopedics Ltd.; K130061, and ■ more) #### Intended Use/Indications for Use #### Piccolo Composite Humeral, Proximal Humerus, Tibial and Femoral Nails Indications include simple fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathological fractures; reconstruction, following tumor resection and grafting. The nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated. #### Piccolo Composite Ankle Arthrodesis Nails - Indicated for tibiotalocalcaneal arthrodesis (fusion). Specific indications include: Avascular necrosis of the talus; Failed total ankle arthroplasty; Trauma (malunited tibial pilon fracture); Severe deformity or instability as a result of talipes equinovarus, cerebral vascular accident, paralysis or other neuromuscular discase; Revision ankle arthrodesis; Neuroarthropathy; Rheumatoid arthritis; Osteoarthritis; Pseudoarthrosis; Post-traumatic arthrosis; Previously infected arthrosis; Charcot foot; Severe endstage degencrative arthritis; Severe defects after tumor resection; Pantalar arthrodesis #### System Description The Piccolo Composite Nailing Systems include nails, interlocking screws and instrumentation sets. The nails are made, in general, of carbon fiber reinforced polyetheretherketone (CFR-PEEK), and are marked with tantalum markers, to provide for their visualization under fluoroscopy. The nails provide for holes at the proximal and distal sections, designed for the insertion of the titanium-alloy-made screws. {2}------------------------------------------------ #### CarboFix Orthopedics Ltd. Piccolo Composite" Nailing System There is no change to the general description of the system as compared to the predicate devices. #### Substantial Equivalence The Piccolo Composite Nailing Systems intended use, design, materials, technological characteristics, and principles of operation are substantially equivalent to those of the predicate devices, as applicable. Evaluation of expected performance characteristics and MR Conditional labeling parameters for the Piccolo Composite Nailing Systems was based on comparison to predicate devices. All the above demonstrate that the device is safe and effective for its intended use. {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10403 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002 October 3, 2013 CarboFix Orthopedics Limited Ms. Yael Rubin Director of Regulatory Affairs 11 Ha'hoshlim Street Herzeliya, 46724 ISRAEL Re: K132774 Trade/Device Name: Piccolo Composite™ Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: September 1, 2013 Received: September 5, 2013 Dear Ms. Rubin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ Page 2 - Ms. Yael Rubin device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Erin | Keith for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(K) Number (if known): Device Name: Piccolo Composite® Nailing System Indications for Use: Piccolo Composite Humeral, Proximal Humerus, Tibial and Femoral Nails Indications include simple fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathological fractures; reconstruction, following turnor resection and grafting. The nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated. ### Piccolo Composite Ankle Arthrodesis Nails Indicated for tibiotalocalcaneal arthrodesis (fusion). Specific indications include: Avascular necrosis of the talus: Failed total ankle arthroplasty; Trauma (malunited tibial pilon fracture); Severe deformity or instability as a result of talipes equinovarus, cerebral vascular accident, paralysis or other neuromuscular disease; Revision ankle arthrodesis; Neuroarthropathy: Rheumatoid arthritis: Osteoarthrosis; Post-traumatic arthrosis; Previously infected arthrosis; Charcot foot; Severe endstage degenerative arthritis; Severe defects after tumor resection; Pantalar arthrodesis Prescription Use _ _ Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Casey L. Hanley, Ph.D. Division of Orthopedic Devices
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...