← Product Code [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB) · K131984

# SYNTHES SYNSONIC ULNA NAIL (K131984)

_Synthes (Usa) · HSB · Mar 31, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K131984

## Device Facts

- **Applicant:** Synthes (Usa)
- **Product Code:** [HSB](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB.md)
- **Decision Date:** Mar 31, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3020
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Synthes Synsonic Ulna Nail System is intended to treat shaft fractures of the ulna, impending pathological fractures, malunions, and nonunions.

## Device Story

The Synthes Synsonic Ulna Nail System is an orthopedic implant system for treating ulnar shaft fractures. It consists of intramedullary nails, end caps, and locking tacks made of PEEK-Optima polymer and stainless steel. The system utilizes an ultrasonic generator, handpiece, and footswitch to deliver ultrasonic energy to fasten locking tacks to the nail, creating a fixed-angle construct for rotational alignment, torsional stability, and axial control. The handpiece includes a force sensor that provides feedback to the surgeon; an audible tone indicates when the correct pressure range is achieved for fixation. The device is used by surgeons in a clinical setting. The fixed-angle construct benefits the patient by stabilizing the fracture site during the healing process.

## Clinical Evidence

Bench testing only. Testing included static cantilever bend, cyclic torsion, tack removal, nail removal, and thermal profile testing to analyze welding-induced temperatures against clinical literature criteria. Electrical safety was verified per IEC 60601-1.

## Technological Characteristics

Materials: PEEK-Optima polymer and implant-grade stainless steel. Principle: Ultrasonic energy delivery for tack fixation. Components: Intramedullary nail, end caps, locking tacks, ultrasonic generator, handpiece, footswitch. Features: Force sensor for load feedback, audible tone for pressure indication. Electrical safety: IEC 60601-1 compliant.

## Regulatory Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

## Predicate Devices

- Synthes Titanium Elastic Intramedullary Nail System ([K042135](/device/K042135.md))
- Synthes Olecranon Osteotomy Nail System ([K073402](/device/K073402.md))
- Bonutti Research, Inc. Unity Ultrasonic System ([K090175](/device/K090175.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a stylized image inside. A small trademark symbol is located to the right of the word.

MAR 3 1 2014

# 6.0 - 510(k) Summary

Date Prepared: June 28, 2013

| Sponsor:                | Synthes<br>Angela F. Lassandro<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>Office: (610) 719-6854<br>Fax: (484)-356-9682                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary<br>Name:    | Synthes Synsonic Ulna Nail System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Classification:         | Classification: 888.3020<br>Product Code: HSB                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate<br>Device:    | Synthes Titanium Elastic Intramedullary Nail System (K042135)<br>Synthes Olecranon Osteotomy Nail System (K073402)<br>Bonutti Research, Inc. Unity Ultrasonic System (K090175)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device<br>Description:  | The Synthes Synsonic Ulna Nail System consists of intramedullary nails with<br>end caps and locking tacks designed to treat fractures of the ulna. The system<br>makes use of an ultrasonic generator to deliver energy through a handpiece to<br>securely fasten the locking tacks to the surface of the nail. The handpiece<br>features a force sensor mechanism that provides feedback to the surgeon on<br>the amount of load applied to the locking tack. An audible tone indicates<br>when the correct pressure range is achieved to allow fixation of the tack to the<br>nail.<br><br>The ulnar nail and locking tack implants are composed of PEEK-Optima® polymer material and implant grade stainless steel. Locking of the tack to the nail creates a fixed angle construct that provides rotational alignment,torsional stability and axial control of the fractured ulna. |
| Indications<br>for Use: | The Synthes Synsonic Ulna Nail System is intended to treat shaft fractures of<br>the ulna, impending pathological fractures, malunions, and nonunions.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                         | The summary information presented supports the substantial equivalence of                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Substantial             | the Synthes Synsonic Ulna Nail System to the predicate devices. The                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Equivalence:            | proposed system has the same intended use, similar indications for use and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                         | employs the same technology.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                         | Bench testing was conducted on the subject device construct to support                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                         | substantial equivalence to the predicate devices including:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                         | Static Cantilever Bend Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                         | Cyclic Torsion Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                         | Tack Removal Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                         | Nail Removal Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                         | Design verification testing was conducted as part of the risk analysis to ensure<br>the safe use of the subject device. This testing included:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                         | Thermal Profile Testing: Conducted to analyze temperatures induced<br>by the welding cycle and compared to criteria for thermal bone injury<br>documented in clinical literature.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                         | The ultrasonic generator, handpiece, and footswitch components the proposed<br>Synsonic Ulna Nail System were evaluated for electrical safety in accordance<br>with the IEC 60601-1.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                         | The testing evaluated all of the components of proposed system, including the<br>Synthes Synsonic Ultrasonic Generator, Handpiece, and Footswitch.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                         | The results of the bench testing indicate that the performance of the proposed<br>Synsonic Ulna Nail System support its substantial equivalence to the<br>predicate devices and will present no new issues with regard to safety or<br>effectiveness. Additionally, the results of the safety tests conducted on the<br>subject system met the established criteria and conform to the requirements of<br>the relevant standards.                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the word "SYNTHES" in a bold, sans-serif font. To the left of the word is a circular logo with a stylized design. A registered trademark symbol is located to the upper right of the word "SYNTHES".

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.

Public Health Service

ood and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 31, 2014

Synthes Ms. Angela F. Lassandro Regulatory Affairs Manager 1301 Goshen Parkway West Chester, Pennsylvania 19380

Re: K131984

> Trade/Device Name: Synthes Synsonic Ulna Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: March 7, 2014 Received: March 10, 2014

#### Dear Ms. Lassandro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Angela F. Lassandro

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/Medica|Devices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

# Lori A. Wiggins

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "SYNTHES" in bold, black font. To the left of the word is a logo that appears to be a stylized image of a flower or plant. To the right of the word is a small circle, likely indicating a registered trademark.

### 5.0 - Indications for Use Statement

K131984 510(k) Number (if known):

Device Name: Synthes Synsonic Ulna Nail System

Indications for Use:

The Synthes Synsonic Ulna Nail System is intended to treat shaft fractures of the ulna. impending pathological fractures, malunions, and nonunions.

Prescription Use _ X (Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Casey L. Hanley, Ph.D.

Division of Orthopedic Devices

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K131984](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSB/K131984)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
