ZIMMER NATURAL NAIL SYSTEM RETROGRADE FEMORAL NAILS
K101622 · Zimmer, Inc. · HSB · Oct 1, 2010 · Orthopedic
Device Facts
Record ID
K101622
Device Name
ZIMMER NATURAL NAIL SYSTEM RETROGRADE FEMORAL NAILS
Applicant
Zimmer, Inc.
Product Code
HSB · Orthopedic
Decision Date
Oct 1, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone. Indications for use of the retrograde femoral nails in the femur include: Compound and simple shaft fractures Proximal, metaphyseal, and distal shaft fractures Segmental fractures Closed supracondylar fractures Severely comminuted supracondylar fractures with articular involvement Fractures involving femoral condyles Comminuted fractures Fractures involving osteopenic and osteoporotic bone Pathological fractures Fractures with bone loss Pseudoarthrosis, non-union, and mal-union Periprosthetic fractures Poly trauma patients
Device Story
The Zimmer Natural Nail System Retrograde Femoral Nails are temporary intramedullary fixation rods used for femoral fracture stabilization. The system consists of nails, screws, nail caps, cortical washers, and cortical nuts. The device is inserted into the femur in a retrograde mode; screws are passed through proximal and distal holes to secure the nail. Used by orthopedic surgeons in a clinical/surgical setting. The device provides mechanical stabilization to facilitate bone healing. Benefits include fixation for various fracture types, including those in compromised bone (osteopenic/osteoporotic).
Clinical Evidence
No clinical data was required or provided for this device. Substantial equivalence was demonstrated through non-clinical bench testing, including proximal and distal nail fatigue tests, midshaft analysis, torsional stiffness analysis, screw bending fatigue, insertion load/torque tests, and shelf-life testing of UHMWPE components.
Technological Characteristics
Intramedullary fixation rod system. Materials: Ti-6Al-4V alloy; UHMWPE (cortical nut). Design: Retrograde femoral nail with proximal/distal screw fixation. Mechanical stabilization principle. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring temporary fracture fixation and stabilization of the femur, including compound/simple shaft, proximal/metaphyseal/distal shaft, segmental, closed/comminuted supracondylar, condylar, pathological, and periprosthetic fractures, as well as fractures with bone loss, pseudoarthrosis, non-union, mal-union, and poly trauma. Suitable for use in osteopenic and osteoporotic bone.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
DePuy ACE Universal and Troch Entry Femoral Nail Systems (K033329)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Zimmer. The logo consists of a large, bold letter "Z" inside of a circle. Below the circle is the word "zimmer" in a smaller, sans-serif font.
K101622 lg2
P.O. Box 708 Warsaw, IN 46581-0708 574 267-6131
## 510(k) Summary
TT 1 2011
Sponsor:
Contact Person:
Date:
Trade Name:
Common Name:
Classification Name and Reference:
Predicate Devices:
Device Description:
Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708
Daniel J. Williman, RAC Specialist, Regulatory Affairs Telephone: (574) 371-8065 Fax: (574) 372-4605
September 1, 2010
Zimmer® Natural Nail™ System Retrograde Femoral Nails
Intramedullary Fixation Rod
Intramedullary fixation rod, product code HSB 21 CFR § 888.3020
Supracondylar Intramedullary Nail, manufactured by Zimmer, Inc. (K962561, cleared September 25, 1996)
DePuy ACE Universal and Troch Entry Femoral Nail Systems, manufactured by DePuy ACE (K033329, cleared November 14, 2003)
The Zimmer Natural Nail System Retrograde Femoral Nails are a family of temporary fixation intramedullary nails designed for fracture fixation and stabilization of the femur. They are inserted into the femur in a retrograde mode and are available in a variety of lengths and diameters to meet assorted anatomical needs. Each of the intramedullary nails is secured by a series of screws that pass through holes manufactured into the proximal and distal sections of each nail. Nail caps, cortical washers and cortical nuts are available for use with the system. All components are
71
{1}------------------------------------------------
K101622 Pg 2g2
manufactured from Ti-6Al-4V alloy. The cortical nut also contains UHMWPE.
The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone.
Indications for use of the retrograde femoral nails in the femur include:
- . Compound and simple shaft fractures
- Proximal, metaphyseal, and distal shaft fractures
- Segmental fractures
- Closed supracondylar fractures ●
- Severely comminuted supracondylar fractures ● with articular involvement
- . Fractures involving femoral condyles
- Comminuted fractures �
- Fractures involving osteopenic and osteoporotic ● bone
- . Pathological fractures
- . Fractures with bone loss
- . Pseudoarthrosis, non-union, and mal-union
- Periprosthetic fractures
- . Poly trauma patients
The Zimmer Natural Nail system is similar or identical in intended use, materials, sterility, and performance characteristics to the predicate device(s).
Non-Clinical Performance and Conclusions: The results of non-clinical (lab) performance testing demonstrate that the devices are safe and effective and substantially equivalent to the predicate devices. Performance testing included: Proximal Nail Fatigue Test, Distal Nail Fatigue Test, Midshaft Nail Analysis, Nail Torsional Stiffness Analysis, Screw Bending Fatigue Test, Screw Insertion Load Test, Screw Insertion Torque Test, Screw Torque to Failure Test, and Shelf Life Testing of UHMWPE.
Clinical Performance and Conclusions: Clinical data and conclusions were not needed for this device.
# Intended Use:
Comparison to Predicate Device:
Performance Data (Nonclinical and/or Clinical):
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The seal is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Zimmer Inc. % Mr. Daniel J. Williman, RAC Specialist, Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581
OCT 4 2010
Re: K101622
Trade/Device Name: Zimmer® Natural Nail™ System Retrograde Femoral Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: September 30, 2010 Received: October 1, 2010
Dear Mr. Williman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
# Page 2 - Mr. Daniel J. Williman, RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
For Doth Dih
Nir
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Zimmer. The logo consists of a circle with a stylized "Z" inside, followed by the word "zimmer" in a bold, sans-serif font. The "Z" in the circle is black, and the word "zimmer" is also black.
# Indications for Use
## 510(k) Number (if known):
#### Device Name:
Zimmer® Natural Nail™ System Retrograde Femoral Nail
## Indications for Use:
The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone.
Indications for use of the retrograde femoral nails in the femur include:
- Compound and simple shaft fractures .
- Proximal, metaphyseal, and distal shaft fractures .
- Segmental fractures .
- . Closed supracondylar fractures
- Severely comminuted supracondylar fractures with articular involvement .
- . Fractures involving femoral condyles
- Comminuted fractures .
- Fractures involving osteopenic and osteoporotic bone .
- . Pathological fractures
- Fractures with bone loss .
- Pseudoarthrosis, non-union, and mal-union .
- . Periprosthetic fractures
- Poly trauma patients .
Prescription Use X (Part 21 CFR 801 Subpart D)
XM
## AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Division Sigr Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K101622
13
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.