K040929 · Smith & Nephew, Inc. · HSB · May 25, 2004 · Orthopedic
Device Facts
Record ID
K040929
Device Name
TRIGEN ADOLESCENT TAN
Applicant
Smith & Nephew, Inc.
Product Code
HSB · Orthopedic
Decision Date
May 25, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated. In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability are indicated for the following: subtrochanteric fractures with lesser trochanteric involvement; ipsilateral femoral shaft/neck fractures; and intertrochanteric fractures. The Smith & Nephew, Inc. TriGen Adolescent TAN is for single use only.
Device Story
TriGen Adolescent TAN is an intramedullary nail system used for orthopedic fracture fixation. Device is implanted surgically by a physician to stabilize long bone fractures, including femoral fractures requiring compression and rotational stability via proximal screws. System provides mechanical support to bone segments during healing. Device is for single use only.
Clinical Evidence
Bench testing only. Mechanical test data demonstrated the device is equivalent to currently marketed devices and capable of withstanding expected in vivo loading without failure.
Technological Characteristics
Intramedullary fixation rod; metallic construction; single-use; design includes proximal holes/slots for screws to provide compression and rotational stability.
Indications for Use
Indicated for patients with simple, comminuted, spiral, oblique, or segmental long bone fractures; nonunions; malunions; polytrauma; impending pathologic fractures; tumor resection/grafting; supracondylar fractures; and bone lengthening/shortening. Also indicated for subtrochanteric fractures with lesser trochanteric involvement, ipsilateral femoral shaft/neck fractures, and intertrochanteric fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
K040212 — TRIGEN INTERTAN · Smith & Nephew, Inc. · Feb 20, 2004
K051557 — TRIGEN RETROGRADE FEMORAL, SUPRACONDYLAR AND TIBIAL NAILS · Smith & Nephew, Inc. · Jun 30, 2005
K061019 — TRIGEN META-NAIL RETROGRADE FEMORAL AND TIBIAL NAILS · Smith & Nephew, Inc. · Jun 6, 2006
K040462 — TRIGEN TROCHANTERIC ANTEGRADE NAIL · Smith & Nephew, Inc. · Mar 23, 2004
K092748 — SURESHOT TAN NAILS AND ACCESSORIES · Smith & Nephew, Inc. · Apr 26, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Kc40929
page 1 of 1
## MAY 2 5 2004
Smith & Nephew, Inc. Summary of Safety and Effectiveness TriGen Adolescent TAN
Date of Summary: April 8, 2004
Contact Person and Address Kim Kelly Project Manager, Regulatory/Clinical Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 East Brooks Road Memphis, TN 38116 (901) 399-6566
Name of Device: TriGen Adolescent TAN Common Name: Intramedullary Nail
#### Device Classification Name
21 CFR 888.3020 Intramedullary fixation rod - Class II
#### Indications for Use
Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated.
In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability are indicated for the following: subtrochanteric fractures with lesser trochanteric involvement; ipsilateral femoral shaft/neck fractures; and intertrochanteric fractures. The Smith & Nephew, Inc. Adolescent TAN is for single use only,
#### Mechanical and Clinical Data
A review of the mechanical test data indicated that the TriGen Adolescent TAN is equivalent to devices currently used clinically and is capable of withstanding expected in vivo loading without failure.
### Substantial Equivalence Information
The substantial equivalence of the TriGen Adolescent TAN is based on its similarities in indications for use, design features, operational principles, and material composition to the following predicate devices – Smith & Nephew's Titanium Nail System (K981529) and Intramedullary Nail System (K983942).
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized representation of three wavy lines, which symbolize the human services aspect of the department's mission.
Public Health Service
MAY 2 5 2004
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Kim Kelly Project Manager, Clinical/Regulatory Affairs Smith & Nephew, Inc. Orthopaedic Division 1450 East Brooks Road Memphis, Tennessee 38116
Re: K040929
Trade/Device Name: TriGen Adolescent TAN Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: HSB Dated: April 8, 2004 Reccived: April 9, 2004
Dear Ms. Kelly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Ms. Kim Kelly
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# TriGen Adolescent TAN Indications Statement
Indications for interlocking intramedullary nails include simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following turnor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of the long bones being treated.
In addition to the indications for interlocking intramedullary nails, devices that contain holes/slots proximally to accept screws that thread into the femoral head for compression and rotational stability are indicated for the following: subtrochanteric fractures with lesser trochanteric involvement; ipsilateral femoral shaft/neck fractures; and intertrochanteric fractures. The Smith & Nephew, Inc. TriGen Adolescent TAN is for single use only.
Mark N Millhuser
neral, Restorative, and Neurological Devices
510(k) Numbar ._
Concurrence of CDRH, Office of Device Evaluation
Prescription Use
(Per 21 CFR 801.109)
OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.