← Product Code [HRY](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRY) · K961900

# TIBIAL TRAY PLUG, A PRODUCT LINE ADDITION TO THE AXIOM TOTAL KNEE SYSTEM (K961900)

_Wrightmedicaltechnologyinc · HRY · Jul 25, 1996 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRY/K961900

## Device Facts

- **Applicant:** Wrightmedicaltechnologyinc
- **Product Code:** [HRY](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRY.md)
- **Decision Date:** Jul 25, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3530
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The AXIOM™ Total Knee System is indicated for use in total knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions: 1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis; 2) inflammatory degenerative joint disease including rheumatoid arthritis; 3) correction of functional deformity; 4) revision procedures where other treatments or devices have failed; and 5) treatment of fractures that are unmanageable using other techniques.

## Device Story

Tibial Tray Plug acts as filler for screw holes in AXIOM Total Knee System tibial bases; prevents bone cement extrusion when metal base implanted without screw fixation. Device machined from UHMWPE; one-piece design with 3-slotted or 4-slotted styles. Surgeon inserts plugs intraoperatively into screw holes from underside of metal tibial base; snap-fit mechanism. Removable via hemostat if necessary. Supplied sterile; available preassembled or in packs of four. Benefits patient by ensuring proper cement containment during total knee arthroplasty.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Material: Ultra high molecular weight polyethylene (UHMWPE) per ASTM F 648. Design: One-piece, 3-slotted or 4-slotted configuration. Mechanism: Snap-fit into tibial base screw holes. Sterilization: Ethylene oxide.

## Regulatory Identification

A knee joint femorotibial metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Tibial Tray PMMA Plug, Wright Medical Technology, Inc.
- MGII tibial tray plug, Zimmer, Inc.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K961900

JUL 25 1996

W

WRIGHT

MEDICAL TECHNOLOGY, INC.

5677 AIRLINE ROAD

ARLINGTON, TN 38002

901-867-9971

# 510(k) Summary

Contact Person: Kim Tompkins
Date Prepared: May 15, 1996

Trade/Common Name: Tibial Tray Plug
Classification Name: Knee joint femorotibial metal/polymer semi-constrained cemented prosthesis.
Predicate Devices: Tibial Tray PMMA Plug, Wright Medical Technology, Inc.
MGII tibial tray plug, Zimmer, Inc.

This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.

# Description

The Tibial Tray Plug is intended to be used with all the tibial bases that are available in the AXIOM™ Total Knee System. The purpose of the Tibial Tray Plug is to provide a filler for the screw holes to prevent extrusion of bone cement through the holes when the metal base is implanted without screw fixation. The Tibial Tray Plug will be machined from ultra high molecular weight polyethylene (UHMWPE) conforming to ASTM F 648. The one-piece design will be offered in two alternative slot styles, 4-slotted and 3-slotted.

The Tibial Tray Plug will be offered sterile in a pack of four for insertion into one or more of the four screw holes of the tibial base by the surgeon intraoperatively. The Tibial Tray Plug will also be available preassembled into the tibial tray. Tibial Tray Plugs are placed into the screw holes from the underside of the metal tibial base and snap fit snugly into place. If desired, the Tibial Tray Plug can be removed by using an instrument such as a hemostat to push the plug from the top of the base back through the screw hole.

# Indications

The AXIOM™ Total Knee System is indicated for use in total knee arthroplasty for reduction or relief of pain and/or improved knee function in skeletally mature patients with the following conditions:

1) noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis;
2) inflammatory degenerative joint disease including rheumatoid arthritis;
3) correction of functional deformity;
4) revision procedures where other treatments or devices have failed; and
5) treatment of fractures that are unmanageable using other techniques.

# Summary of differences

The Tibial Tray Plug is similar to either of the two predicate devices with respect to material, design, availability, and intended use. The only difference left between the Tibial Tray Plug and both predicate devices is the sterilization method. The predicate devices are sterilized with gamma radiation and the Tibial Tray Plug will be sterilized with ethylene oxide, both well accepted for achieving a sterility assurance level of $10^{-6}$.

000128

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRY/K961900](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRY/K961900)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
