← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K242343

# BPS Wrist Fracture System (K242343)

_Orthonovis, Inc. · HRS · Oct 4, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K242343

## Device Facts

- **Applicant:** Orthonovis, Inc.
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Oct 4, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The BPS Wrist Fracture System is indicated for the fixation of simple and complex intra-articular and extra-articular fractures, and for osteotomies of the distal radius in adults. The device is indicated for fixation of Fractures AO types A2, A3, B1, B3, C1, C2, C3.

## Device Story

BPS Wrist Fracture System consists of metallic bone plates and screws for distal radius fracture fixation; osteotomies; fusions. Plates manufactured from Commercially Pure Titanium Grade 4; screws from Titanium Alloy Grade 5 (Ti-6Al-4V ELI). System includes various plate sizes and locking/non-locking/cortical screw styles to accommodate patient anatomy. Used by orthopedic surgeons in clinical settings. Surgeon selects appropriate hardware to stabilize bone fragments; hardware provides mechanical fixation to facilitate healing. Benefits include restoration of anatomical alignment and stability for distal radius fractures.

## Clinical Evidence

Bench testing only. Testing included static four-point bending (ASTM F382), torsional yield strength (ASTM F543), driving torque (ASTM F543), and axial pullout strength (engineering analysis). All tests met performance criteria outlined in FDA guidance for orthopedic fracture fixation plates and metallic bone screws.

## Technological Characteristics

Materials: Commercially Pure Titanium Grade 4 (ASTM F67) for plates; Titanium Alloy Grade 5 (ASTM F136) for screws. Mechanical fixation via locking/non-locking/cortical screws. Standards: ASTM F382 (bending), ASTM F543 (screws).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- AFFINITY Variable Angle Distal Radius System ([K191641](/device/K191641.md))
- VariAx 2 Distal Radius System ([K233919](/device/K233919.md))
- 2.4mm VA-LCP Volar Distal Radius System ([K083694](/device/K083694.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image is completely white and contains no discernible content. There are no objects, shapes, or patterns visible. The image appears to be a blank canvas.

October 4, 2024

OrthoNovis, Inc. % Hannah Taggart Regulatory Associate Empirical Technologies 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K242343

Trade/Device Name: BPS Wrist Fracture System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: August 7, 2024 Received: August 7, 2024

Dear Hannah Taggart:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# CHRISTOPHER FERREIRA -S

Christopher Ferreira, MS Assistant Director, Biomedical Engineer DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.

Submission Number (if known)

K242343

Device Name

BPS Wrist Fracture System

Indications for Use (Describe)

The BPS Wrist Fracture System is indicated for the fixation of simple and complex intra-articular and extra-articular fractures, and for osteotomies of the distal radius in adults. The device is indicated for fixation of Fractures AO types A2, A3, B1, B3, C1, C2, C3.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## K242343 510(k) Summary

| Submitter's Name:              | OrthoNovis, Inc.                                                                           |
|--------------------------------|--------------------------------------------------------------------------------------------|
| Submitter's Address:           | 1 Hargrove Grade, 2F<br>Palm Coast, Florida 32137                                          |
| Submitter's Telephone:         | 833-950-0127                                                                               |
| Contact Person:                | Hannah Taggart, MS<br>Empirical Technologies<br>719-457-1152<br>htaggart@empiricaltech.com |
| Date Summary was Prepared:     | October 2, 2024                                                                            |
| Trade or Proprietary Name:     | BPS Wrist Fracture System                                                                  |
| Device Classification Name:    | Plate, Fixation, Bone                                                                      |
| Classification & Regulation #: | Class II per 21 CFR §888.3030 and 21 CFR §888.3040                                         |
| Product Code:                  | HRS                                                                                        |
| Additional Product Code:       | HWC                                                                                        |
| Classification Panel:          | Orthopedic - Restorative, Repair and Trauma Devices (DHT6C)                                |

## DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The BPS Wrist Fracture System includes bone plate and bone screw implants designed for fixation of certain fractures, fusions, or osteotomies in the distal radius. To meet various patient anatomical needs, the BPS Wrist Fracture System features several sizes of distal radius plates as well as various styles and sizes of locking, nonlocking, and cortical bone screws. The plates are manufactured from Commercially Pure Titanium Grade 4 per ASTM F67. The screws are manufactured from Titanium Alloy Grade 5 (Ti-6AI-4V ELI) per ASTM F136.

### INDICATIONS FOR USE

The BPS Wrist Fracture System is indicated for the fixation of simple and complex intra- articular and extraarticular fractures, and for osteotomies of the distal radius in adults.

The device is indicated for fixation of Fractures AO types A2, A3, B1, B3, C1, C2, C3.

### TECHNOLOGICAL CHARACTERISTICS

The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. The predicates included in this submission were selected based on the best practices described in the FDA Draft Guidance document Best Practices for Selecting a Predicate Device to Support a Premarket Notification [510(k)] Submission. Predicate clearance information and labeling documentation is shown in Appendix B. Specifically, the following characteristics are the same between the subject and predicates:

- Indications for Use
- Anatomical Site for Use
- . Materials of manufacture
- . Plate Sizes
- Screw Styles and Sizes ●

Minor differences in design do not raise questions for the safety and efficacy of the subject device since the performance testing of the subject device established the substantial equivalence despite those differences.

{5}------------------------------------------------

#### Predicate Devices

| 510k #  | Trade or Proprietary or Model Name           | Manufacturer                      | Product Code | Predicate Type |
|---------|----------------------------------------------|-----------------------------------|--------------|----------------|
| K191641 | AFFINITY Variable Angle Distal Radius System | Industrias Medicas Sampedro S.A.S | HRS, HWC     | Primary        |
| K233919 | VariAx 2 Distal Radius System                | Stryker GmbH                      | HWC          | Additional     |
| K083694 | 2.4mm VA-LCP Volar Distal Radius System      | Synthes (USA)                     | HRS          | Additional     |

#### PERFORMANCE DATA

The BPS Wrist Fracture System has been tested in the following test modes:

- Static Four-Point Bending, conducted per ASTM F382, of the worst-case subject plates met the . performance criteria as outlined in the FDA Guidance 'Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway'.
- Torsional yield strength, conducted per ASTM F543, of the worst-case subject screws met the performance criteria as outlined in the FDA Guidance 'Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathwav'.
- . Driving torque of the worst-case subject screws was conducted per ASTM F543 and met the performance criteria as outlined in the FDA Guidance 'Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway'.
- . Axial Pullout Strength of the worst-case subject screws was evaluated using an engineering analysis, which met the performance criteria as outlined in the FDA Guidance 'Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway'.

The results of this non-clinical testing show that the strength of the BPS Wrist Fracture System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

#### CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the BPS Wrist Fracture System is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K242343](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K242343)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
