← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K233863

# TriMed ASET Foot Plating System (1ST MTP  9-hole Long Plate), TriMed ASET Foot Plating System (1st Met Osteotomy Plate), TriMed ASET Foot Plating System (H-Plate), TriMed ASET Foot Plating System (Talonavicular Plate), TriMed ASET Foot Plating System (Calcaneal Slide Osteotomy Plate), TriMed ASET Foot Plating System (Evans Osteotomy Plate), TriMed ASET Foot Plating System (Medial Column Fusion Plate), TriMed ASET Foot Plating System (Straight Plate), TriMed ASET Foot Plating System (T- (K233863)

_TriMed, Inc. · HRS · Apr 17, 2024 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K233863

## Device Facts

- **Applicant:** TriMed, Inc.
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Apr 17, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The TriMed ASET™ Foot Plating System is indicated for use in stabilization of fractures, revision procedures, joint fusions, reconstructions, deformity corrections, osteotomies, and non-unions of small bones of the feet. Specific examples include: First metatarsal osteotomies for hallux valgus and hallux varus correction including: Opening/Closing base wedge osteotomies, Distal and Proximal Chevron osteotomies, Ludloff Osteotomy, Arthrodesis of the first metatarsophalangeal joint (MTP) including: Primary MTP Fusion due to hallux rigidus and/or hallux valgus, Revision MTP Fusion Revision of failed first MTP Arthroplasty implant Metatarsal and Phalanges: Metatarsal and Phalanges fractures and osteotomies; Mid / Hindfoot Fractures: Cuneiform Fracture, Cuboid Fracture, Navicular Fracture Mid / Hindfoot Fusions: Tarsometatarsal (TMT) Fusions/Stabilization, Intercuneiform Fusions, Navicular Cuneiform (NC) Fusion, Talonavicular (TN) Fusion, Calcaneocuboid (CC) Fusion, Medial Column Fusion, Medial Column Fusion (talus, navicular, cuneiform, metatarsal) for neuropathic osteoarthropathy (Charcot), Lateral Column Fusion (calcaneus, cuboid, metatarsal) for neuropathic osteoarthropathy (Charcot) Mid / Hindfoot Osteotomies (Deformity Corrections): Lateral Column Lengthening (Evans Osteotomy), Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy), Medial Displacement Calcaneal Osteotomy

## Device Story

TriMed ASET Foot Plating System consists of metallic bone plates and screws for orthopedic fixation in forefoot, midfoot, and hindfoot. System provides surgical options for fractures, osteotomies, and arthrodesis. Implants include various plate geometries (e.g., H-Plate, T-Plate, Evans Osteotomy Plate) and variable angle/non-locking screws. Used by orthopedic surgeons in clinical/OR settings to stabilize bone segments, facilitate fusion, or correct deformities. Device provides mechanical support to bone during healing process; benefits patient by restoring structural integrity and enabling bone union. No software or AI components.

## Clinical Evidence

No clinical data. Bench testing only. Static and dynamic four-point bending tests conducted per ASTM F382-17 to evaluate mechanical performance against predicate devices.

## Technological Characteristics

Materials: Implant-grade titanium and titanium alloy. Components: Plates and 2.7mm, 3.5mm, 4.0mm variable angle/non-locking screws. Mechanical testing: Static and dynamic four-point bending per ASTM F382-17. No software, electronics, or energy sources.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- TriMed ASET Foot Plating System ([K192696](/device/K192696.md))
- Baby Gorilla/Gorilla Plating System ([K222194](/device/K222194.md))
- Mondeal Extremity Bone Fixation System ([K072740](/device/K072740.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 17, 2024

TriMed. INC Anderson David Regulatory Affairs Contractor 27533 Avenue Hopkins Valencia, California 91355

### Re: K233863

Trade/Device Name: TriMed ASET Foot Plating System (1ST MTP 9-hole Long Plate), TriMed ASET Foot Plating System (1st Met Osteotomy Plate), TriMed ASET Foot Plating System (H-Plate), TriMed ASET Foot Plating System (Talonavicular Plate), TriMed ASET Foot Plating System (Calcaneal Slide Osteotomy Plate), TriMed ASET Foot Plating System (Evans Osteotomy Plate), TriMed ASET Foot Plating System (Medial Column Fusion Plate), TriMed ASET Foot Plating System (Straight Plate), TriMed ASET Foot Plating System (T-Plate) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: December 5, 2023 Received: December 6, 2023

Dear Anderson David:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{1}------------------------------------------------

## K233863 - Anderson David

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE

{2}------------------------------------------------

by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali-S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

Submission Number (if known)

K233863

Device Name

TriMed ASET Foot Plating System (1ST MTP 9-hole Long Plate); TriMed ASET Foot Plating System (1st Met Osteotomy Plate); TriMed ASET Foot Plating System (H-Plate); TriMed ASET Foot Plating System (Talonavicular Plate); TriMed ASET Foot Plating System (Calcaneal Slide Osteotomy Plate); TriMed ASET Foot Plating System (Evans Osteotomy Plate); TriMed ASET Foot Plating System (Medial Column Fusion Plate); TriMed ASET Foot Plating System (Straight Plate): TriMed ASET Foot Plating System (T-Plate)

Indications for Use (Describe)

The TriMed ASET™ Foot Plating System is indicated for use in stabilization of fractures, revision procedures, joint fusions, reconstructions, deformity corrections, osteotomies, and non-unions of small bones of the feet.

Specific examples include:

First metatarsal osteotomies for hallux valgus and hallux varus correction including:

Opening/Closing base wedge osteotomies, Distal and Proximal Chevron osteotomies, Ludloff Osteotomy,

Arthrodesis of the first metatarsophalangeal joint (MTP) including: Primary MTP Fusion due to hallux rigidus and/or hallux valgus, Revision MTP Fusion Revision of failed first MTP Arthroplasty implant

Metatarsal and Phalanges: Metatarsal and Phalanges fractures and osteotomies;

Mid / Hindfoot Fractures: Cuneiform Fracture, Cuboid Fracture, Navicular Fracture

Mid / Hindfoot Fusions:

Tarsometatarsal (TMT) Fusions/Stabilization, Intercuneiform Fusions, Navicular Cuneiform (NC) Fusion, Talonavicular (TN) Fusion, Calcaneocuboid (CC) Fusion, Medial Column Fusion, Medial Column Fusion (talus, navicular, cuneiform, metatarsal) for neuropathic osteoarthropathy (Charcot), Lateral Column Fusion (calcaneus, cuboid, metatarsal) for neuropathic osteoarthropathy (Charcot)

Mid / Hindfoot Osteotomies (Deformity Corrections):

Lateral Column Lengthening (Evans Osteotomy), Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy), Medial Displacement Calcaneal Osteotomy

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

{4}------------------------------------------------

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

| 510(k) #: | K233863                 |
|-----------|-------------------------|
|           | 510(k) Summary          |
|           | Prepared on: 2024-03-19 |

Contact Details

| Applicant Name              | TriMed, INC                                          |
|-----------------------------|------------------------------------------------------|
| Applicant Address           | 27533 Avenue Hopkins Valencia CA 91355 United States |
| Applicant Contact Telephone | (574) 377-0111                                       |
| Applicant Contact           | David Anderson                                       |
| Applicant Contact Email     | david.anderson@tech2medllc.com                       |

Device Name

| Device Trade Name   | TriMed ASET Foot Plating System (1ST MTP 9-hole Long Plate);<br>TriMed ASET Foot Plating System (1st Met Osteotomy Plate);<br>TriMed ASET Foot Plating System (H-Plate);<br>TriMed ASET Foot Plating System (Talonavicular Plate);<br>TriMed ASET Foot Plating System (Calcaneal Slide Osteotomy Plate);<br>TriMed ASET Foot Plating System (Evans Osteotomy Plate);<br>TriMed ASET Foot Plating System (Medial Column Fusion Plate);<br>TriMed ASET Foot Plating System (Straight Plate);<br>TriMed ASET Foot Plating System (T-Plate) |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name         | Single/multiple component metallic bone fixation appliances and accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification Name | Plate, Fixation, Bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Regulation Number   | 888.3030                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code        | HRS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |

Legally Marketed Predicate Devices

| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K192696     | TriMed ASET Foot Plating System                          | HRS          |
| K222194     | Baby Gorilla/Gorilla Plating System                      | HRS          |
| K072740     | Mondeal Extremity Bone Fixation System                   | HRS          |

Device Description Summary
The TriMed ASET Foot Plating System is a multi-indication total foot plating system which provides surgical options for fractures, osteotomes and arthrodesis of the forefoot, midfoot and hindfoot. Plates and screws are made from implant grade titanium andtomes and arthrodesis of the forefoot, midfoot and hindfoot. Plates and screws are made from implant grade titanium and titanium alloy. Implant offerings are as followed: 1st MTP Osteotomy plates, H-Plate, Talonavicular Plate, Calcaneal Slide Osteotomy Plate, Evans Osteotomy Plate, Medial Column Fusion Plate, Universal Hook plates, and Universal T-Plates. 2.7mm, 3.5mm, and 4.0mm variable angle and non-locking screws are also included.

# Intended Use/Indications for Use

Page 1 of 2

{6}------------------------------------------------

The TriMed ASET™ Foot Plating System is indication of fractures, revision procedures, joint fusions, reconstructions, deformity corrections, osteotomies, and non-unions of small bones of the feet.

Specific examples include:

First metatarsal osteotomies for hallux valgus and hallux varus correction including:

Opening/Closing base wedge osteotomies, Distal and Proximal Chevron osteotomies, Ludloff Osteotomy,

Arthrodesis of the first metatarsophalangeal joint (MTP) including: Primary MTP Fusion due to hallux rigidus and/or hallux valgus, Revision MTP Fusion Revision of failed first MTP Arthroplasty implant

Metatarsal and Phalanges: Metatarsal and Phalanges fractures and osteotomies;

Mid / Hindfoot Fractures: Cuneiform Fracture, Cuboid Fracture, Navicular Fracture

Mid / Hindfoot Fusions:

Tarsometatarsal (TMT) Fusions/Stabilization, Intercuneiform Fusions, Navicular Cuneiform (NC) Fusion, Talonavicular (TN) Fusion, Calcaneocuboid (CC) Fusion, Medial Column Fusion (talus, navicular, cuneiform, metatarsal) for neuropathic osteoarthropathy (Charcot), Lateral Column Fusion (calcaneus, cuboid, metatarsal) for neuropathy (Charcot)

Mid / Hindfoot Osteotomies (Deformity Corrections):

Lateral Column Lengthening (Evans Osteotomy), Plantar Flexion Opening Wedge Osteotomy of the Medial Cuneiform (Cotton Osteotomy), Medial Displacement Calcaneal Osteotomy

## Indications for Use Comparison

The TriMed ASET Foot Plating system has the same Indications for Use as the referenced predicate devices.

## Technological Comparison

TriMed ASET Foot Plating System devices are substantially equivalent to the predicate devices in which basic design features, manufacturing, packaging, and labeling are the same. Any differences between the proposed device and the predicate device are considered minor and do not raise different questions concerning safety or effectiveness.

#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

The TriMed ASET Foot Plating System implants were tested per the recommendations cited in the FDA Guidance Document, Orthopedic Fracture Fixation Plates - Performance Criteria for Safety and Performance Based Pathway: Guidance for Industry and Food and Drug Administration Staff. Static and dynamic four-point bending were conducted per ASTM F382-17.

The TriMed ASET Calcaneal Sliding Ostected for static bending. Testing showed that loading was greater or equal to the referenced predicate device.

Clinical studies were not conducted for the subject devices.

Based on the indications for use, technological characteristics, and the summary of data submitted, TriMed Inc. hat the proposed devices are substantially equivalent to the currently marketed predicate device.

## 21 CFR 807.92(a)(5)

21 CFR 807.92(a)(6)

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K233863](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K233863)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
