← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K210290

# JACE Medical Thoracic Plating System (K210290)

_Jace Medical, LLC · HRS · Apr 30, 2021 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K210290

## Device Facts

- **Applicant:** Jace Medical, LLC
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Apr 30, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The JACE Medical Thoracic Plating System is indicated for use in the stabilization and fixation of fractures of the chest wall including sternal fixation following sternotomy and sternal reconstructive surgical procedures, trauma, or planned osteotomies.

## Device Story

JACE Medical Thoracic Plating System provides rigid fixation for chest wall fractures and sternal closures. System comprises titanium alloy plates (flat, pre-contoured concave/convex) and locking screws. Used by surgeons in clinical/OR settings for sternotomy, reconstructive procedures, or trauma. Plates are manually bent to fit patient anatomy and secured with self-drilling locking screws. Device provides mechanical stabilization to facilitate bone healing. Provided non-sterile; requires steam sterilization by end-user. MR conditional.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including theoretical stress analysis, longitudinal shear-static, lateral distraction-static and cyclic, and four-point bending (ASTM F382-17). Biocompatibility evaluated per ISO 10993:2018. MR safety verified per ASTM F2182-19e2, F2119-13, F2052-15, and F2213-17.

## Technological Characteristics

Materials: Titanium alloy (Ti-6-Al-4 ELI) per ASTM F136. Components: Plates and 3.5mm locking screws. Mechanical testing: ASTM F382-17. MR safety: ASTM F2182-19e2, F2119-13, F2052-15, F2213-17. Sterilization: Steam (non-sterile delivery).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Zimmer Biomet, Sternalock Blu-Biomet Microfixation Sternal Closure System (Sternalock) ([K121302](/device/K121302.md))
- JACE Medical, JACE Sternal Closure System ([K142484](/device/K142484.md))
- Ethicon, Inc., Stainless Steel Suture Wire ([K170767](/device/K170767.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

April 30, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract design, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" on the first line, "U.S. FOOD & DRUG" on the second line, and "ADMINISTRATION" on the third line. The logo is clean and professional, reflecting the organization's role in regulating food and drugs.

JACE Medical, LLC % Dawn N. Norman Partner MRC Global 9085 E. Mineral Circle, Suite 110 Centennial, Colorado 80112

Re: K210290

Trade/Device Name: JACE Medical Thoracic Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: February 1, 2021 Received: February 2, 2021

Dear Dawn Norman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K210290

Device Name

JACE Medical Thoracic Plating System

Indications for Use (Describe)

The JACE Medical Thoracic Plating System is indicated for use in the stabilization and fixation of fractures of the chest wall including sternal fixation following stemotomy and sternal reconstructive surgical procedures, trauma, or planned osteotomies.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

JACE Medical Thoracic Plating System April 01, 2021

#### SUBMITTER INFORMATION 1.

| Name of Sponsor:           | JACE Medical, LLC                       |
|----------------------------|-----------------------------------------|
| Address:                   | 3516 Commerce Drive<br>Warsaw, IN 46580 |
| Primary Contact:           | Dawn Norman, Partner<br>MRC Global, LLC |
| Phone:                     | 618-604-3064                            |
| Email:                     | Dawn.Norman@askMRCGlobal.com            |
| Company/Secondary Contact: | Scott Steffensmeier                     |

| Company/Secondary Contact: | Scott Steffensmeier |
|----------------------------|---------------------|
| Phone:                     | 574-527-9427        |

#### II. DEVICE

| Trade Name:          | JACE Medical Thoracic Plating System                                                             |
|----------------------|--------------------------------------------------------------------------------------------------|
| Common Name:         | Plate, Fixation, Bone                                                                            |
| Classification Name: | Single/multiple component metallic bone fixation<br>appliances and accessories (21 CFR 888.3030) |
| Regulatory Class:    | Class II                                                                                         |
| Panel:               | Orthopedic                                                                                       |
| Product Code:        | HRS                                                                                              |

#### III. PREDICATE DEVICES

Primary Predicate:

- Zimmer Biomet, Sternalock Blu-Biomet Microfixation Sternal . Closure System (Sternalock), K121302
Additional Predicates:

- . JACE Medical, JACE Sternal Closure System, K142484
- . Ethicon, Inc., Stainless Steel Suture Wire, K170767

#### IV. DEVICE DESCRIPTION

The JACE Medical Thoracic Plating System consists of plates and accessories designed to provide fixation following sternotomy and sternal reconstructive 

{4}------------------------------------------------

surgical procedures, trauma, or planned osteotomies. The system includes instrumentation which has been specifically designed for use with this system of implant. The system is comprised of an assortment of various plate configurations, locking screws, and accompanying instrumentation. Variants of the plates include flat and pre-contoured concave and convex plates. Plates may be bent to fit patient anatomy. Plates are secured using previously cleared JACE self-drilling locking bone screws (K142484), which are offered in a single diameter of 3.5mm and have a range of lengths from 10mm to 20mm. There are no changes to the screws and they are not within the scope of this 510(k). Like the predicate device, all implant components are manufactured from titanium alloy (Ti-6-Al-4 ELI) per ASTM F136. All implants in the JACE Medical Thoracic Plating System are intended for single use only. The plates, screws, and all associated instrumentation will be provided nonsterile, to be steam sterilized by the end user. MR compatibility has been established.

## V. INDICATIONS FOR USE

The JACE Medical Thoracic Plating System is indicated for use in the stabilization and fixation of fractures of the chest wall including sternal fixation following sternotomy and sternal reconstructive surgical procedures, trauma, or planned osteotomies.

The Indications for Use statement for the JACE Medical Thoracic Plating System is not identical to the predicate device; however, the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both the subject and primary predicate device have the same basic intended use for the treatment of fractures of the chest wall, by stabilization and fixation through attachment of plates.

#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTIS WITH THE PREDICATE DEVICE

The subject JACE Medical Thoracic Plating System components are substantially equivalent with respect to indications for use, design, dimension, principles of operation, general surgical technique and materials to the primary predicate and the JACE Sternal Closure System, previously cleared by the FDA.

There are insignificant differences between the subject device and primary predicate device. The subject device plate thickness range and screw diameter slightly differs from the predicate device. The subject device has a range of screw lengths of 10-20mm while the primary predicate has screw lengths from 8-20mm. Additionally, the subject device and primary predicate device each have multiple geometric configurations for plate shapes. While many plates have the same or similar configurations, the subject device and primary predicate systems do not contain the exact same configurations.

{5}------------------------------------------------

## VII. PERFORMANCE TESTING

The following performance testing was performed on the subject device in support of substantial equivalence to the primary predicate device.

### Biocompatibility

A biological evaluation has been performed in accordance with ISO 10993:2018. The predicate device is an implanted device that is in contact with tissue and bone for greater than 30 days. The JACE Medical Thoracic Plating System plates are fabricated from Titanium Alloy (ASTM F136) identical to that of the original JACE Medical Sternal Closure System (K142484).

## Magnetic Resonance Testing

Non-clinical testing is provided to support the conditional safety of the JACE Medical Thoracic Plating System components in the MR environment including:

- Radio frequency heating (ASTM F2182-19e2) .
- . lmage artifacts (ASTM F2119-13)
- . Magnetically induced displacement force (ASTM F2052-15)
- . Torque (ASTM F2213-17).

#### Non-Clinical Test Summary

Bench tests were conducted to verify that the subject device met all design specifications:

- . Theoretical Stress Analysis
- Longitudinal Shear-Static .
- Lateral Distraction-Static and Cyclic .
- . Four-Point Bending (ASTM F382-17)

#### Clinical Test Summary

No clinical studies were conducted.

#### VIII. CONCLUSIONS

Based on the information submitted, the JACE Medical Thoracic Plating System is substantially equivalent to the currently marketed devices. The subject device and primary predicate device are made of similar materials and have similar design and characteristics. Therefore, the subject device is as safe, as effective, and performs as well or better than the legally marketed primary predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K210290](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K210290)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
