← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K181091

# AxSOS 3 Ti (K181091)

_Stryker GmbH · HRS · Oct 11, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K181091

## Device Facts

- **Applicant:** Stryker GmbH
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Oct 11, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Intended Use

AxSOS 3 Ti is intended for long bone fracture fixation. Indications include: - Diaphyseal, metaphyseal,epiphyseal,extra- and intra-articular fractures - Non-unions and malunions - Normal and osteopenic bone - Osteotomies - Periprosthetic fractures of the femur and proximal tibia. The AxSOS 3 Ti Waisted Compression Plates are also indicated from fracture fixation of: - Periprosthetic fractures - Diaphyseal and metaphyseal areas of long bones in pediatric patients The 4mm Waisted Compression Plate indications also include fixation of the scapula and the pelvis.

## Device Story

AxSOS 3 Ti is an internal bone fixation system comprising anatomically contoured monoaxial locking plates, screws (locking, cortical, cancellous), blind screws, and auxiliary components. This submission introduces a Variable Angle Extension Arm connector assembly for periprosthetic fractures. The system is used by surgeons in clinical settings to stabilize bone fragments. The device is implanted to provide mechanical support during the bone healing process. The extension arm is affixed to existing plates using a blind screw and secured to bone fragments with a locking screw. The system provides structural stability to facilitate fracture repair and patient recovery.

## Clinical Evidence

No clinical data. Bench testing only. MR compatibility testing performed per ASTM F2052, F2213, F2182, and F2119.

## Technological Characteristics

Materials: Titanium alloy (ASTM F-136) for plates/screws; Variable Angle Extension Arm frame is anodized titanium (Ti-6Al-4V, ASTM F136) with Co-Cr inlay (ASTM F1537). System includes monoaxial locking plates, screws, and extension arms. Non-powered, mechanical fixation device. Sterilization: Sterile or non-sterile.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- AxSOS 3 Ti System ([K172350](/device/K172350.md))

## Reference Devices

- Synthes 3.5mm Locking Attachment Plate ([K083573](/device/K083573.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Stryker GmbH Saad Attiyah Sr. Regulatory Affairs Manager Bohnackerweg 1 Selzach, 2545 Ch

October 11, 2018

#### Re: K181091

Trade/Device Name: AxSOS 3 Ti Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: April 23, 2018 Received: April 24, 2018

Dear Saad Attiyah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## Jesse Muir -S 2018.10.11 22:41:44 -04'00'

For:

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

#### K181091

Device Name AxSOS 3 Ti

Indications for Use (Describe)

AxSOS 3 Ti is intended for long bone fracture fixation.

Indications include:

- * Diaphyseal, metaphyseal,epiphyseal,extra- and intra-articular fractures
- * Non-unions and malunions
- * Normal and osteopenic bone
- * Osteotomies
- * Periprosthetic fractures of the femur and proximal tibia.

The AxSOS 3 Ti Waisted Compression Plates are also indicated from fracture fixation of:

- * Periprosthetic fractures
- * Diaphyseal and metaphyseal areas of long bones in pediatric patients

The 4mm Waisted Compression Plate indications also include fixation of the scapula and the pelvis.

Type of Use (Select one or both, as applicable):

| <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|--------------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>                    |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Proprietary Name:                              | AxSOS 3 Ti                                                                                                                                       |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:                                   | Plate, Fixation, Bone and Screw, Fixation, Bone                                                                                                  |
| Regulation Number &<br>Regulation Description: | 21 CFR 888.3030: Single/multiple component<br>metallic bone fixation appliances and accessories                                                  |
|                                                | 21 CFR 888.3040: Smooth or threaded metallic bone<br>fasteners                                                                                   |
| Product Code:                                  | HRS, HWC                                                                                                                                         |
| Device Class:                                  | Class II                                                                                                                                         |
| Sponsor:                                       | Stryker GmbH<br>Bohnackerweg 1<br>2545 Selzach / Switzerland                                                                                     |
| Contact Person:                                | Saad Attiyah<br>Sr. Regulatory Affairs Manager<br>Bohnackerweg 1<br>2545 Selzach, Switzerland<br>saad.attiyah@stryker.com<br>Phone: 201.831.5655 |
| Date Prepared:                                 | April 23, 2018                                                                                                                                   |
| Primary Predicate:                             | AxSOS 3 Ti System (K172350)                                                                                                                      |
| Reference Predicate:                           | Synthes 3.5mm Locking Attachment Plate<br>(K083573)                                                                                              |

{4}------------------------------------------------

### Description

The AxSOS 3 Ti system (also referred to as "AxSOS 3 Ti Locking Plate System") is a system used for internal fixation applications. The system includes anatomically contoured monoaxial locking plates, screws, blind screws, and auxiliary components. The plates have been designed with holes that can accommodate either locking or non-locking screws, holes for suture and/or Kirschner wires (K-wires). The screws come in 3 types (locking, cortical, and cancellous) and are available in several different diameters and lengths. Implants can be delivered sterile or nonsterile. The various plates and screws are manufactured from titanium alloy (ASTM F-136).

This traditional 510(k) submission is intended to introduce an additional component into the AxSOS 3 Ti System. The Variable Angle Extension Arm connector is an assembly made from an anodized titanium frame (Ti-6Al-4V, ASTM F136) and Co-Cr inlay (ASTM F1537, CoCr28Mo6). These preassembled Variable Angle Extension Arms can be affixed to the previously cleared Distal Lateral Femur Plates, the 5.0 mm narrow waisted compression plates, or the 5.0 mm broad waisted compression plates. The Variable Angle Extension Arms are affixed to the plate using the 4.0 mm blind screw and use a 4.0 mm locking screw to secure it into the desired targeted location of the fragment. These Variable Angle Extension Arms are used for periprosthetic fractures.

#### Indications for Use

AxSOS 3 Ti is intended for long bone fracture fixation.

Indications include:

- Diaphyseal, metaphyseal, epiphyseal, extra- and intra-articular fractures ●
- Non-unions and malunions ●
- Normal and osteopenic bone 0
- Osteotomies
- o Periprosthetic fractures of the femur and proximal tibia.

The AxSOS 3 Ti Waisted Compression Plates are also indicated for fracture fixation of:

- Periprosthetic fractures
- Diaphyseal and metaphyseal areas of long bones in pediatric patients ●

The 4mm Waisted Compression Plate indications also include fixation of the scapula

{5}------------------------------------------------

and the pelvis.

#### Summary of Technologies

A comparison of the systems demonstrated the subject AxSOS 3 Ti system is substantially equivalent to the previously cleared AxSOS 3 Ti system (K172350) with regards to intended use, design, and operational principles.

### Non-Clinical Testing

Bench testing was conducted comparing the cleared plates and screws with the Variable Angle Extension Arms against the predicate and reference devices.

Testing was performed to determine the compatibility of the system in a MR environment. These tests included an assessment of:

- Magnetically Induced Displacement Force per ASTM F2052 ●
- 0 Magnetically Induced Torque per ASTM F2213
- Heating by RF Fields per ASTM F2182 ●
- Image Artifacts per ASTM F2119

Testing demonstrated that the subject device is substantially equivalent to the marketed predicate device.

## Animal Testing

Animal testing was not required for this submission.

## Clinical Testing

Clinical testing was not required for this submission.

#### Conclusion

Documentation is provided demonstrating the AxSOS 3 Ti System is substantially equivalent to the predicate and reference devices in of design, indications for use, technological characteristics, and performance characteristics.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K181091](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K181091)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
