← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K173121

# Ankle Fusion Plating System (K173121)

_In2bones Sas · HRS · Nov 28, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K173121

## Device Facts

- **Applicant:** In2bones Sas
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Nov 28, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The In2Bones Ankle Fusion plating system is indicated for anterior fixation of ankle arthrodesis and fractures, including the distal tibia, talus and calcaneus. The addition of a compression screw through the tibiotalar joint (example IBS™ 6.5mm screw) is required.

## Device Story

Ankle Fusion Plating System provides rigid fixation for ankle arthrodesis and fractures; consists of anatomically contoured titanium plate and screws. Used in hospital or outpatient surgery centers by orthopedic surgeons. Device requires manual bone preparation and screw insertion using dedicated instrument set; includes targeting device for mandatory tibiotalar compression screw. Plate and screws stabilize bone segments to facilitate fusion or fracture healing. Benefits patient by providing mechanical stability during bone healing process.

## Clinical Evidence

Bench testing only. Static and dynamic compression tests performed by independent laboratory. Results met all acceptance criteria. No clinical or animal testing performed.

## Technological Characteristics

Implants (plate and screws) manufactured from Ti6Al4V per ISO 5832-1 and ASTM F136. Anatomically contoured plate with 3.5mm or 4.5mm diameter screws (10-100mm length). Sterile, single-use, gamma-irradiated. Mechanical fixation principle.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Tibiaxys ([K073375](/device/K073375.md))
- Peri-Loc Ankle Fusion Plating System ([K083032](/device/K083032.md))
- Ankle Fusion Plating System ([K141735](/device/K141735.md))
- ExtremiLock Ankle Fusion Plating System ([K133691](/device/K133691.md))
- LCP Ankle Arthrodesis Plate ([K022255](/device/K022255.md))
- Ortholoc3Di Ankle Fusion Plating system ([K121425](/device/K121425.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 28, 2017

In2Bones SAS % Ms. Christine Scifert Executive Vice President MRC-X, LLC 6075 Poplar Ave, Suite 500 Memphis, Tennessee 38119

Re: K173121

Trade/Device Name: Ankle Fusion Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: September 27, 2017 Received: September 29, 2017

Dear Ms. Scifert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

## Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K173121

Device Name Ankle Fusion Plating System

Indications for Use (Describe)

The In2Bones Ankle Fusion plating system is indicated for anterior fixation of ankle arthrodesis and fractures, including the distal tibia, talus and calcaneus.

The addition of a compression screw through the tibiotalar joint (example IBS™ 6.5mm screw) is required.

| Type of Use (Select one or both, as applicable)                                                |                                                                                               |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for IN2BONES. The logo is set on a gray oval shape. The text "IN2BONES" is in white, with the "2" in orange. A white curved line runs from the top left to the center of the oval.

K173121 Page 1 of 3

# 

| Sponsor<br>identification                                                                       | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully – France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29                                                                                                                                                                                                                                                                                                                                                                                       |
|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>registration number                                                            | 3010470577                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Date of preparation                                                                             | November 8, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Contact person                                                                                  | Christine Scifert, MS, MEM<br>MRC-X, LLC<br>6075 Poplar Avenue, suite 500<br>Memphis, TN 38119<br>Cell: 901-831-8053<br>Email: Christine.Scifert@mrc-x.com                                                                                                                                                                                                                                                                                                                                                   |
| Proprietary Name                                                                                | Ankle Fusion Plating System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Common name                                                                                     | Ankle Fusion Plating System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device<br>classification<br>regulation                                                          | 21 CFR 888.3030: Single/multiple component metallic bone<br>fixation appliances and accessories<br>Class II                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Product<br>Code and Panel                                                                | HRS: plate, fixation, bone<br>HWC: screw, fixation, bone<br>87 orthopedics                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Device Description                                                                              | The Ankle Fusion Plating System is composed of an<br>anatomically contoured plate, with a distal part adapted to the<br>talar neck, and a proximal part adapted to the distal tibial<br>epiphysis and diaphysis. Rigid fixation is achieved by screws.                                                                                                                                                                                                                                                       |
|                                                                                                 | The Ankle Fusion Plating System implants (plate and screws) are<br>manufactured from Ti6Al4V. A dedicated instrument set is<br>available for the bone preparation, screw sizing and insertion of<br>the Ankle Arthrodesis Plating System. A specific targeting<br>device is available for the insertion of the additional I.B.S®<br>6.5mm screw.                                                                                                                                                             |
|                                                                                                 | Sizes:<br>The Ankle Fusion Plate is available in one size, with left and<br>right versions.<br>Fixation screws are available in 3.5 or 4.5mm diameter, with<br>length ranging from 10 to 100mm.                                                                                                                                                                                                                                                                                                              |
|                                                                                                 | Material: The Ankle Fusion Plating System implants (plate and<br>screws) are manufactured from Ti6Al4V, as per ISO 5832-1 and<br>ASTM F136.                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                                                 | Single use: The Ankle Fusion Plating System is designed for<br>single use only.                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                 | Sterilization: The Ankle Fusion Plating System is supplied<br>sterile, using gamma irradiation.                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                 | Place of use: The Ankle Fusion Plating System is indicated for<br>use in a hospital, or outpatient surgery center where sterile field<br>may be created and maintained.                                                                                                                                                                                                                                                                                                                                      |
| Predicate Devices                                                                               | Primary Predicate:<br>Tibiaxys (K073375), Newdeal, Acquired by Integra<br>LifeSciences                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                                 | Additional Predicates:<br>Peri-Loc Ankle Fusion Plating System (K083032), Smith and<br>Nephew<br>Ankle Fusion Plating System (K141735), Arthrex<br>ExtremiLock Ankle Fusion Plating System (K133691),<br>Osteomed<br>LCP Ankle Arthrodesis Plate (K022255), Synthes<br>Ortholoc3Di Ankle Fusion Plating system (K121425), WMT                                                                                                                                                                                |
| Indications for use:                                                                            | The In2Bones Ankle Fusion plating system is indicated for<br>anterior fixation of ankle arthrodesis and fractures, including the<br>distal tibia, talus and calcaneus.<br>The addition of a compression screw through the tibiotalar joint<br>(example IBSTM 6.5mm screw) is required.                                                                                                                                                                                                                       |
| Comparison of the<br>indications for use<br>with the predicate<br>devices:                      | The indications for use for the In2Bones Ankle Fusion Plating<br>System are similar to the predicate devices Newdeal Tibiaxys<br>(K073375), Smith and Nephew Peri-Loc Ankle Fusion Plating<br>System (K083032), Arthrex Ankle Fusion Plating System<br>(K141735), Osteomed ExtremiLock Ankle Fusion Plating<br>System (K133691), Synthes LCP Ankle Arthrodesis Plate<br>(K022255), and Wright Medical Ortholoc3Di Ankle Fusion<br>Plating system (K121425)                                                   |
| Comparison of<br>Technological<br>characteristics and<br>Substantial<br>Equivalence<br>Summary: | The In2Bones Ankle Fusion Plating System is similar to the<br>predicate devices Newdeal Tibiaxys (K073375), Smith and<br>Nephew Peri-Loc Ankle Fusion Plating System (K083032),<br>Arthrex Ankle Fusion Plating System (K141735), Osteomed<br>ExtremiLock Ankle Fusion Plating System (K133691), Synthes<br>LCP Ankle Arthrodesis Plate (K022255), and Wright Medical<br>Ortholoc3Di Ankle Fusion Plating system (K121425) in<br>intended use, design, size ranges, principle of operation and<br>materials. |
| Summary<br>Performance Data                                                                     | Performance testing of the Ankle Fusion Plating System was<br>assessed through mechanical bench testing performed by an<br>independent test laboratory, animal and clinical testing being<br>considered not applicable.<br>Testing performed included static and dynamic compression<br>tests on Ankle Fusion Plating System. The results indicate that<br>the Ankle Fusion Plating System met the acceptance criteria.                                                                                      |
| Pyrogen testing                                                                                 | The method used to make the determination that the device meets<br>pyrogen limit specification is the Limulus Amebocyte Lysate<br>(LAL) test in accordance with ANSI/AAMI ST72:2011:<br>Bacterial endotoxins - Test methods, routine monitoring and<br>alternative to batch testing.                                                                                                                                                                                                                         |
| CONCLUSION                                                                                      | Based on the comparison of indications for use and<br>technological characteristics and the results of the testing<br>performed, the Ankle Fusion Plating System is substantially<br>equivalent to the predicate devices identified in the 510(k)<br>submission.                                                                                                                                                                                                                                             |

{4}------------------------------------------------

{5}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K173121](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K173121)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
