← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K121601

# SYNTHES 2.7/3.5MM VARIABLE ANGLE LCP ANKLE TRAUMA SYSTEM-ANTEROLATERAL DISTAL TIBIA PLATE (K121601)

_Synthes USA · HRS · Jul 6, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K121601

## Device Facts

- **Applicant:** Synthes USA
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Jul 6, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System is intended for fixation of the ankle and is indicated in adults and adolescents (12-21) in which the growth plates have fused, and particularly in osteopenic bone. Specifically, • Anterolateral Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia. • Medial and Anteromedial Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia, • Distal Tibia T Plates and Distal Tibia L Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia, and • Lateral Distal Fibula Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal fibula.

## Device Story

System consists of metallic plates and screws for orthopedic fixation of distal tibia and fibula fractures; used in surgical settings by orthopedic surgeons. Plates feature Combi holes allowing for dynamic compression and locking; compatible with various cortical, locking, and variable angle (VA) locking screws. Device provides mechanical stabilization of bone fragments to facilitate healing. Benefits include structural support for osteopenic bone and versatility in screw placement via VA-LCP technology.

## Clinical Evidence

Bench testing only. Engineering analysis and fatigue strength testing were performed to demonstrate mechanical equivalence to predicate devices.

## Technological Characteristics

Metallic bone fixation plates and screws; materials include stainless steel and titanium alloy (TAN). Features Combi holes for dynamic compression and locking. Available in sterile and non-sterile configurations. Variable angle locking technology.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes 2.7/3.5 VA-LCP Ankle Trauma System ([K120854](/device/K120854.md))
- Synthes 4.5/3.5 LCP Metaphyseal Plates ([K033805](/device/K033805.md))
- Zimmer Periarticular and ECT Plate Systems ([K040593](/device/K040593.md), [K042598](/device/K042598.md), [K043227](/device/K043227.md), [K043560](/device/K043560.md), [K030121](/device/K030121.md), [K051098](/device/K051098.md), and Pre-amendment)
- 3.5mm Cortex Screws ([K112583](/device/K112583.md))
- 3.5 mm Locking Screws ([K000684](/device/K000684.md))
- Synthes 2.7mm VA Locking Screws ([K100776](/device/K100776.md))
- 4.5mm VA Curved Condylar Plate Set ([K110354](/device/K110354.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KIZI 601

### 6 2012 JUL

# 510(k) Summary

Date Prepared: June 22, 2012

| Sponsor:                             | Synthes<br>Angela F. Lassandro<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>(610) 719-6854                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:                         | Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System - Anterolateral<br>Distal Tibia Plates                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classification:                      | Classification: Class II, §888.3030, Single/multiple component metallic bone<br>fixation appliances and accessories.<br>Product Code:  HRS, HWC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Predicate Device:                    | Synthes 2.7/3.5 VA-LCP Ankle Trauma System (K120854)<br>Synthes 4.5/3.5 LCP Metaphyseal Plates (K033805)<br>Zimmer Periarticular and ECT Plate Systems (K040593, K042598, K043227,<br>K043560, K030121, K051098, and Pre-amendment)<br>3.5mm Cortex Screws (K112583)<br>3.5 mm Locking Screws (K000684)<br>Synthes 2.7mm VA Locking Screws (K100776)<br>4.5mm VA Curved Condylar Plate Set (K110354)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device<br>Description:               | The Synthes Anterolateral Distal Tibia Plates are intended to treat fractures of<br>the distal tibia. These plates can be used with a new screw configuration,<br>3.5mm VA Locking Screws.<br>The Anterolateral Distal Tibia Plates and 3.5mm VA Locking Screws will be<br>offered in both stainless steel and titanium alloy (TAN), and in both sterile and<br>non-sterile configurations. The Anterolateral Distal Tibia Plates will be offered<br>in left and right configurations.<br>The system accepts existing cortical screws, locking screws, dynamic locking<br>screws, and metaphyseal screws (K100776, K063049, K112583, K000684,<br>K111230, K120070, K110592, and K043185) as well as new 3.5mm VA<br>Locking Screws, and allows for both dynamic compression and locking through<br>Combi holes.                                                                                                                                                                                                                                                                                                                                 |
| Intended Use/<br>Indications for Use | The Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System is intended<br>for fixation of the ankle and is indicated in adults and adolescents (12-21) in<br>which the growth plates have fused, and particularly in osteopenic bone.<br>Specifically,<br>• Anterolateral Distal Tibia Plates are intended for fixation of<br>osteotomies, fractures, nonunions, malunions, and replantations of<br>bones and bone fragments of the diaphyseal and metaphyseal regions<br>of the distal tibia.<br>• Medial and Anteromedial Distal Tibia Plates are intended for fixation<br>of osteotomies, fractures, nonunions, malunions, and replantations of<br>bones and bone fragments of the diaphyseal and metaphyseal regions<br>of the distal tibia,<br>• Distal Tibia T Plates and Distal Tibia L Plates are intended to buttress<br>partial articular fractures and bone fragments of the distal tibia, and<br>• Lateral Distal Fibula Plates are intended for fixation of osteotomies,<br>fractures, nonunions, malunions, and replantations of bones and bone<br>fragments of the diaphyseal and metaphyseal regions of the distal<br>fibula. |
| Substantial<br>Equivalence:          | Both the subject Synthes 2.7/3.5mm VA-LCP Ankle Trauma System and<br>predicate Synthes 2.7/3.5 VA-LCP Ankle Trauma System (K120854), Synthes<br>4.5/3.5 LCP Metaphyseal Plate (K033805) and Zimmer Periarticular and ECT<br>Plate (K040593, K042598, K043227, K043560, K030121, K051098, and Pre-<br>amendment) Systems have similar indications, design characteristics, materials,<br>and performance characteristics. The subject system has been shown to be at<br>least as strong as the predicate devices through engineering analysis and fatigue<br>strength testing. Similarly, the new screws have been shown to be substantially<br>equivalent to existing screws through analysis and dimensional comparison to<br>the predicate 3.5mm Cortex Screws (K112583) and 3.5mm Locking Screws<br>(K000684).                                                                                                                                                                                                                                                                                                                               |

{1}------------------------------------------------

·

・

·

t

.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, suggesting community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure, emphasizing the department's role and national scope.

#### Public Health Service

JUL

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

6 2012

Synthes % Ms. Angela F. Lassandro Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380

Re: K121601

Trade/Device Name: Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System -Anterolateral Distal Tibia Plates

Regulation Number: 21 CFR 888.3030

Regulation Name: Single/multiple component metallic bone fixation appliances and accessories

Regulatory Class: II Product Code: HRS, HWC Dated: May 30, 2012 Received: June 1, 2012

Dear Ms. Lassandro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

Page 2 - Ms. Angela F. Lassandro

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Mark N. Melkerson
Director
Division of Surgical, Orthopedic
and Restorative Devices
Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

### Indications for Use Statement

510(k) Number (if known): ______________________________________________________________________________________________________________________________________________________________________________

Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System Device Name:

Indications for Use:

The Synthes 2.7/3.5mm Variable Angle LCP Ankle Trauma System is intended for fixation of the ankle in adults and adolescents (12-21) in which the growth plates have fused, and particularly in osteopenic bone. Specifically,

- · Anterolateral Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia.
- · Medial and Anteromedial Distal Tibia Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal tibia,
- · Distal Tibia T Plates and Distal Tibia L Plates are intended to buttress partial articular fractures and bone fragments of the distal tibia, and
- · Lateral Distal Fibula Plates are intended for fixation of osteotomies, fractures, nonunions, malunions, and replantations of bones and bone fragments of the diaphyseal and metaphyseal regions of the distal fibula.

Prescription Use _ X (Per 21 CFR 801.109)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DÓ NOT WRITE BÈLOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ast

(Division Sign-Off) Division o: urgical, Orthopedic, and Restorative Devices

K121601 510(k) Number_

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K121601](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K121601)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
