← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K121452

# GRIDLOCK PLATING SYSTEM (K121452)

_Trilliant Surgical, Ltd. · HRS · Aug 20, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K121452

## Device Facts

- **Applicant:** Trilliant Surgical, Ltd.
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Aug 20, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Gridlock Plating System is intended for use in trauma and reconstructive procedures of the small bones in the hand/foot, ankle, and other bones appropriate for the size of the device. The plates (implant), screws (implant), olive wires (instrument), and guide wires (instrument) are intended for single use only.

## Device Story

Gridlock Plating System consists of metallic bone plates and screws for small bone orthopedic fixation; trauma; reconstruction. Features: low profile; limited bone contact; dynamic/manual compression; angulated-locking threaded screw holes. Used by surgeons in clinical settings for bone stabilization. System provides mechanical support to bone fragments during healing. Benefits: facilitates bone alignment and fixation in small bone procedures.

## Clinical Evidence

No clinical studies were performed. Equivalence established via bench testing.

## Technological Characteristics

Materials: CP Titanium (ASTM F67), Titanium alloy (ASTM F136). Components: Plates, screws, olive wires, guide wires. Features: Low profile, limited bone contact, dynamic/manual compression, angulated-locking threaded screw holes. Testing: Static four-point bending (ASTM F382), dynamic bending (ASTM F382), static axial pull-out (ASTM F543), driving torque (ASTM F543), torque to failure (ASTM F543).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- OsteoMed Foot Plating System ([K091614](/device/K091614.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K1214.52

p.1/2

## 510(k) Summa for the Gridlock Plating System

In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the Gridlock Plating System

AUG 20 2012

#### GENERAL INFORMATION 1.

Date Prepared: May 11, 2012

Trade Name: Gridlock Plating System

Common Name: bone plate

Classification Name: Single/multiple component metallic bone fixation appliances and accessories

Class: II

Product Code: HRS / HWC

CFR section: 21 CFR section 888.3030 / 888.3040

Device panel: Orthopedic

Legally Marketed

Predicate Device: OsteoMed Foot Plating System - K091614

Submitter: Trilliant Surgical LTD 602 Sawyer Street, Suite 120

Houston, TX 77007

Contact: J.D. Webb

1001 Oakwood Blvd Round Rock, TX 78681 512-388-0199 Tele 512-692-3699 Fax e-mail: jdwebb@orthomedix.net

### 2. DEVICE DESCRIPTION

The Gridlock Plating System consists of various shape and size plates for the management of small bone orthopedic osteotomies, reconstruction, and trauma. Features include a low profile, limited bone contact, capability of dynamic/manual compression, and angulated-locking threaded screw holes. The system also consists of multiple locking/standard screw lengths and diameters.

### Materials:

CP Titanium per ASTM F67 Titanium alloy per ASTM F136

#### SUBSTANTIAL EQUIVALENCE CLAIMED TO PREDICATE DEVICES 3.

The Gridlock Plating System is substantially equivalent to the predicate device in terms of intended use, design, materials used, mechanical safety and performances.

#### INTENDED USE ধं

The Gridlock Plating System is intended for use in trauma and reconstructive procedures of the small bones in the hand/foot, ankle, and other bones appropriate for the size of the device.

The plates (implant), screws (implant), olive wires (instrument), and guide wires (instrument) are intended for single use only.

{1}------------------------------------------------

#### NON-CLINICAL TEST SUMMARY 5.

The following tests were performed:

## Plates

- Static four point bending per ASTM F382 (Annex A1) .
- . Dynamic bending per ASTM F382 (Annex A2)

Screws

- Static Axial Pull-out per ASTM F543 .
- Driving Torque per ASTM F543 .
- Torque to Failure per ASTM F543 .

The results of this testing indicate that the Gridlock Plating System is equivalent to predicate devices.

### CLINICAL TEST SUMMARY 6.

No clinical studies were performed

### 7. CONCLUSIONS NONCLINICAL AND CLINICAL

This summary includes the conclusions drawn from the nonclinical tests (discussed above) that demonstrate that Gridlock Plating System is as safe, as effective, and performs as well as the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, rendered in black. The overall design is simple and conveys a sense of official government authority.

### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

AUG 2 0 2012

Trilliant Surgical Limited % The OrthoMedix Group, Incorporated Mr. J. D. Webb Authorized Contact Person 1001 Oakwood Boulevard Round Rock, Texas 78681

Re: K121452

Trade/Device Name: Gridlock Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II

Product Code: HRS, HWC Dated: August 9, 2012 Received: August 14, 2012

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

# Page 2 - Mr. J. D. Webb

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Mark A. Millmann

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K121452.

# INDICATIONS FOR USE

510(k) Number (if known):

Device Name: ___ Gridlock Plating System_

Indications for Use:

The Gridlock Plating System is intended for use in trauma and reconstructive procedures of the small bones in the hand/foot, ankle, and other bones appropriate for the size of the device.

The plates and screws are intended for single use only.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Aush

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K121452

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K121452](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K121452)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
