← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K112819

# RECONSTRUCTION LOCKING PLATE AND 3.5MM LOCKING SCREW (K112819)

_Xiamen Double Engline Medical Material Co., Ltd. · HRS · Dec 16, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K112819

## Device Facts

- **Applicant:** Xiamen Double Engline Medical Material Co., Ltd.
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Dec 16, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Reconstruction Locking Plate and 3.5mm Locking Screw are used together and intended to treat fractures of various bones, including the clavicle and pelvis.

## Device Story

The Reconstruction Locking Plate and 3.5mm Locking Screw system is a metallic bone fixation implant used by orthopedic surgeons to treat fractures of the clavicle and pelvis. The system consists of titanium alloy (Ti-6AL-4V) plates of varying lengths and hole counts, paired with 3.5mm self-tapping locking screws. The device is provided non-sterile and must be sterilized by the hospital via autoclave prior to use. The plates and screws provide mechanical stabilization to bone fragments to facilitate healing. The device is used in a surgical setting.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included mechanical verification per ASTM F382-99 (metallic bone plates) and ASTM F543-07 (metallic bone screws) to demonstrate performance equivalent to the predicate.

## Technological Characteristics

Materials: Titanium Alloy (Ti-6AL-4V) per ASTM F136. Components: Reconstruction locking plates (various lengths/holes) and 3.5mm self-tapping locking screws. Sterilization: Moist heat (autoclave) per ISO 17665-1:2006. Mechanical standards: ASTM F382-99, ASTM F543-07.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Syntec Osteo - plate and screw Fixation (k092889)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1/2

Additional Information I K 112819 - Attachment IV 510(k) Summary

DEC 1 6 2011

## Attachment IV 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) Number:

- 1. Date of Submission: September 23, 2011
- 2. Sponsor

Xiamen Double Engine Medical Material Co., Ltd. No.218, Houxiang Road, Haicang District, Xiamen, 361022, China

Contact Person: Jian Wu Position: General manager assistant Tel: 86-0592-6087671 Fax: 86-0592-6587078 Email: xmde.ltd@gmail.com

- 3. Submission Correspondent Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd . P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
- 4. Proposed Device Identification

Proposed Device Name: Reconstruction Locking Plate and 3.5mm Locking Screw

Classification: II Product Code: HRS, HWC Regulation Number: 21 CFR 888.3030, 21 CFR 888.3040 Review Panel: Orthopedic

Intended Use Statement:

The Reconstruction Locking Plate and 3.5mm Locking Screw are used together and intended to treat fractures of various bones, including the clavicle and pelvis.

{1}------------------------------------------------

K112814 Page 2/2

## Additional Information I K112819 - Attachment IV 510(k) Summary

- 5. Predicate Device Identification
510(k) Number: k092889 Product Name: Syntec Osteo - plate and screw Fixation Manufacturer: Syntec Scientific Corporation

- Device Description 6.
The proposed devices of Reconstruction Locking Plate and 3.5mm Locking Screw are intended to treat fractures of various bones, including the clavicle and pelvis.

They are made of Titanium Alloy (Ti-6AL-4V), which meet ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well known biocompatibility.

The plates vary essentially through different lengths and number of plate holes. The screws are self-tapping with 3.5mm major diameter, which are applied with the reconstruction locking plates together.

The proposed devices are not provided sterile. It is required to be sterilized via autoclave method to reach a SAL of 10 by the hospital prior to surgery. The sterilization method is presented in the user manual, which was validated per ISO 17665-1: 2006 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.

- 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

ASTM F 382-99, Standard Specification and Test Method for Metallic Bone Plates ASTM F 543-07, Standard Specification and Test Methods for Metallic Medical Bone Screws

- Substantially Equivalent Conclusion 8.
The proposed device, Reconstruction Locking Plate and 3.5mm Locking Screw, is determined to be Substantially Equivalent (SE) to the predicate device, Syntec Osteo - plate and screw Fixation, in respect of safety and effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the address for the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Mail Center - WO66-G609, Silver Spring, MD 20993-0002. The text is left-aligned and uses a serif font.

## DEC 1 6 2011

Xiamen Double Engine Medical Material Co., Ltd. % Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023 Shanghai, 200237, China

Re: K112819

Trade/Device Name: Reconstruction Locking Plate and 3.5mm Locking Screw Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliance and accessories Regulatory Class: II Product Code: HRS, HWC Dated: September 23td, 2011 Received: September 28th, 2011

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Diana Hong

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Erin Keith

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

//2819

## Section II Indications for Use

510(k) Number: Device Name: Reconstruction Locking Plate and 3.5mm Locking Screw

Indications for Use:

The Reconstruction Locking Plate and 3.5mm Locking Screw are used together and intended to treat fractures of various bones, including the clavicle and pelvis.

OPRESCRIPTION USE (Part 21 CFR 801 Subpart D) ■OVER-THE-COUNTER USE (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

| |- |

Hyllett Burr

-(Division Sign-Off)
(Division of Surgical, Orthopedic,
and Restorative Devices

510(k) Number K112819

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K112819](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K112819)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
