← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K110320

# COMPREHENSIVEPROXIMAL HUMERAL PLATING SYSTEM (K110320)

_Biomet Manufacturing Corp · HRS · Jul 21, 2011 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K110320

## Device Facts

- **Applicant:** Biomet Manufacturing Corp
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Jul 21, 2011
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Intended Use

The Comprehensive® Proximal Humeral Plating System is indicated for fractures, fracture dislocations, osteotomies and non-unions of the proximal humerus.

## Device Story

Comprehensive® Proximal Humeral Plating System is an orthopedic bone fixation device consisting of anatomic, precontoured stainless steel plates and various screw types (non-locking, locking, variable angle). Plate sizing and contouring were determined using Biomet's IntelliFIT Technology, which utilized contour analysis of cadaveric specimens to map bone patterns; this software was used only to define standard plate sizes, not for patient-specific customization. The system is intended for surgical use by orthopedic surgeons to stabilize proximal humeral fractures, dislocations, osteotomies, and non-unions. The device provides mechanical fixation to support bone healing. It is MR Conditional per ASTM F2503-08.

## Clinical Evidence

No clinical data provided. Substantial equivalence determined via engineering analysis and bench testing comparing plate strength to predicate devices.

## Technological Characteristics

Materials: Stainless Steel (ASTM F138, ASTM F139, ASTM F-2229). Components: Anatomic precontoured plates (75-255mm length, 3-4mm thickness) and 3.5mm diameter screws (10-60mm length). Sensing/Actuation: None (mechanical fixation). Connectivity: None. Sterilization: Not specified. Software: IntelliFIT Technology used for pre-defined plate sizing (non-patient-matched).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- EBI OptiLock® Upper Extremity Plating System ([K062494](/device/K062494.md))
- Synthes (USA) 3.5mm LCP Periarticular Proximal Humeral Plates ([K082625](/device/K082625.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

<110320

Image /page/0/Picture/1 description: The image shows the logo for Biomet Manufacturing Corp. The word "BIOMET" is displayed in a stylized, blocky font above the words "MANUFACTURING CORP." The text is black and the background is white. The logo appears to be a scan or photocopy, as the edges of the letters are slightly pixelated.

JUL 2 1 2011

## 510(k) SUMMARY

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92

| Submitter Information                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                                             | Biomet Manufacturing Corp.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Address                                                          | 56 East Bell Drive<br>Warsaw, IN 46581-0857                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Phone number                                                     | (574) 267-6639                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Fax number                                                       | (574) 371-1027                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Establishment<br>Registration Number                             | 1825034                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Name of contact person                                           | Patricia Sandborn Beres<br>Senior Regulatory Specialist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Date prepared                                                    | January 26, 2011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Name of device                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Trade or proprietary<br>name                                     | Comprehensive® Proximal Humeral Plating System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common or usual<br>name                                          | plate, fixation, bone screw, fixation, bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Classification name                                              | Single/multiple component metallic bone fixation appliances and accessories Smooth or threaded metallic bone fixation fastener                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification panel                                             | Orthopedics                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulation                                                       | 21 CFR 888.3030 21 CFR 888.3040                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product Code(s)                                                  | HRS HWC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Legally marketed device(s)<br>to which equivalence is<br>claimed | K062494 - EBI OptiLock® Upper Extremity Plating System<br>K082625 - Synthes (USA) 3.5mm LCP Periarticular Proximal Humeral Plates                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Reason for 510(k)<br>submission                                  | New device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Device description                                               | The Comprehensive® Proximal Humeral Plating System is<br>comprised of anatomic plates in seven lengths and non-locking,<br>locking and variable angle screws in multiple lengths. Plate<br>sizing and contouring was developed through the use of<br>Biomet's IntelliFIT Technology which uses contour analysis to<br>map patterns in complex bone on cadaveric specimens to<br>determine plate sizing. (Note, the software was used to<br>determine a set of pre-defined plate sizes and is not used to<br>create individual, patient matched plates.) |
| Intended use of the device                                       | Bone fixation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

Malling Address:
P.O. Box 5811681
P.Oresaw, M.G.G.G.G.G.G.G.G.G.G.C.C.C.
Translaw, B000.3 AM.C.S.A.G.C.C.C.
Collice: Sp. 2007.37 A.C.T.C.T.C.
Main Fax: 574.767.8137 ww.blomet.com

Shipping Address:
 56 East Bell Drive
 Warsaw, IN 46582

\$\cdot l*{D} f*{2}\$

{1}------------------------------------------------

K 110320

510(k) Summary
Comprehensive® Proximal Humeral Plating System Page 2 of 2

く

:

| Indications for use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | The Comprehensive® Proximal Humeral Plating System is<br>indicated for fractures, fracture dislocations, osteotomies and<br>non-unions of the proximal humerus. |                         |                   |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|-------------------|
| Summary of the technological characteristics of the device compared to the predicate                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                 |                         |                   |
| Characteristic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | New Device                                                                                                                                                      | Predicate Device*       |                   |
| Plate Design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Anatomic, precontoured                                                                                                                                          | K062494, K082625        |                   |
| Plate Material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Stainless Steel<br>ASTM F138 & ASTM F139                                                                                                                        | K062494, K082625        |                   |
| Plate Dimensions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Length: 75-255mm<br>Thickness: 3-4mm                                                                                                                            | K062494, K082625        |                   |
| Screw Design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Non-Locking, Locking and Variable Angle                                                                                                                         | K062494, K082625        |                   |
| Screw Material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Stainless Steel<br>ASTM F138, ASTM F139, ASTM F-2229                                                                                                            | K062494, K082625        |                   |
| Screw Dimensions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Diameter: 3.5mm<br>Length: 10-40mm or 10-60mm                                                                                                                   | K062494, K082625        |                   |
| PERFORMANCE DATA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                 |                         |                   |
| SUMMARY OF NON-CLINICAL TESTS CONDUCTED FOR DETERMINATION OF<br>SUBSTANTIAL EQUIVALENCE                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                 |                         |                   |
| Performance Test Summary-New Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                 |                         |                   |
| Characteristic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Standard/Test/FDA Guidance                                                                                                                                      | Results Summary         |                   |
| Plate Strength                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Engineering Analysis                                                                                                                                            | Equivalent to predicate |                   |
| Comparative Performance Information Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                 |                         |                   |
| Characteristic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Requirement                                                                                                                                                     | New Device              | Predicate Device* |
| Plate Strength                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Meet or exceed<br>predicate                                                                                                                                     | Meet                    | K062494           |
| SUMMARY OF CLINICAL TESTS CONDUCTED FOR DETERMINATION OF SUBSTANTIAL<br>EQUIVALENCE AND/OR OF CLINICAL INFORMATION                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                 |                         |                   |
| Clinical Performance Data/Information: None                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                 |                         |                   |
| MAGNETIC RESONANCE (MR) ENVIROMENT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                 |                         |                   |
| Biomet® has performed non-clinical Magnetic Resonance Imaging (MRI) studies on Plating<br>Systems manufactured of 316L Stainless Steel per ASTM F138. These Plating Systems are<br>determined to be MR Conditional in accordance to ASTM F2503-08 Standard Practice for<br>Marking Devices and Other Items for Safety in the Magnetic Resonance Environment. MR<br>Conditional refers to an item that has been demonstrated to pose no known hazards in a<br>specified MR environment with specified conditions of use. |                                                                                                                                                                 |                         |                   |
| CONCLUSIONS DRAWN FROM NON-CLINICAL AND CLINICAL DATA                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                 |                         |                   |
| No mechanical or clinical testing was necessary for a determination of substantial equivalence. The<br>results of engineering analysis indicated the devices performed within the intended use, did not<br>raise any new safety and efficacy issues and were found to be substantially equivalent to the<br>predicate devices.<br>*Any statement made in conjunction with this submission regarding and/or a determination of substantial equivalence to any other                                                      |                                                                                                                                                                 |                         |                   |
| product is intended only to relate to whether the product can be lawfully marketed without pre-market approval or redassification and is                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                 |                         |                   |

not intended to be Interpreted as an admission or any other type of evident infingenent litigation. [Establishment
Registration and Premarket Notification Procedures, Final R

p. 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be an abstract design of an object with three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Biomet Manufacturing Corp. % Ms. Patricia Beres Senior Regulatory Specialist 56 East Bell Drive P.O. Box 587 Warsaw, Indiana 46581-0587

JUL 2 1 2011

Re: K110320

Trade/Device Name: Comprehensive Proximal Humeral Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: June 27, 2011 Received: June 28, 2011

Dear Ms. Beres:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Patricia Beres

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerso Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known):_ k 1) ωろ乙ム

Device Name: Comprehensive® Proximal Humeral Plating System

Indications For Use: The Comprehensive® Proximal Humeral Plating System is indicated for fractures, fracture dislocations, osteotomies and non-unions of the proximal humerus.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use NO (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

for M. Melkerm

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K110320

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K110320](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K110320)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
